光谱学与光谱分析 |
|
|
|
|
|
System Parameters Selection and Optimization of Tunable Diode Laser Absorption Spectroscopy |
GAO Nan1, DU Zhen-hui1*, TANG Miao2, YANG Jie-wen1, YANG Chun-mei1, WANG Yan1 |
1. College of Precision Instruments and Opto-Electronics Engineering, Tianjin University, Tianjin 300072, China 2. Tianjin Environmental Monitoring Center, Tianjin 300191, China |
|
|
Abstract The system performance of tunable diode laser absorption spectroscopy (TDLAS) is affected by the modulation parameters such as modulation index, modulation frequency, scanning amplitude and scanning frequency. There is a lack of definite parameters selection basis in practical measurement. Aiming at this problem, the influence of modulation parameters on second harmonic signals was observed by experiment based on a certain theory in the present paper, and the basis and method of modulation parameters optimization for various system functions and demands were summarized by analyzing the signal characteristic including amplitude, signal to noise ratio, symmetry and peak width. For the system of concentration or temperature detection the amplitude and signal to noise ratio will be taken into prior consideration which require optimum modulation index, lower modulation frequency and lower scanning frequency. In condition of pressure detection deduced by lineshape the signal symmetry and peak width are more important to ascertain the modulation parameters according to practical demands. Scanning amplitude will be adjusted to obtain complete signal waveforms, then scanning frequency can be adjusted according to system speed and accuracy requirement. The result of the experiment provided a definite basis for conforming the working state of such system.
|
Received: 2010-03-06
Accepted: 2010-06-08
|
|
Corresponding Authors:
DU Zhen-hui
E-mail: duzhenhui@tju.edu.cn
|
|
[1] Hinkley E D. Appl. Phys. Lett., 1970, 16(9): 351. [2] Reid J, Wchun J S, Garside B K, et al. Appl. Opt., 1978, 17: 300. [3] DU Zhen-hui, ZHAI Ya-qiong, LI Jin-yi, et al(杜振辉,翟雅琼,李金义,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2009, 29(12): 3199. [4] Werle P, Spectrochimica Acta Part A, 1998, 54: 197. [5] Silver J A. Appl. Opt., 1992, 31: 707. [6] Supplee J M, Whittaker E A, Lenth W. Appl. Opt., 1994, 33: 6294. [7] Kluczynski P, Axner O. Appl. Opt., 1999, 38: 5803. [8] Reid J, Labrie D. Appl. Phys. B, 1981, 26: 203. [9] Axner O, Kluczynski P, Lindberg A M. Journal of Quant. Spectrose. Radiant. Trans., 2001, 68: 299. [10] Silver J A, Kane D J. Meas. Sci. Technol., 1999, 10: 845. [11] SHAO Jie, GAO Xiaoming, ZHANG Weijun, et al. Chinese Physics Letter, 2005, 22(1): 57.
|
[1] |
ZHANG Le-wen1, 2, WANG Qian-jin1, 3, SUN Peng-shuai1, PANG Tao1, WU Bian1, XIA Hua1, ZHANG Zhi-rong1, 3, 4, 5*. Analysis of Interference Factors and Study of Temperature Correction Method in Gas Detection by Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 767-773. |
[2] |
SHAO Guo-dong1, LI Zheng-hui1, GUO Song-jie1, ZOU Li-chang1, DENG Yao1, LU Zhi-min1, 2, 3, YAO Shun-chun1, 2, 3*. Research on Gas Concentration Measurement Method Based on Gradient Descent Method to Fit Spectral Absorption Signal Directly[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(10): 3256-3261. |
[3] |
SHANG Hou-fei1, DU Zhen-hui2, 3, GAO Nan1, 3*, LI Jin-yi4, XIAO Yan-jun1, MENG Zhao-zong1, ZHANG Zong-hua1. Parameter Analysis of Tunable Diode Laser Absorption Spectroscopy Based on Spectral Frequency Characteristics[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(11): 3359-3364. |
[4] |
XIAO Hu-ying1, YANG Fan1, XIANG Liu1, HU Xue-jiao2*. Jet Vacuum Enhanced Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(10): 2993-2997. |
[5] |
LI Zheng-hui1,3, YAO Shun-chun1,3*, LU Wei-ye2, ZHU Xiao-rui1,3, ZOU Li-chang1,3, LI Yue-sheng2, LU Zhi-min1,3. Study on Temperature Correction Method of CO2 Measurement by TDLAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(07): 2048-2053. |
[6] |
LI Meng, GUO Jin-jia*, YE Wang-quan, LI Nan, ZHANG Zhi-hao, ZHENG Rong-er. Study on TDLAS System with a Miniature Multi-Pass Cavity for CO2 Measurements[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(03): 697-701. |
[7] |
WANG Zhe1, WANG Yan1, ZHANG Rui2, ZHAO Xue-hong1*, LIU Qiao-jun2, LI Cong-rong2 . A Singular Value Decomposition Method for Tunable Diode Laser Absorption Spectroscopy System to Remove Systematic Noises[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(10): 3369-3376. |
[8] |
YAN Jie, ZHAI Chang, WANG Xiao-niu, HUANG Wen-ping . The Research of Oxygen Measurement by TDLAS Based on Levenberg-Marquardt Nonlinear Fitting[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(06): 1497-1500. |
[9] |
QI Ru-bin1, HE Shu-kai1, LI Xin-tian1, WANG Xian-zhong2 . Simulation of TDLAS Direct Absorption Based on HITRAN Database [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(01): 172-177. |
[10] |
CAI Ting-dong1, 2, GAO Guang-zhen1, WANG Min-rui1, WANG Gui-shi2, GAO Xiao-ming2 . Measurements of CO2 Concentration at High Temperature and Pressure Environments Using Tunable Diode Laser Absorption Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34(07): 1769-1773. |
[11] |
SHAO Xin1, 2, LIU Fu-gui1, CHEN Wen-liang3 . Comparative Study on Software Demodulation for Continuous Wave and Quasi-Continuous Wave Wavelength Modulation Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(12): 3419-3424. |
[12] |
LI Han, LIU Jian-guo, HE Ya-bai, HE Jun-feng, YAO Lu, XU Zhen-yu, CHEN Jiu-ying, YUAN Song, KAN Rui-feng* . Simulation and Analysis of Second-Harmonic Signal Based on Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(04): 881-885. |
[13] |
LIU Qing-jie1, 2, JING Lin-hai1, 2, WANG Meng-fei3, LIN Qi-zhong1, 2 . Hyperspectral Remote Sensing Image Classification Based on SVM Optimized by Clonal Selection[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(03): 746-751. |
[14] |
DOU Wen1, SUN Hong-quan2, CHEN Yun-hao2* . Comparison among Remotely Sensed Image Fusion Methods Based on Spectral Response Function[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(03): 746-752. |
[15] |
ZHANG Shuai, DONG Feng-zhong*, ZHANG Zhi-rong,WANG Yu, KAN Rui-feng,ZHANG Yu-jun,LIU Jian-guo, LIU Wen-qing. Monitoring of Oxygen Concentration Based on Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(10): 2593-2596. |
|
|
|
|