Research on Gas Pressure Measurement Method Based on Absorption Spectroscopy and Laser Interference Technology |
ZHANG Bo-han, YANG Jun, HUANG Qian-kun, XIE Xing-juan |
AVIC Changcheng Institute of Metrology & Measurement, Beijing 100095, China
|
|
|
Abstract Optical non-contact measurement of gas pressure is currently one of the important application fields of laser technology,and the temperature coupling problem in the process of gas pressure measurement is a difficult point in the current research work.Therefore, a combined measurement method of spectral and laser interference technology is proposed, which realizes the calculation of gas pressure value and temperature value by integrating absorbance and refractive index. The principle of direct absorption measurement of tunable semiconductor laser spectroscopy (TDLAS) and the principle of laser interferometry based on the refractive index are analyzed, and the pressure measurement model based on absorption spectrum and the laser interferometric pressure measurement model based on the refractive index are established. The method of fitting the intensity function of the absorption spectrum by the cubic polynomial equation establishes a mathematical model for the solution of the gas pressure value and temperature value based on the integral absorbance and refractive index. In the experiment, a gas pressure detection system based on TDLAS technology and laser interference technology was built. A tunable semiconductor laser with a center wavelength of 2 004 nm and a laser interferometer with a fixed wavelength of 632.8 nm wasused. The length of the gas cell was 24.8 cm. CO2 was selected forthe research. Use the measurement results of the high-precision pressure controller and temperature sensor as the pressure and temperature reference values, and use the vacuum signal as the background signal to measure and calculate the integrated absorbance and refractive index values after the gas pressure changes in a room temperature environment. Finally, the gas pressure and temperature values are obtained by the solution. The realization results show that the maximum relative error of the pressure measurement result is 3.61%, the minimum relative error is 0.5%, and the average relative error is 1.99%. Under the premise of temperature in Kelvin, the maximum absolute error of the temperature solution is 7.66 K, the absolute minimum error is 0.78 K, and the average absolute error of the measurement is 3.29 K. The measurement results are in good agreement with the reference results. This work can provide a reference for future analysis and research on the influence of optical methods on gas pressure and temperature.
|
Received: 2021-10-14
Accepted: 2022-04-23
|
|
|
[1] Felix Sharipov, Yuanchao Yang, Jacob E Ricker, et al. Metrologia, 2016, 53(5): 1177.
[2] Ahmed Z, Klimov N N, Douglass K, et al. Research Gate, 2016, 3: 07690.
[3] XU Yu-rong, LIU Yang-yang, WANG Jin, et al(许玉蓉, 刘洋洋, 王 进, 等). Acta Physica Sinica(物理学报), 2020, 69(15): 150601.
[4] ZANG Yi-peng, NIE Wei, XU Zhen-yu, et al(臧益鹏, 聂 伟, 许振宇, 等). Acta Optica Sinica(光学学报), 2018,38(11):1130004.
[5] YU Ying-chun, WANG Han, WANG Zhi-min, et al(禹迎春,王 晗,王志敏,等). Journal of Applied Optics(应用光学) , 2020, 41(6): 1255.
|
[1] |
ZENG Si-xian1, REN Xin1, HE Hao-xuan1, NIE Wei1, 2*. Influence Analysis of Spectral Line-Shape Models on Spectral Diagnoses Under High-Temperature Conditions[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(09): 2715-2721. |
[2] |
PENG Wei, YANG Sheng-wei, HE Tian-bo, YU Ben-li, LI Jin-song, CHENG Zhen-biao, ZHOU Sheng*, JIANG Tong-tong*. Detection of Water Vapor Concentration in Sealed Medicine Bottles Based on Digital Quadrature Phase-Locked Demodulation Algorithm and TDLAS
Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 698-704. |
[3] |
ZHANG Le-wen1, 2, WANG Qian-jin1, 3, SUN Peng-shuai1, PANG Tao1, WU Bian1, XIA Hua1, ZHANG Zhi-rong1, 3, 4, 5*. Analysis of Interference Factors and Study of Temperature Correction Method in Gas Detection by Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 767-773. |
[4] |
LONG Jiang-xiong1, 2, ZHANG Yu-jun1*, SHAO Li1*, YE Qing1, 2, HE Ying3, YOU Kun3, SUN Xiao-quan1, 2. Traceable Measurement of Optical Path Length of Gas Cell Based on Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(11): 3461-3466. |
[5] |
LI Cong-cong1, LUO Qi-wu2, ZHANG Ying-ying1, 3*. Determination of Net Photosynthetic Rate of Plants Based on
Environmental Compensation Model[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1561-1566. |
[6] |
CHEN Hao1, 2, JU Yu3,HAN Li1. Research on the Relationship Between Modulation Depth and Center of High Order Harmonic in TDLAS Wavelength Modulation Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3676-3681. |
[7] |
CHEN Yang, DAI Jing-min*, WANG Zhen-tao, YANG Zong-ju. A Near-Infrared TDLAS Online Detection Device for Dissolved Gas in Transformer Oil[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3712-3716. |
[8] |
WANG Guo-shui1, GUO Ao2, LIU Xiao-nan1, FENG Lei1, CHANG Peng-hao1, ZHANG Li-ming1, LIU Long1, YANG Xiao-tao1*. Simulation and Influencing Factors Analysis of Gas Detection System Based on TDLAS Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(10): 3262-3268. |
[9] |
CHEN Hao1, 2, JU Yu3, HAN Li1, CHANG Yang3. Curve Fitting of TDLAS Gas Concentration Calibration Based on Relative Error Least Square Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(05): 1580-1585. |
[10] |
HUANG An1, 2, XU Zhen-yu1, XIA Hui-hui1, YAO Lu1, RUAN Jun1, HU Jia-yi1, ZANG Yi-peng1, 2, KAN Rui-feng1*. Measurement Method of Two-Dimensional Distribution of Temperature and Components in Gas Turbine Combustor Based on Wavelength Modulated Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(04): 1144-1150. |
[11] |
JU Yu1,CHEN Hao2, 3,HAN Li2,CHANG Yang1, ZHANG Xue-jian1. Based on TDLAS Technology Gas Concentration Calibration Algorithm for a Large Range[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(12): 3665-3669. |
[12] |
CHEN Hao1, 2, JU Yu3, HAN Li1, CHANG Yang3. Algorithms for Calculating the Concentration of Gas Mixture Containting Different Background Gases in TDLAS Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(10): 3015-3020. |
[13] |
XIAO Hu-ying1, YANG Fan1, XIANG Liu1, HU Xue-jiao2*. Jet Vacuum Enhanced Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(10): 2993-2997. |
[14] |
ZHANG Bu-qiang1, 2, XU Zhen-yu1, LIU Jian-guo1, XIA Hui-hui1, FAN Xue-li1, NIE Wei1, 2, YUAN Feng1, 2, KAN Rui-feng1. Modulation Characteristics of Laser Based on Wavelength Modulation Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(09): 2702-2707. |
[15] |
LI Zheng-hui1,3, YAO Shun-chun1,3*, LU Wei-ye2, ZHU Xiao-rui1,3, ZOU Li-chang1,3, LI Yue-sheng2, LU Zhi-min1,3. Study on Temperature Correction Method of CO2 Measurement by TDLAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(07): 2048-2053. |
|
|
|
|