The Evaluation of Uncertainty in the Results for Elements Rubidium, Strontium, Yttrium and Zirconium in Silicate Geological Samples by Polarized Energy Dispersive X-Ray Fluorescence Spectrometry
WANG Yi-ya,ZHAN Xiu-chun,YUAN Ji-hai,FAN Xing-tao
National Research Center for Geoanalysis,Beijing 100037, China
Abstract:A method for evaluation of uncertainty was established with standard deviation of relative error. Utilizing a polarized energy dispersive X ray fluorescence spectrometer(P-EDXRF)X-lab 2000 with pressed polyethylene-backed pellets, 76 national reference materials and 89 geological examination samples were analyzed, the results indicated that the relative errors consist with the normal distribution with confidence level 95%. The section standard deviations of relative errors acted as method global relative uncertainty and expanded factor was 2. The section relative uncertainty caused by precision was analyzed and relative uncertainty caused by accuracy based on the error transfer formula was isolated. The ratio of relative uncertainty caused by accuracy to the global relative uncertainty was different with different levels and elements. Two methods validated that the evaluation of global uncertainty is reasonable, with the first method being the formula of audited results in laboratory, and the second being the comparison of standard value with expanded uncertainty and a revised value with expanded uncertainty.
王祎亚,詹秀春,袁继海,樊兴涛. 地质样品铷锶钇锆元素偏振能量色散X射线荧光光谱分析结果不确定度的评估 [J]. 光谱学与光谱分析, 2011, 31(06): 1707-1711.
WANG Yi-ya,ZHAN Xiu-chun,YUAN Ji-hai,FAN Xing-tao. The Evaluation of Uncertainty in the Results for Elements Rubidium, Strontium, Yttrium and Zirconium in Silicate Geological Samples by Polarized Energy Dispersive X-Ray Fluorescence Spectrometry. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(06): 1707-1711.
[1] Yamini Y,Amiri N,Karimi M. X-Ray Spectrometry,2009,38:474. [2] Aydin Gunes,Ali Inala,Yusuf K Kadioglu. X-Ray Spectrometry,2009,38:451. [3] Kitov B I,Mukhachev Yu S. X-Ray Spectrometry,2010,39:32. [4] Gazulla M F,Gómez M P,Ordua M,et al . X-Ray Spectrometry,2009,38:3. [5] Yasuhiro Shibata,Junnosuke Suyama,Masaru Kitano,et al. X-Ray Spectrometry,2009,38:410. [6] Cevik U,Koz B,Makarovska Y. X-Ray Spectrometry,2010,39:202. [7] REN Chun-sheng,ZHANG Ai-zhen,YING Hai-song,et al(任春生,张爱珍,应海松,等). Metal Mine(金属矿山),2008,379(1):76. [8] LI Yu-wu,MA Li(李玉武,马 莉). Rock and Mineral Analysis(岩矿测试),2007,26(3):219. [9] CHEN Ai-ping,WANG Ye,WANG Su-ming(陈爱平,王 烨,王苏明). Rock and Mineral Analysis(岩矿测试),2006,25(3):270. [10] HU Zheng-yang,XING Hua-bao,SHI Hou-yi,et al(胡正阳,邢华宝,史厚义,等). Metallurgical Analysis(冶金分析),2005,25(1):82. [11] JIN De-long,WANG Cheng-zhong,LU Xiao-ming,et al(金德龙,王承忠,陆晓明,等). Metallurgical Analysis(冶金分析),2005,25(3):88. [12] Sieber J R,Yu L L,Marlow A F,et al. X-Ray Spectrometry,2005,34:153. [13] JI Ang,TAO Guang-yi,ZHUO Shang-jun,et al(吉 昂,陶光仪,卓尚军,等). X-Ray Fluorescence Spectrometry Analysis(X射线荧光光谱分析). Beijing:Science Press(北京:科学出版社),2003. 218. [14] Rousseau R M. The Rigaku Journal,2001,18(2):33. [15] Wegrzynek D,Markowicz A,Chinea-Cano E. X-Ray Spectrom,2003,32:317. [16] Rashmi,Dhawan U,Suri D K. X-Ray Spectrom,1999,28:157. [17] Padilla lvarez R,Markowicz A,Wegrzynek D,et al. X-Ray Spectrometry,2007,36:27. [18] Accreditation Criteria for Testing and Calibration Laboratories(ISO/IEC 17025—2005)(检测和校准实验室认可准则). CNAL(中国实验室认可委员会),2005. 13. [19] WANG Yi-ya,ZHAN Xiu-chun,LIU Yi-jian,et al(王祎亚,詹秀春,刘以建,等). Chinese Journal of Analysis Laboratory(分析试验室),2009,28(9):90. [20] DZ/T 0130—2006. The Specification of Testing Quality Management for Geological Laboratories(地质矿产实验室测试质量管理规范). Geology and Mineral Resources of the Industry Standard of the People’s Republic of China(中华人民共和国地质矿产行业标准),2006.