High Precision Tunable Laser Absorption Spectroscopy Temperature
Detection Method Based on Transmission Curve Fitting |
GAO Hui, YAO Shu-zhi, ZHANG Meng, JIANG Meng, ZHANG Zi-hao, WANG Xue-feng, YANG Yong* |
Beijing Institute of Aerospace Control Devices, Beijing 100089, China
|
|
|
Abstract The tunable diode laser absorption spectroscopy technology is widely used in flow field detection. However, the harsh environment of on-site application puts higher requirements on the test system's environmental adaptability and anti-noise interference ability. Direct absorption spectroscopy technology can obtain all the information on the absorption spectrum and is suitable for many types of test environments, such as wind tunnels and engine flow fields. However, due to the influence of a harsh on-site test environment, the signal-to-noise ratio of the obtained signals is usually poor, and it isn't easy to achieve high-precision detection. This article analyzes the data processing and analysis algorithm of direct absorption spectroscopy technology and proposes a high-precision temperature analysis method based on transmission curve fitting for this problem. A temperature detection system based on direct absorption spectroscopy technology was developed using the absorption spectra of water vapor near 7 185.6 and 7 444.3 cm-1. The processing and analysis of test signals in the temperature range of 573~1 173 K were carried out. The analysis results of the baseline-absorption line fitting algorithm(BALF) and the transmission curve fitting(TCF) algorithm were compared. When the detection temperature was 773 K, compared with the BALF algorithm, the peak-to-peak fitting error near 7 444.3 cm-1 of the TCF algorithm can be reduced by about 30%, and the peak-to-peak fitting error near 7 185.6 cm-1 can be reduced by about 16%. In the temperature range of 573~1 173 K, the maximum error of the TCF algorithm for temperature inversion is 13 K, which is reduced by 23 K compared with the BALF algorithm. When random noise signals with different amplitudes were added to the test signals, the standard deviation of test results using these two algorithms increased with noise amplitude. Moreover, the higher the temperature, the greater the standard deviation of test results. When the peak-to-peak noise value was 20, 60, and 100 mV respectively, for different temperatures, the minimum standard deviation of test results using the BALF algorithm was 18 K, and the maximum was 313 K. The minimum standard deviation of test results using the TCF algorithm is 4 K, and the maximum is 44 K. Experimental analysis results show that compared with the BALF algorithm, the TCF algorithm can correct the baseline fitting error of laser power, improve the fitting accuracy of transmission signals, and reduce temperature analysis errors. Comparing the temperature analysis results of test signals under different noise levels shows that the TCF algorithm can achieve higher detection accuracy and precision under noise interference and has stronger anti-noise interference ability.
|
Received: 2023-10-31
Accepted: 2024-03-21
|
|
Corresponding Authors:
YANG Yong
E-mail: youngbrave@qq.com
|
|
[1] Zhao W, Xu L, Huang A, et al. IEEE Sensors Journal, 2020, 20(8): 4179.
[2] Wang Zhenzhen, Zhou Wangzheng, Yan Junjie, et al. Measurement Science and Technology, 2020, 31(3): 035203.
[3] Huang A, Cao Z, Zhao W, et al. IEEE Transactions on Instrumentation and Measurement, 2020, 69(11): 9087.
[4] Wang Z, Zhou W, Kamimoto T, et al. Applied Spectroscopy, 2020, 74(2): 210.
[5] ZHOU Wang-zheng, WANG Zhen-zhen, YAN Jun-jie, et al(周王峥, 王珍珍, 严俊杰,等). Journal of Propulsion Technology(推进技术), 2021, 42(9): 2129.
[6] Jang H, Choi D. Applied Sciences, 2020, 10(1): 285.
[7] Qiu S, Cao Z, Wen J, et al. IEEE Transactions on Instrumentation and Measurement, 2023, 72: 9506614. doi: 10.1109/TIM.2023.3271738.
[8] Zhai Y, Wang F, Chi Y, et al. Optical Engineering, 2022, 61(3): 034103.
[9] Liu C, Xu L, Chen J, et al. Opt. Express, 2015, 23(17): 22494.
[10] Zhang H, Cao Z, Zhao W, et al. IEEE Transactions on Instrumentation and Measurement, 2020, 69(10): 8226.
[11] HU Bing, DANG Feng, WANG Bao-lin(胡 兵,党 峰,王宝林). Navigation and Control(导航与控制), 2020, 19(3): 95.
[12] Huang A, Cao Z, Wang C, et al. IEEE Transactions on Instrumentation and Measurement, 2021, 72:4506911. doi: 10.1109/TIM.2021.3115210.
[13] Li R, Li F, Lin X, et al. Microwave and Optical Technology Letters, 2023, 65(5): 1024.
[14] Ji Y, Duan K, Lu Z, et al. Sensors and Actuators B: Chemical, 2022, 371: 132574.
[15] Wang Z, Deguchi Y, Kamimoto T, et al. Fuel, 2020, 268: 117370.
[16] Xia H, Kan R, Xu Z, et al. Optics and Lasers in Engineering, 2017, 90: 10.
