Abstract:The determination method for trace Pb in whole blood was studied, which was diluted by 0.5% HNO3. In the measurement, Zeeman-GFAAS was used and the relative condition was discussed. This method shows that the detection limit was 10.0 pg, the relative standard deviation of standards (RSD) was 0.36%-1.8%,the relative standard deviation of samples (RSD) was 1.4%-2.3%, and the recovery rate was 96.0%-101.7%. The concurrent interference was low. This method is simple and rapid. It is an ideal method to determine trace Pb in whole blood.
[1] KAI Xiao-ming, ZHANG Qun, LING Bi-wen,CHEN You-cun(开小明,张 群,凌必文, 陈友存). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2001,21(2):244. [2] Miier D T, Paschal D C, Gunter E W et al. Analyst, 1987, 112: 1701. [3] JIN Mi-cong,WANG Li et al(金米聪,王 立等). Physical Testing and Chemical Analysis, Part B: Chemical Analysis(理化检验·化学分册)2002,38(6):293. [4] QIN Hong, LI Xin(秦 红,李 鑫). Chinese Journal of Public Health(中国公共卫生学报),1999,18(4):216.