Determination of Total Selenium in Water by Hydride Generation Atomic Fluorescence Spectrometry
XU Yong-xin1, 2, LUO Kun-li1*
1. Institute of Geographic Sciences and Natural Resources Research, Chinese Academy of Sciences, Beijing 100101, China 2. Graduate University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:In order to establish a rapid and accurate hydride generation-atomic fluorescence spectrometry method for the determination of selenium content in water samples, the influence factors on Se nitrolysis in the water such as nitrolysis temperature and the amount of hydrochloric acid were studied. The authors determined a simple and practical test method that is to add 3 mL hydrochloric acid at 3 ℃. We measured the precision and accuracy of the method by the standard liquid of 3, 6 and 10 μg·L-1, and the test results were 2.96, 5.43 and 9.66 mg·L-1.The recovery varied from 90.50% to 98.67%. Meanwhile we tested 29 water samples by different methods. The error of standard sample was 8.95%~25.46%by hydride generation-atomic fluoresence spectrometry in the GB/T 5750.6—2006, and the error of standard sample was 1.33%~3.40% by new method. Compared with the national standards of GB/T 5750.6—2006, the method has the advantages of operating simplicity and and easy control, and it is also reliable and economical.
Key words:Fluorescence spectroscopy;Test method;Toal selenium in water;Improved
徐永新1, 2,雒昆利1* . 氢化物发生-原子荧光光谱法对水中总硒含量测试方法改进研究[J]. 光谱学与光谱分析, 2012, 32(02): 532-536.
XU Yong-xin1, 2, LUO Kun-li1* . Determination of Total Selenium in Water by Hydride Generation Atomic Fluorescence Spectrometry . SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(02): 532-536.
[1] Helina H. Journal of Trace Elements in Medicine and Biology, 2005,18: 305. [2] Sirichakwal P, Puwastien P, Polngam J, et al. Journal of Food Composition and Analysis, 2005, 18: 47. [3] Chand V, Prasad S. Journal of Hazardous Materials, 2009, 165: 780. [4] ZHAO Mei-song, MAO Li-zhong(赵美松, 毛礼钟). Bulletin of Biology(生物学通报), 1996, 31(6): 15. [5] Gamiz-Gracia L, Luque M D. Analytical Chemistry Division, 1999, 20:10. [6] Araz Bidari, Elham Zeini Jahromi, Assadi Yaghoub, et al. Microchemical Journal, 2007, 87: 6. [7] Mehdi Ashournia, Alireza Aliakbar. Journal of Hazardous Materials, 2009, 168: 542. [8] YANG Xiu-lin(杨秀琳). Chemical Analysis and Meterage(化学分析计量), 2004, 13(1): 32. [9] XU Kun-ming, WANG Jia, DENG Hao, et al(许昆明, 王 佳, 邓 浩, 等). Journal of Preventive Medicine Information(预防医学情报杂志), 2005, 21(1): 126. [10] LI Jing-na, LIANG Gao-dao, WU Yan, et al(李静娜, 梁高道, 吴 衍, 等). Chinese Journal of Health Laboratory Technology(中国卫生检验杂志), 2001, 11(1): 51. [11] FAN Zhi-hong(樊智虹). Environmental Science & Technology(环境科学与技术), 2011, 34(1): 117. [12] Standard Examination Methods for Drinking Water Metal Parameters(生活饮用水标准检验方法金属指标). National Standards of the People’s Republic of China(中华人民共和国国家标准)GB/T 5750.6-2006. [13] LUO Kun-li, XU Li-rong, XIANG Lian-hua(雒昆利, 徐立荣, 向连华). Acta Geologica Sinica(地质学报), 2002, 76(3): 389. [14] Andersen M B, Stirling C H, Porcell D, et al. Earth and Planetary Science Letters, 2007, 259(1-2): 171.