光谱学与光谱分析 |
|
|
|
|
|
Quick Determination of Formaldehyde Remnant in the Pharmic Intermediates by Homemade Microspectrometer |
XU Yi1,2,3,ZHANG Wen-pin1,2,TIAN Peng1,2,ZHANG Bo1,2,WEN Zhi-yu1,2,3 |
1. Microsystem Research Center, Chongqing University, Chongqing 400030, China 2. College of Chemistry and Chemical Engineering, Chongqing University, Chongqing 400030, China 3. Key Lab for Optoelectronic Technology & Systems of Ministry of Education, Chongqing University, Chongqing 400030, China |
|
|
Abstract The rapid and real-time analytical system based on photometrical methodology of a homemade holographic flat field concave grating integrated microspectrometer was established,which was applied to detect the remnant formaldehyde in the intermediate of ranitidine synthesis. Based on the chromogenic reaction of formaldehyde with MBTH and NH4Fe(SO4)2 in acid medium, the methodology of formaldehyde detection was developed. Furthermore, factors affecting the chromogenic reaction were optimized in details. Under the optimized conditions, a good linear relationship in the range of 0.04-0.20 μg·mL-1 (r=0.993 8) was obtained and the lowest detection limit achieved 0.04 μg·mL-1. In order to demonstrate the methodology and reliability of the microspectrometer analytical system, comparative experiments were carried on among the homemade microspectrometer, SHIMADZU UV-2550 spectrophotometer and U.S.A. Ocean Optic HR2000 Microspectrometer. The F-test was employed to check the reliability and efficiency of the proposed methodology. No significant difference was shown among them. The experimental results showed that the set up analytical system could satisfy the needs of quick and real-time detection in practical applications for experimental analysis, industrial monitoring and medicinal diagnosis.
|
Received: 2007-08-10
Accepted: 2007-12-20
|
|
Corresponding Authors:
XU Yi
E-mail: xuyibbd@sina.com
|
|
[1] Gaylin M Yee, Nadim I Maluf, Paul A. Sensors and Actuators, 1997, 58(1): 61. [2] FAN Shi-fu(范世福). Modern Scientific Instruments(现代科学仪器), 2000, 3: 10. [3] Stephen J Dearden. A Comparative Review of Some Commercial Spectrographs[EB/OL]. http://users.erols.com/njastro/faas/pages/paper001.htm. [4] Goldman D S, White P L, Anheier N C. Applied Optics, 1990, 29(31): 4583. [5] Morris Michael J, Walters Roy A. SPIE, 1993, 1796: 141. [6] WEN Zhi-yu, CHEN Gang, WEN Zhong-quan, et al(温志渝, 陈 刚, 温中泉, 等). Acta Optica Sinica(光学学报), 2003, 23(6): 740. [7] CHEN Gang, WEN Zhi-yu, WEN Zhong-quan, et al(陈 刚, 温志渝, 温中泉, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2003, 23(6): 1232. [8] HU Song, WEN Zhi-yu, LIANG Yu-qian, et al(胡 松, 温志渝, 梁玉前, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2006, 26(9): 1769. [9] CHEN Gang, WEN Zhi-yu, WEN Zhong-quan, et al(陈 刚, 温志渝, 温中泉, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2005, 25(3):439. [10] ZHANG Bo, WEN Zhi-yu(张 波, 温志渝). Semiconductor Optoelectronic(半导体光电),2007, 28(1): 147. [11] XU Bo-hong, YAN Hui-fang(徐伯洪,闫慧芳). Methods for Monitoring Hazardous Substances in the Air of Workplace(工作场所有害物质监测方法). Beijing: Chinese People’s Public Security University Press(北京:中国人民公安大学出版社), 2003. 191. [12] XIONG Yu-hong, WEN Zhi-yu, CHEN Gang, et al(熊宇虹, 温志渝, 陈 刚, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2005, 25(8): 1332. [13] CHEN Gang, WEN Zhi-yu, LIANG Yu-qian, et al(陈 刚, 温志渝, 梁玉前, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2005, 25(3): 346. |
[1] |
ZONG Zhi-fang1, XU Wei-cheng2, CHEN De-peng1*, TANG Gang1, ZHOU Xiao-hui1, DONG Wei1, WU Yu-xi2. Preparation Mechanism of Decylic Acid-Palmitic Acid/SiO2@TiO2
