Development of Multi-Spectral Thermometer for Explosion Flame True Temperature Measurement——Field Experiments and Measurement Accuracy Analysis
SUN Kun1, 2, SUN Xiao-gang3, YU Xiao-yang1, SUN Xiao-ming1
1. The Higher Educational Key Laboratory for Measuring & Control Technology and Instrumentations of Heilongjiang Province, Harbin University of Science and Technology, Harbin 150080, China 2. Key Laboratory of Engineering Dielectric and Its Application, Ministry of Education, Harbin University of Science and Technology, Harbin 150080, China 3. Instrument Science and Technology, Harbin Institute of Technology, Harbin 150001, China
摘要: 研制了可用于大型爆炸现场的、 测量爆炸火焰真温的多光谱温度计(量程为800~3 500 ℃,波长范围为0.6~1.1 μm)。测量原理进一步改进,加入亮温逼近法解决了二次测量法初值选取困难的问题,并应用此高温计在空旷场地对3 kg TNT炸药爆炸的全过程进行测量。通过实验结果的分析可知,此高温计可以测量爆炸火焰真温变化全过程,对波阵面瞬时温度与燃烧火球温度的测量均具有很好的效果。同时,分析了影响此高温计测量精度的各个因素,得出目前制约多光谱高温计测量精度提高的主要因素仍然为真温算法及标定方法的误差,这为今后研制高精度高温计明确了方向。
关键词:多光谱;爆炸火焰;真温测量;TNT;测量不确定度
Abstract:In the present article, the measurement principle of a multi-target multi-spectral thermometer developed previously is improved in that the brightness temperature approximation method is adopted to overcome the second measurement method initial value selection difficulty. The thermometer is used to measure the explosion true temperature of 3KG TNT explosive in open fields. The experimental results show that this pyrometer can measure changes in the explosion flame true temperature in the whole process, and functions well in the wavefront instantaneous temperature measurement and burning fireball temperature measurement. Finally, through the analysis of the factors which affect the thermometer measurement accuracy, it was concluded that the main error sources restricting the measurement accuracy improvement are still the true temperature algorithm error and the calibration method error. The analysis result has provided the direction for the high-precision thermometer development.
Key words:Multi-spectral;Explosion flame;True temperature measurement;TNT explosive;Measurement uncertainty
孙 崐1, 2,孙晓刚3,于晓洋1,孙晓明1. 测量爆炸火焰真温的多光谱温度计的研制——现场实验与测量精度分析[J]. 光谱学与光谱分析, 2013, 33(06): 1719-1722.
SUN Kun1, 2, SUN Xiao-gang3, YU Xiao-yang1, SUN Xiao-ming1. Development of Multi-Spectral Thermometer for Explosion Flame True Temperature Measurement——Field Experiments and Measurement Accuracy Analysis. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(06): 1719-1722.
[1] DAI Jing-min, LU Xiao-dong(戴景民, 卢小冬). Journal of Infrared and Millimeter Waves(红外与毫米波学报), 2000, 19(1): 62. [2] Kenneth,Cashdollar L, Isaac A Zlochower. Journal of Loss Prevention in the Process Industries, 2007, 20: 337. [3] Sun Xiaogang, Yuan Guibin, Dai Jingmin. International Journal of Thermophysics,2005, 26(4): 1255. [4] SUN Xiao-gang, DAI Jing-min, WANG Xue-feng(孙晓刚, 戴景民, 王雪峰). Journal of Infrared and Millimeter Waves(红外与毫米波学报),2003, 22(2): 141. [5] ZHAN Jian-ming, WEN De-sheng, WANG Hong(占建明, 汶德胜, 王 宏). Semiconductor Technology(半导体技术),2011, 36(4): 304. [6] YU Hang, JIAN Tian-cheng, WANG Hong-jin(于 航, 建天成, 王宏瑾). Modern Electronics Technique,2011, 34(2): 196. [7] SUN Xiao-gang, SUN Kun, DAI Jing-min(孙晓刚, 孙 崐, 戴景民). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2012, 32(1): 273. [8] SUN Xiaogang, ZHAO Wei, YUAN Guibin. Journal of Harbin Institute of Technology (New Series),2006, 13(4): 421.