Evaluation of Flow Injection-Solution Cathode Glow Discharge With an Interference Filter Wheel for Spectral Discrimination |
ZHENG Pei-chao, ZHONG Chao, WANG Jin-mei*, LUO Yuan-jiang, LAI Chun-hong, WANG Xiao-fa, MAO Xue-feng |
Chongqing University of Posts and Telecommunications, Chongqing 400065, China |
|
|
Abstract Solution cathode glow discharge-atomic emission spectroscopy is a novel technology for metal ion detection in the aqueous solution that has emerged in recent years. It has the remarkable characteristics of rapid, real-time and low-cost detection. In this paper, an industrial injection pump was used to realize the flow injection analysis for the solution cathode glow discharge excitation source, and the narrow-band filter was used to extract the metal element spectral signals of Na, K, Ca, Li, Sr, and Cs, and the photomultiplier and ammeter were employed to obtain optical signals to electric signals, to realize the detection of metal elements in aqueous solution. The effects of injection volumes of 100 and 166 μL on the signal intensity of 1 mg·L-1 Na element was analyzed. The relative standard deviations (RSD) of the signal intensity are 4.64% and 1.95%, respectively, indicating that both injections had good repeatability. In order to obtain better analysis performance, the influence of parameters such as DC discharge voltage, slit width and photomultiplier tube supply voltage on signal strength were explored. The experimental results show that a high signal-to-back ratio is obtained when the DC discharge voltage is 1 000 V, the slit width is 70 μm, and the photomultiplier tube supply voltage is -800 V. the detection limits of six metal elements Na, K, Ca, Li, Sr and Cs were measured when the equipment works at flow injection mode, which was 2.78, 4.23, 589, 9.45, 981 and 83.6 μg·L-1. Na and K elements in the mixed solution were quantitatively analyzed and measured, the measurement errors were 7.5% and 6.67%, and the precisions were 1.24% and 0.89%, respectively, indicating that the homemade equipment can detect metal elements with high accuracy.
|
Received: 2020-01-22
Accepted: 2020-05-09
|
|
Corresponding Authors:
WANG Jin-mei
E-mail: wangjm@cqupt.edu.cn
|
|
[1] Brian P, Amir Liba A, Nelson J. Journal of Analytical Atomic Spectrometry, 2015, 30: 1179.
[2] Bocca B, Caimi S, Senofonte O, et al. Science of The Total Environment, 2018, 630(14): 922.
[3] Etorki A. International Journal of Environmental Analytical Chemistry, 2016, 3(2): 2381.
[4] Zou H J, Wang Z, Li Q, et al. Chinese Journal of Analytical Chemistry, 2017, 45(7): 973.
[5] Wang Z, Schwartz A J, Ray S J, et al. Journal of Analytical Atomic Spectrometry, 2013, 28(2): 234.
[6] Wang Z, Gai R, Zhou L, et al. Journal of Analytical Atomic Spectrometry, 2014, 29(11): 2042.
[7] Doroski T A, King A M, Fritz M P, et al. Journal of Analytical Atomic Spectrometry, 2013, 28(7): 1090.
[8] Wang J M, Tang P F, Zheng P C, et al. Journal of Analytical Atomic Spectrometry, 2017, 32(10): 1925.
[9] Peng X, Guo X, Ge F, et al. Journal of Analytical Atomic Spectrometry, 2019, 34(2): 394.
[10] Schwartz A J, Ray S J, Hieftje G M. Journal of Analytical Atomic Spectrometry, 2016, 31(6): 1278.
[11] Zheng P C, Zhai X, Wang J M, et al. Analytical Letters, 2018, 51(14): 2304.