Photocatalytic Phase Change Microcapsules Based on
Multiple Spectrum Analysis[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(04): 1306-1313. |
[2] |
LIU Guo-hua, LI Qi-hua*, OU Jin-ping, XU Heng, ZHU Peng-cheng, LIU Hao-ran. Passive Spectrum Measurement of HCHO in Chongqing Area Based on MAX-DOAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(01): 243-247. |
[3] |
HE Qi-xin, LI Jia-kun, FENG Qi-bo*. Development of a Mid-Infrared Cavity Enhanced Formaldehyde Detection System[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(07): 2077-2081. |
[4] |
YANG Chuan-xiao, GONG Wei-bin, TANG Fan, SUN Xiang-ying. Determination of Sodium Hexametaphosphate by Ratiometric Fluorescence Method Based on Formaldehyde Functionalized Polyethyleneimine/Eosin Y System[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(02): 454-459. |
[5] |
ZHU Hong-qiu, HU Hao-nan, ZHENG Guo-liang, ZHOU Can*, LI Yong-gang. High-Quality UV-Vis Spectrum Signal Reconstruction Algorithms Based on Binary Search[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(06): 1979-1983. |
[6] |
ZHANG Hao1,2, GAO Qing1, HAN Xiang-xiang1, RUAN Gao-yang1, LIU Xiu-yu1. Mechanism Analysis of Formaldehyde Degradation by Hot Braised Slag Modified Activated Carbon Based on XRF and XRD[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(05): 1447-1451. |
[7] |
ZHANG Hao1, 2, 3, ZHANG Lei3, LONG Hong-ming1, 2*. Spectroscopic Analysis of Preparation of Ecological Activated Carbon Based on Electric Furnace Slag Ultrafine Powder Modified Biomass Waste Material[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(03): 861-866. |
[8] |
ZHENG Guo-liang, ZHU Hong-qiu*, LI Yong-gang. Spectral Signal Denoising Algorithm Based on Improved LMS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(02): 643-649. |
[9] |
WEI Min-hong1,2, LIU Cheng2*, LI Su-wen1, CHEN Zheng-hui1, MOU Fu-sheng1. Measurement of Tropospheric HCHO by MAX-DOAS Based on QDOAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(08): 2332-2336. |
[10] |
CAO Si-min, LIU Yang-yi, ZHOU Zhong-neng, CHEN Jin-quan, XU Jian-hua*. Study on Spectral Characteristics of a Novel Formaldehyde Probe Based on Fluoral-P Derivatives[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(03): 828-833. |
[11] |
LIU Xing-bin, WANG Yue*, HAN Xiao-ri*, MA Bin. Feasibility of Rapid Evaluation of Urea-Formaldehyde Fertilizer by Infrared Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(03): 750-755. |
[12] |
XIE Ying-ke1,2,3, WEN Qun1,2*, WEN Zhi-yu1,2, MO Zhi-hong1,2, WEI Kang-lin1,2. On-Line Determination of Chemical Oxygen Demand and Total Phosphorus in Water by Using Fenton Reagents via a Micro-Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(05): 1642-1648. |
[13] |
SHEN Shi-liang1, WANG Shan-shan1*, ZHOU Bin2 . Investigation of Atmospheric Formaldehyde and Glyoxal Based on Differential Optical Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(08): 2384-2390. |
[14] |
LI Jia-jia, LI Rui, MA Xue, FAN Ting-ting, LIN Yi-qing, SONG Yu-ting, HAN Jing-qi, LI Xia* . Fluorescence Properties and the Detection of Benzaldehyde of Lanthanide Complex with 2-Sufoterephthalalic Acid [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(02): 345-348. |
[15] |
LI Yang-jun1, JIAO Ning2, WANG Rui-jun2, LEI Shu-feng2, WANG Gao1 . Research on Formaldehyde Concentration Detection System Based on Photo-Elastic Modulation[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(12): 3308-3311. |
|
|
|
|