Apparatus, Method and Result Analysis of International Key Comparison for Spectral Irradiance |
DAI Cai-hong, WANG Yan-fei, LI Ling, WU Zhi-feng*, XIE Yi-hang |
Division of Optical Metrology, National Institute of Metrology, China, Beijing 100029, China
|
|
|
Abstract Based on a high-temperature blackbody and a double grating spectral comparison measurement system, the new primary standard apparatus of spectral irradiance was developed at the National Institute of Metrology (NIM) in the spectral wavelength from 230 to 2 550 nm. Stable 1 000 W tungsten halogen lamps are the secondary primary standard for spectral irradiance values. The temperature measurement of the high-temperature blackbodies was traced to the Pt-C and Re-C fixed-point blackbodies and checked against the WC-C fixed-point blackbody at 3 021 K. From 2017 to 2019, NIM participated in the new international comparison of spectral irradiance CCPR-K1 using the newly developed primary standard apparatus. An organized by BIPM. The comparison was piloted by VNIIOFI from Russia, with 12 laboratories participating. The comparison results show that, except for a few laboratories, the consistency of the participated laboratories is ±2.0%, ±1.0%, and ±2.0% at wavelengths of 250 to 290 nm, 300 to 1 300 nm, and 1 500 to 2 500 nm. The average relative deviation between NIM and the KCRV(key comparison reference values)is 0.13%, 0.28% and 0.15%, respectively, in the ultraviolet, visible, and near-infrared wavelengths. The whole wavelength from 250 to 2 500 nm, the average relative deviation between NIM and the KCRV is 0.17%. Compared with the average relative deviation of 0.9% in 2004, the measurement capability of spectral irradiancehas significantly improved.
|
Received: 2024-03-18
Accepted: 2024-07-10
|
|
Corresponding Authors:
WU Zhi-feng
E-mail: wuzf@nim.ac.cn
|
|
[1] DAI Cai-hong, WU Zhi-feng, OUYANG Hui-quan, et al(代彩红,吴志峰,欧阳慧泉,等). Acta Metrologica Sinica(计量学报), 2013, 34(3):201.
[2] Dai Caihong, Wu Zhifeng, Yu Jialin. SPIE. 8910: 89100D-1~8, International Symposium on Photoelectronic Detection and Imaging 2013. Beijing, China.
[3] Suzuki M,Ooba N. Metrologia, 1976,12(3): 123.
[4] Walker J H, Robert D Saunders, Jackson K, et al. Journal of Research of the National Institute of Standards and Technology, 1991, 96(6): 647.
[5] Woolliams Emma R, Fox Nigel P, Cox Maurice G, et al. Metrologia, 2006,43(2):S98.
[6] Woolliams E R, Cox M G, Fox N P, et al. Final Report of the CCPR K1-a Key Comparison of Spectral Irradiance 250 nm to 2500 nm. 2006.
[7] Dai Caihong, Boris Khlevnoy, Wu Zhifeng, et al. MAPAN-Journal of Metrology Society of India, 2017, 32(3): 243.
[8] Dai Caihong, Wu Zhifeng, Wang Yanfei, et al. Applied Optics, 2020, 59(27): 8494.
[9] Dai Caihong,Wang Yanfei, Li Ling, et al. Metrologia, 2022, 59(2): 024001.
[10] Khlevnoy Boris, Solodilov Maksim, Kolesnikova Svetlana, et al. Metrologia, 2023, 60(1A): 02002.
|
[1] |
DAI Cai-hong, WANG Yan-fei*, LI Ling, WU Zhi-feng, XIE Yi-hang. A New Method for Realization National Primary Scale of Spectral
Irradiance Based on Large Area High Temperature Fixed Point
Blackbody[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(04): 925-931. |
[2] |
XIE Yi-hang, DAI Cai-hong*, WANG Yan-fei, WU Zhi-feng, LI Ling, HE Shu-fang. A New Method for Inflection Point Temperature Calculation of Large-Area High-Temperature Fixed-Point Blackbody Used in Spectral Irradiance Scale Realization[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3942-3948. |
[3] |
WU Zhi-feng, DAI Cai-hong, ZHAO Wei-qiang, XU Nan, LI Ling, WANG Yan-fei, LIN Yan-dong. Spectral Irradiance Responsivity Calibration Using Tunable Lasers[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(03): 853-857. |
[4] |
DAI Cai-hong, WU Zhi-feng, WANG Yan-fei, LI Ling. Realization of National Primary Standard Apparatus of Spectral Irradiance from 200 to 400 nm[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(08): 2640-2644. |
[5] |
DONG Hai-long, WANG Jia-chun*, ZHAO Da-peng, CHEN Zong-sheng, LIU Rui-huang, SHI Jia-ming. Analysis of Measurement Uncertainty in THz-TDS Carried by Delay-Line Position Deviation[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(11): 3379-3384. |
[6] |
LI Zhi-gang, LI Fu-tian*. A New Generation High-Temperature Blackbody and Synchrotron Radiation Facility and Its Application in Extraterrestrial Solar Spectral Irradiance Measurements[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(10): 3242-3248. |
[7] |
GAO Rui-feng1, 2, GAO Meng-chao3, LING Rui3, ZHANG Ning1* . Improving Analytical Methods by Uncertainty Evaluation with the Case of Determination of Aluminum in Starch Products by ICP-MS [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(04): 1211-1216. |
[8] |
SUN Kun1, 2, SUN Xiao-gang3, YU Xiao-yang1, SUN Xiao-ming1. Development of Multi-Spectral Thermometer for Explosion Flame True Temperature Measurement——Field Experiments and Measurement Accuracy Analysis[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(06): 1719-1722. |
|
|
|
|