[17] Song J, Hong Y, Wang G, et al. Applied Physics B, 2013, 112(4): 529.
[18] Li J, Schwarm K K, Wei C, et al. Measurement Science and Technology, 2021, 32(4): 045502.
[19] Liu X, Ma Y. Sensors, 2022, 22(16): 6095.
[20] Lee J, Bong C, Yoo J, et al. Optics Express, 2020, 28(14): 21121.
[21] Gong T, Shi W, Zhu H, et al. Microwave and Optical Technology Letters, 2024, 66(1): e33747.
|
[1] |
ZHANG Xue-jun1, CHEN Qin-gen2, YANG Zhan1, DENG Qin1, HE Shuan-ling3, PENG Zhi-min3*. On Line Simultaneous Measurement of CO/CO2/H2S Concentration Based on Laser Absorption Spectrum[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(05): 1412-1416. |
[2] |
TIAN Si-di1, WANG Zhen1, DU Yan-jun2, DING Yan-jun1, PENG Zhi-min1*. High Precision Measurement of Spectroscopic Parameters of CO at 2.3 μm Based on Wavelength Modulation-Direct Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(07): 2246-2251. |
[3] |
ZHANG Li-fang1, YANG Yan-xia1, ZHAO Guan-jia1, MA Su-xia1, GUO Xue-mao2. Comparison of Numerical Iterative Algorithms for Two-Dimensional Absorption Spectral Reconstruction[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(05): 1367-1375. |
[4] |
PAN Sun-qiang, HU Peng-bing, CHEN Zhe-min, ZHANG Jian-feng, LIU Su-mei. Measurement of Vapor Hydrogen Peroxide Based on Mid Infrared Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(04): 1102-1106. |
[5] |
CHEN Jia-jin1, 2, WANG Gui-shi1*, LIU Kun1, TAN Tu1, CHENG Gang1, TIAN Xing1, GAO Xiao-ming1, 2*. High Sensitivity Detection of Carbon Dioxide Based on Portable Cylindrical Multi-Pass Cell[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(01): 292-296. |
[6] |
LIU Yong-sheng, HE Jian-jun*, ZHU Gao-feng, YANG Chun-hua, GUI Wei-hua. A New Method for Second Harmonic Baseline Correction and Noise Elimination on Residual Oxygen Detection in Packaged Xilin Bottle[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(08): 2598-2602. |
[7] |
FANG Zheng, OUYANG Qi-nan, ZENG Fu-rong, CHEN Si-yuan, MA Sheng-lin* . Study on the Estimation Algorithm of the Temperature Based on Mid-Wave Infrared Remote Sensing[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(04): 960-966. |
[8] |
XIA Hui-hui, LIU Jian-guo, XU Zhen-yu, KAN Rui-feng, HE Ya-bai, ZHANG Guang-le, CHEN Jiu-ying . Reconstruction Research for Gas Concentration and Temperature of Flame Based on Algebraic Reconstruction Technique[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(10): 2697-2702. |
[9] |
ZHANG Li-fang, WANG Fei*, YU Li-bin, YAN Jian-hua, CEN Ke-fa . The Research for Trace Ammonia Escape Monitoring System Based on Tunable Diode Laser Absorption Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(06): 1639-1642. |
[10] |
WANG He-lin1,2, ZHANG Yu1,2*, LIU Wen-yan1, WANG Guo-guang2, ZHANG Tie-qiang2 . Study of PL Spectra of PbSe Quantum Dots for IC Chip Temperature Dependence[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(05): 1169-1172. |
[11] |
SHAO Xin1, 2, LIU Fu-gui1, DU Zhen-hui3*, WANG Wei3 . Absorption Spectrum of Quasi-Continuous Laser Modulation Demodulation Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34(05): 1186-1190. |
[12] |
WANG Gao, LIU Shao-cong, WEN Qiang, ZHAO Hui, ZHAO Yu . Research on Granary Temperature Network Monitoring System Based on the Linear Frequency Shift of Spectrum [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(04): 1146-1150. |
[13] |
JIANG Zhi-shen, WANG Fei*, XING Da-wei, XU Ting, YAN Jian-hua, CEN Ke-fa. The Reconstruction of Two-Dimensional Distributions of Gas Concentration in the Flat Flame Based on Tunable Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(11): 2891-2896. |
[14] |
ZOU De-bao1,2, CHEN Wen-liang1, 2*, DU Zhen-hui1, JIA Hao1, QI Ru-bin1, LI Hong-lian1,3, ZHEN Yang1, HOU Yan-xia1, XU Ke-xin1. Selection of Digital Filtering in the Escaping Ammonia Monitoring with TDLAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(09): 2322-2326. |
[15] |
DU Zhen-hui, GAO Dong-yu, QI Ru-bin, XU Xiao-bin, JIAO Meng. Comparison of Quasi-Continuous and Continuous Tunable Diode Laser Absorption Spectroscopy for Gas Detection [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(06): 1580-1583. |
|
|
|
|