Abstract:A novel chemiluminescence method for the determination of Levofloxacin is based on its enhancement effect on the reaction of luminol with H2O2. The calibration graph is linear over the range of 5.53×10-11-2.21×10-8 mol·L-1 and the detection limit is 1.38×10-11 mol·L-1. The relative standard deviation (RSD) is 2.56% (n=9). The method has been successfully applied to the determination of Levofloxacin in capsule samples.
[1] Hairui L, Kays M B, Sowinski K M. J. Chromatogr. B, 2002, 772: 53. [2] YIN Guo-zhen, JIN Ou, LI Hui-lin et al(殷国真,金 瓯,李会林等). J. North-west Pharm.(西北药学杂志),1998,13(5):195. [3] FU Chun-mei, LI Zhang-wan(付春梅,李章万). J. of Pharm. Analysis(药物分析杂志),2000,20(6):373. [4] CI Wei, CHAI Yi-feng(慈 薇,柴逸峰). J. of Pharm. Analysis(药物分析杂志),2000,20(4):240. [5] ZHU Ka-lin, XU Qing(朱卡林,徐 庆). J. West-China Pharm.(华西药学杂志),1999,14(2):124. [6] WU Dong-fang, SUN Yan-rong(吴东方,孙艳蓉). J. West-China Pharm.(华西药学杂志),2000,15(2):103. [7] DU Li-ming, JIN Wei-jun, DONG Chuan et al(杜黎明,晋卫军,董 川等). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2001,21(4):518.