Dark Current Analysis and Processing Method of Environment
Monitoring Instrument Nadir and Limb |
LIN Fang1, 2, LIU Wen-qing1, 2*, WANG Yu3, CHANG Zhen2, ZHANG Quan2, SI Fu-qi2 |
1. School of Environmental Science and Optoelectronic Technology, University of Science and Technology of China, Hefei 230026, China
2. Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230031, China
3. Institutes of Physical Science and Information Technology, Anhui University, Hefei 230039, China
|
|
|
Abstract EMI-NL satellite-borne imaging spectrometer utilizing charge-coupled device CCD275 is applied to monitor the global trace gases of the atmosphere. This spectrometer is the first application of CCD275 in China. Since its overall volume and power consumption are more significant than usual CCD detectors, CCD275 uses the thermal design of radiant coolers with heaters. The satellite-borne imaging spectrometer works in a sun-synchronous orbit. It captures original spectrum images during the day and dark background images at night. The original spectrum and dark background images must be captured under the same temperature and exposure time conditions. During the data processing, the dark background images are subtracted from the original spectrum images to obtain valid spectral images for subsequent retrieval of the trace gases. During the on-orbit operation of the radiant cooler, there is a risk of cooling efficiency decrease or complete failure. The risk leads to the temperature control failure of the CCD275, and collecting original spectrum images and dark background images at the same temperature becomes a problem. This paper proposes a method to calculate the dark background images instead of acquiring one to solve the problem of dark background collection at a specific temperature after thermal control failure. This method could calculate the dark background images at any specific temperature and exposure time by collecting two sets of dark background images with known temperatures and different exposure times. This paper discusses the working principle, the thermal design, and the dark background collection method of the satellite-borne spectrometer. The relationship formula, which the method mentioned above is based on, between pixel value, temperature, and exposure time is deduced from the analysis of dark current characteristics and the continuous frame transfer imaging mode. To verify this method, this paper built an experimental platform that included a vacuum system, cooling system, and imaging circuit. The platform can set the detector's working temperature and exposure time, obtaining dark background images. Statistic results show that the correlation coefficient between the calculated and measured images reaches 99.96% for a single point, and the pixel value deviation range is 0.06%~6.9%. For a full-frame image under 278 K, the mean error between the calculated dark background and the measured one is 0.23%, the variance error is -3.20%, the correlation coefficient is 99.94%, the pixel value deviation range is -0.70%~2.37%. The feature point matching rate reaches 97.9%. When temperature control fails, the calculated dark background can replace the measured one for spectral image processing.
|
Received: 2023-12-17
Accepted: 2024-07-22
|
|
Corresponding Authors:
LIU Wen-qing
E-mail: wqliu@aiofm.ac.cn
|
|
[1] Veefkind J P, Aben I, McMullan K, et al. Remote Sensing of Environment, 2012, 120: 70.
[2] WANG Yu, LU Yi-huai, ZHAO Xin, et al(王 煜,陆亦怀,赵 欣,等). Laser & Infrared(激光与红外), 2015, 45(6): 663.
[3] Bodegom E, Widenhorn R, Iordache D A. 2005 International Semiconductor Conference, 2005. IEEE, 2005, 2: 327.
[4] ZHU Jian-bing(朱建炳). Vacuum & Cryogenics(真空与低温), 2010,(4): 187.
[5] Teledyne E2V. CCD275 Datasheet, May. 2018.
[6] Zevenbergen P, Doornink J, Butters T, et al. Thermal Design of the Sentinel 5 Precursor TROPOMI Instrument. 42nd International Conference on Environmental Systems,2012: 3580.
[7] DONG De-ping, ZHANG Yu-lin, LU Yan(董德平,张玉林,陆 燕). Infrared and Laser Engineering(红外与激光工程), 2012, 41(1): 119.
[8] ZHANG Quan, HUANG Shu-hua, ZHAO Xin, et al(张 泉,黄书华,赵 欣,等). Acta Photonica Sinica(光子学报), 2017, 46(3): 0311004.
[9] ZHAO Min-jie, SI Fu-qi, ZHOU Hai-jin, et al(赵敏杰,司福祺,周海金,等). Journal of Atmospheric and Enviroment Optics(大气与环境光学学报), 2019,(1): 66.
[10] European Machine Vision Association. EMVA Standard 1288 Standard, 16 June 2021.
[11] CHANG Zhen, WANG Yu, SI Fu-qi, et al(常 振,王 煜,司福祺,等). Chinese Journal of Lasers(中国激光), 2017, 44(8): 0804002.
[12] SU Jing-ming, ZHAO Min-jie, ZHOU Hai-jin, et al(苏静明,赵敏杰,周海金,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2022, 42(3): 686.
[13] Lowe D G. International Journal of Computer Vision, 2004, 60: 91.
|
[1] |
ZHAO Min-jie, SI Fu-qi*, LU Yi-huai, WANG Yu, WANG Shi-mei, JIANG Yu, ZHOU Hai-jin, LIU Wen-qing . Spectral Calibration of Space-born Imaging Spectrometers Using Spectrum-Matching Technique[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(07): 2049-2053. |
[2] |
LI Ying1,LI Ping1,2*,XIAO Hai-bo1,HU Dong2,3,YUAN Chang-ying2,4. Detonation Temperature Measurement of Epoxypropane Using Instantaneous Spectrum Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2008, 28(03): 490-493. |
|
|
|
|