Study On Resistance to Motion Artifact in Pulse Oximetry Measurement Using Segment Filter Method
LI Qing-bo1, NIE Xin1, ZHANG Guang-jun1*, WU Jin-guang2
1. Key Laboratory of Precision Opto-Mechatronics Technology, Ministry of Education, College of Instrument Science and Opto-Electronics Engineering,Beihang University, Beijing 100191, China 2. The State Key Laboratory of Rare Earth Materials and Applications, College of Chemistry and Molecular Engineering, Peking University, Beijing 100871, China
Abstract:The authors proposed a method which combined the differential threshold segment with median filter to eliminate abrupt artifact motion interference in the use of pulse oximeter, and further more, by adding the window moving average filter, this method could reduce the periodic artifact motion interrupt in some way. In the experiment, the authors intently simulated the common finger motion to get the artifact motion “corrupted” signal with our self-made reflectance oximeter system. The comparison of the original PPG signal and processed PPG signal shows that the method we proposed performed well in an effort to reduce the common artifact motion interrupt. Finally, in order to figure out what effort this method can make for the accuracy of oxygen saturation measurement, the ration of AC and DC component in the red and infrared signal directly associated with the pulse oxygen saturation was calculated. The result indicates that this method can reduce the error caused by the artifact motion and stabilize the rate of false reading of pulse oximeter under the finger motion state. More importantly, this method needs only a small calculation cost, so it could be applied in the real-time PPG signal processing without the need of powerful hardware.
Key words:Oxygen saturation;Motion artifact;Differential threshold;Median filter;Windows moving average
李庆波1,聂 鑫1,张广军1*,吴瑾光2 . 分段滤波用于消除脉搏血氧检测中多种运动干扰的研究[J]. 光谱学与光谱分析, 2009, 29(11): 2994-2997.
LI Qing-bo1, NIE Xin1, ZHANG Guang-jun1*, WU Jin-guang2 . Study On Resistance to Motion Artifact in Pulse Oximetry Measurement Using Segment Filter Method . SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(11): 2994-2997.
[1] HUANG Jian-xin, LIU Huai(黄建新,刘 怀). Journal of Nanjing Normal University(南京师范大学学报), 2006, 6(1): 17. [2] Yitzhak Mendelson. Burt D Ochs. IEEE Transactions on Biomedical Engineering, 1988, 35(10): 798. [3] DAI Li-juan, WANG Hui-nan, QIAN Zhi-yu, et al(戴丽娟,王惠南,钱志余,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2008,28(7): 1469. [4] Rhee S, Yang B, Asada H H. IEEE Transactions on Biomedical Engineering, 48(7): 795. [5] Constantine Dumas, Joyce A Wahr, et al. Anesth Analg.,1996,83: 269. [6] Gary Comtois, Yitzhak Mendelson, Piyush Ramuka. Proceeding of IEEE/EMBS,Conference,Lyon France, IEEE Press, 2007. 1528. [7] Yao Jianchu, Steve Warren. Proceedings of the IEEE/EMBS Conference, San Francisco. CA USA, IEEE Press, 2004. 2153. [8] Graybeal J M, Petterson M T. Proceedings of the IEEE/EMBS Conference, San Francisco, CA USA, IEEE Press, 2004. 5363. [9] Diab. US Patent, 5482036, 1996. [10] Relente A R, Sison L G. Proceeding of the Joint IEEE/EMBS Conference,Houstom TX, USA, IEEE Press, 2002. 1769. [11] Sang Hyun Kim, Dong Wan, Changseok Bae. Proceedings of IEEE/EMBS Conference, Lyon France, 2007, 2567. [12] Seyedtabaii S, Seyedtabaii L. International Journal of Electronics,2007, 2(1): 31. [13] Levi B Wood, Harry H Asada. Proceeding of IEEE/EMBS Conference,Lyon France, IEEE Press,2007. 652. [14] Stetson P F. Proceeding of the IEEE/EMBS Conference, San Francisco, CA, USA, IEEE Press, 2004. 231. [15] Byung S Kim, Sun K Yoo. IEEE Transactions on Biomedical Engineering, 2006, 53(3): 566. [16] YANG Zhu-qing, LI Yong, HU De-wen(杨竹青,李 勇,胡德文). Acta Automatica Sinica(自动化学报),2002, 28(5): 762. [17] Yao Jianchu, Steve Warren. Proceeding of IEEE-Engineering in Medicine and Biology Confereoce,2005. 3585. [18] Matthew James Hayes,Peter R Smith. IEEE Transaction on Biomedical Engineering, 2001, 48(4): 452. [19] Johnston W S, Mendelson Y. Proceedings of IEEE/EMBS Conference, New York City, USA, IEEE Press, 2006. 5888. [20] YU Wei, GU Qing-en, YAO Xiang,et al(于 巍,古庆恩,姚 翔,等). Chinese Medical Equipment Journal(医疗卫生装备),2007,28(10): 4. [21] Lee Hanwook, Lee Juwon, Jung Wongeun, et al. International Journal of Control, 2007, 5(6): 701. [22] CAO Zhi-hua, SONG Bin-heng(曹治华,宋斌恒). Application Research of Computers(计算机应用研究),2006,23(3): 86. [23] ZHU Bing-lian, PAN Zhe-ming, LI Dan-dan(朱冰莲,潘哲明,李单单). Signal Processing(信号处理),2008,24(4): 684. [24] WANG Yi-yan, WANG Xuan(王益艳,王 晅). Computer Engineering and Design(计算机工程与设计),2008,29(11): 2841. [25] Lee Juwon, Jung Wongeun, Kang IkTae, et al. Computer Standards & Interfaces, 2004, 26: 241. [26] George Zonios, Uday Shankar, Vijay K Iyer. IEEE Transactions on Biomedical Engineering, 2004, 51(5): 818.