[12] Tsizin G I, Statkus M A, Zolotov Y A, et al. Journal of Analytical Chemistry, 2015, 70(11): 1289. |
[1] |
YUAN Liang-jing1, JIA Yun-hai1, 2, CHENG Da-wei1, 2*. Study on Methods of Detection Limit in XRF Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(02): 412-418. |
[2] |
ZHANG Shuai-shuai1, GUO Jun-hua1, LIU Hua-dong1, ZHANG Ying-li1, XIAO Xiang-guo2, LIANG Hai-feng1*. Design of Subwavelength Narrow Band Notch Filter Based on
Depth Learning[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1393-1399. |
[3] |
ZHENG Pei-chao, LIU Ran-ning, WANG Jin-mei, FENG Chu-hui, HE Yu-tong, WU Mei-ni, HE Yu-xin. Solution Cathode Glow Discharge-Atomic Emission Spectroscopy Coupled With Hydride Generation for Detecting Trace Mercury and Tin in Water[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(04): 1139-1143. |
[4] |
ZHENG Pei-chao, LUO Yuan-jiang, WANG Jin-mei*, HU Qiang, YANG Yang, MAO Xue-feng, LAI Chun-hong, FENG Chu-hui, HE Yu-tong. Determination of Strontium in Strontium-Rich Mineral Water Using Solution Cathode Glow Discharge-Atomic Emission Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(01): 272-276. |
[5] |
LI Peng-cheng1, 2, LIU Han1, 2, ZHAO Long-lian1, 2, LI Jun-hui1, 2*. Key Parameters for Maize Leaf Moisture Measurement Using NIR Camera With Filters Based on Hyperspectral Data[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(10): 3184-3188. |
[6] |
CAI Shun-yan1, 2, ZHOU Jian-bin1*, TUO Xian-guo1, YU Jie1. Optimized Filter Selection for Measuring Copper and Molybdenum Contents by EDXRF[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(06): 1934-1939. |
[7] |
LI Ling, WU Zhi-feng, WANG Yan-fei, DAI Cai-hong. The Effects of Light Source Differences on Array Spectroradiometer Measurements and the Research of Ultraviolet Stray Light Correction[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(03): 837-841. |
[8] |
ZHANG Ying-hua1, 2, 3, LI Ang1*, XIE Pin-hua1, HUANG Ye-yuan1, HU Zhao-kun1, ZHANG Chao-gang1. Ultraviolet Two-Dimensional Non-Dispersive Imaging of SO2 Column Density in Power Plant Plume[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(03): 688-693. |
[9] |
LI Hong-bo1, 2, HU Bing-liang1*, YU Lu1, 2, WEI Rui-yi1, YU Tao1. A Fast Lossless Data Compression Method for the Wedge Filter Spectral Imager[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(01): 297-302. |
[10] |
XU Song-ning, LI Chuan-xiang, NING Ri-bo, LI Qian, JIANG Ran. Spectral Characteristics of Pb Plasma in Soil at Low Pressure[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(09): 2877-2880. |
[11] |
ZHENG Pei-chao, TANG Peng-fei, WANG Jin-mei*, LI Shi-yu. Analysis of Metal Elements Manganese Using Solution Cathode Glow Discharge-Atomic Emission Spectrometry with Portable Spectrographs[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(05): 1567-1571. |
[12] |
WANG Xia1, LIAO Ning-fang1, LI Ya-sheng1, CHENG Hao-bo1, 2, CAO Bin1, YANG Wen-ming1, LIN Kai1. The Research on Reconstruction of Spectral Reflectance in LCTF Imaging System Based on Comparative Measurement[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(01): 290-295. |
[13] |
GONG Xiao-long1, TIAN Hong-xiang1*, SUN Yun-ling1, HE Wei1, LI Jing1, YANG Kun2. Research on Fuel-Dilution Monitoring of Engine Lubricant by UV Fluorescence[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(12): 3758-3762. |
[14] |
ZHENG Pei-chao, TANG Peng-fei, WANG Jin-mei*, YANG Rui. Quantitative Analysis of Solution Cathode Glow Discharge-Atomic Emission Spectroscopy Coupled with Internal Standard Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(09): 2900-2904. |
[15] |
YANG Fan-chao1, 2, LI Yong1, HU Bing-liang1, KONG Liang1 , WEI Ru-yi1, LI Hong-bo1, 2. Design of Hyperspectral Polarimetic Imaging System Based on LCTF and LCVR[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(08): 2610-2614. |
|
|
|
|