Abstract:A new flow injection electrogenerated chemiluminescence (ECL) method for the determination of genyamicin sulfate was developed.The method is based on the sensitizing effect of genyamicin sulfate on the weak ECL signal of electrochemical oxidation of luminol on the platinum flake electrode in the medium of Na2B4O7.The experimental conditions that influence the ECL intensity were investigated and optimized.Under the optimum conditions, the ECL response is in the range of 1.2×10-9-4.0×10-6 g·mL-1 genyamicin sulfate with a detection limit of 8.0×10-10 g·mL-1.The relative standard deviation is 2.0%.This method was used to determine genyamicin sulfate in injection samples with satisfactory results.
[1] Pharmacopoeia Committe of Ministry of Public Health, the People's Republic of China(中华人民共和国卫生部药典委员会编).The Pharmacopoeia of the People's Republic of China(Part Ⅱ)(中华人民共和国药典·二部).Beijing: Chemical Industry Press(北京: 化学工业出版社),1995.902. [2] JIAO Geng-sheng, LI Yuan, LANG Hui-yun(焦更生,李 媛,郎惠云).Chinese Journal of Analysis Laboratory(分析试验室),2001,20(6): 55. [3] HUAI You-wang, LIAN Dong-liu, YUE Sun, et al.Talanta, 2000,52(2): 201. [4] ZHANG Qiang, LIAO Gong-tie, SHOU Dan(张 强,廖工铁,寿 旦).Chinese J.of Hospital Pharmacy(中国医院药学杂志),1995,15(4): 147. [5] MENG Fan-de, ZHAO Li, LIU Xue-qi(孟凡德,赵 莉,刘学启) .Northwest Pharm.J.(西北药学杂志),1997,12(3): 103. [6] JIANG Hong, LIU Shao-pu, HU Xiao-li(江 虹,刘绍璞,胡小莉).Chinese Journal of Analytical Chemistry(分析化学),2003,31(9): 1053. [7] Karsten A Fahnrich, Miloslav Pravda, George G Guilbault.Talanta, 2001, 54: 531. [8] HUANG Bing-qiang, GUO Wen-ying, XU Yang, TU Yi-feng(黄炳强,郭文英,徐 扬, 屠一锋).Spectroscopy and Spectral Analysis(光谱学与光谱分析),2003,23(5): 849.