Abstract:A DNA analysis device based on poly(methyl methacrylate) (PMMA) microfluidic chips was developed. A PMMA chip with cross microchannels was fabricated by a simple hot embossing. Microchannels were modified with a static adsorptive coating method using 2% hydroxyethyl cellulose. A high-voltage power unit, variable in the range 0-1 800 V, was used for on-chip DNA sample injection and gel electrophoretic separation. High speed, high resolution DNA analysis was obtained with the home-built PMMA chip in a sieving matrix containing 2% hydroxyethyl cellulose with a blue intercalating dye, TO-PRO-3 (TP3), by using diode laser induced fluorescence detection based on optical fibers with a 670 nm long-pass filter. The DNA analysis device was applied for the separation of φX-174/HaeⅢ DNA digest sample with 11 fragments ranging from 72 to 1 353 bp. A separation efficiency of 1.14×106 plates/m was obtained for the 603 bp fragments, while the R of 271/281 bp fragments was 1.2. The device was characterized by simple design, low cost for fabrication and operation, reusable PMMA chips, and good reproducibility. A portable microfluidic device for DNA analysis can be developed for clinical diagnosis and disease screening.
[1] Vilkner T, Janasek D, Manz A. Anal. Chem., 2004, 76: 3373. [2] Whitesides G M. Nature, 2006, 442: 368. [3] Dittrich P S, Tachikawa K, Manz, A. Anal. Chem., 2006, 78: 3887. [4] Chen Y, Zhang L Y, Chen G. Electrophoresis, 2008, 29: 1801. [5] Chen R, Guo H Z, Shen Y W, et al. Sens. and Actuators B: Chemical, 2006, 114: 1100. [6] Woolley A T, Mathies R A. Anal. Chem., 1995, 67: 3676. [7] Chen L, Manz A, Day P J R. Lab Chip, 2007, 7: 1413. [8] DU Xiao-guang, GUAN Yan-xia, WANG Fu-ren, et al(杜晓光, 关艳霞, 王福仁, 等). Chem. J. Chinese Universities(高等学校化学学报), 2003, 24: 1962. [9] Du X G, Fang Z L. Electrophoresis, 2005, 26: 4625. [10] XU Zhang-run, WANG Shi-li, FAN Xiao-feng, et al(徐章润,王世立, 樊晓峰, 等). Chinese Anal. Chem.(分析化学), 2003, 31: 1527. [11] ZHANG Hua-shan, WANG Hong, ZHAO Yuan-yuan(张华山, 王 红, 赵媛媛). Molecule Probes and Detection Reagents(分子探针与检测试剂). Beijing: Science Press(北京: 科学出版社), 2002. 318. [12] Rampal S, Liu M, Chen F A. J. Chromatogr. A, 1997, 781: 357.