光谱学与光谱分析 |
|
|
|
|
|
Research on the Method of Transient Spectrum Detection Based on Array CCD |
ZHU Bing-li1, 3, BAI Yong-lin2*, WANG Bo2, LIU Bai-yu2, OUYANG Xian2,YANG Wen-zheng2, BAI Xiao-hong2, QIN Jun-jun2, ZHAO Jun-ping2, GOU Yong-sheng2, LU Kai1, 3 |
1. State Key Laboratory of Transient Optics and Photonics, Xi’an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi’an 710119, China 2. Key Laboratory of Ultrafast Photoelectric Diagnostic Technology, Xi’an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi’an 710119, China 3. Graduate University of Chinese Academy of Sciences, Beijing 100049, China |
|
|
Abstract Based on the characteristic of high speed line scanning for CCD in transient spectrum detection, a method of transient spectrum detection with array CCD is presented. The high speed line scanning with array CCD was realized by changing the mode of charge transfer. In order to explore the feasibility of this method, a fast detection system of single point based on linear CCD was designed and fabricated. Seven different LED pulses were measured when the system worked at fast detection mode of single point and normal mode respectively. The results demonstrate that the method of fast detection of single point based on linear CCD is feasible, and the rate of single point detection reaches up to 20 MHz. Thus, in theory, it was proved that transient spectrum detection with array CCD by changing the mode of charge transfer is also feasible.
|
Received: 2011-08-23
Accepted: 2011-11-20
|
|
Corresponding Authors:
BAI Yong-lin
E-mail: baiyonglin@opt.ac.cn
|
|
[1] LI Jing-zhen(李景镇). Science in China Ser. E Technological Sciences(中国科学 E辑 技术科学),2009,39(12): 1887. [2] SU Hong-yu, ZHOU Yan, YU De-ping, et al(苏红雨,周 彦,余德萍,等). China Measurement Technology(中国测试技术),2005,31(4):5. [3] DENG Xing-kai,YANG Yong-jun(邓兴凯,杨永军). Metrology & Measurement Technology(计测技术),2011,31(1):45. [4] SHI Bao-song, SUN Shou-hong, ZHANG Wei(石宝松,孙守红,张 伟). Electronic Measurement Technology(电子测量技术),2010,33(11):84. [5] WANG Bo, BAI Yong-lin, OUYANG Xian, et al(王 博,白永林,欧阳娴,等). Acta Photonica Sinaca(光子学报), 2010,39(3):441. [6] Reich R K, Rathman D D, O’Mara D M, et al. 27th International Congress on High-Speed Photography and Photonics, 2007, 6279: 1K1. [7] Reich R K, Rathman D D, O’Mara D M, et al. Review of Scientific Instruments, 2003, 74(3): 2027. [8] Kosonocky W F, Yang G, Kabra R K, et al. IEEE Transactions on Electron Devices, 1997, 44(10): 1617. [9] Goji T Etoh, Poggemann Dirk, Kreider Greg, et al. IEEE Transactions on Electron Devices, 2003, 50(1): 144. |
[1] |
WU Jin-hui1,2, YANG Rui-feng1,2, WANG Gao1,2 . Research on Transient Temperature Detection by Spectrum Function Variation of Speckle Pattern Interferometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(07): 1852-1855. |
[2] |
ZHENG Jin-kun1,3, BAI Yong-lin2*, WANG Bo2, LIU Bai-yu2, YANG Wen-zheng2, BAI Xiao-hong2, QIN Jun-jun2, ZHAO Jun-ping1, GOU Yong-sheng1, LIU Hui1,3 . Real-Time Measurement of Detonation Transient Temperature [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(11): 3060-3063. |
[3] |
YAN Xiao-yan, QIN Jian-min, WU Jin-hui, QIAO Ji-ping . Based on Moire Interference Accurate Detection of the Laser Center Wavelength [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(07): 1873-1877. |
[4] |
DAI Bing1,YUAN Yin-nan2,BAO Zhi-hua1,MEI De-qing3 . An Improved Method for Inversion of Particle Size Distribution from Scattering Spectrum [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(02): 539-542. |
[5] |
DU Li-li1, 2, YI Wei-ning1, 2, ZHANG Dong-ying1, 2, HUANG Hong-lian1, 2, QIAO Yan-li1, 2, ZHANG Xie3 . Radiometric Calibration of LCTF-Based Multispectral Area CCD Camera [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(01): 272-276. |
[6] |
WANG Chun-hong1, 2, 3,XIANG Yang1* . MTF of Line-Array CCD on Detection Performance for Hyperspectral Imager[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(06): 1478-1481. |
|
|
|
|