Determination of Rare-Earth and Trace Elements in Teeth Enamel by Inductively Coupled Plasma Mass Spectroscopy
WANG Song-jun1,CHANG Ping1,SU Wei-na1,SUN Chun-hua1,WANG Li-juan1,ZHANG Gui-ying2,HOU Tian-ping3
1. Center of Test Science,Jilin University, Changchun 130026, China 2. College of Public Health,Jilin University, Changchun 130021, China 3. Jilin Teachers Institute of Engineering and Technology, Changchun 130052, China
Abstract:Determination of rare earth and trace elements in teeth enamel was studied by inductively coupled plasma mass spectroscopy. The sample was completely dissolved by HNO3-H2O2. The range of precision is 0.50%-6.79% given by the relative standard deviation and the range of accuracy is 94.45%-104.78% obtained by measuring standard recovery rate. The above data satisfied completely the requirements of biological sample. While this study provides experimental data of rareearth and trace elements of teeth enamel for carious teeth and none carious teeth of human for prevention and cure of decayed tooth. A reasonable method is provided for mouth medicine research.
Key words:ICP-MS;Teeth enamel;Rareearth and trace elements
[1] Chew L T et al. Applied Radiation and Isotopes,2000,53(4-5):633. [2] YANG Xiao-feng et al(杨晓峰等). Rare Earth(稀土),2000,21(5):23. [3] Anerjee A et al. Caries Research,2000,34(2):144. [4] WANG Song-jun, CHANG Ping(王松君,常 平). Spectroscopy and Spectral Analysis(光谱学与光谱分析),1992,12(1):79. [5] WANG Rui-ru et al(王瑞儒等). China Mouth Medicine Magazine(中华口腔医学杂志),1995,30(2):107. [6] ZHAO Ling-xia et al(赵凌霞等). Humanics Transaction(人类学学报),1999,19(2):102.