Abstract:Some certified reference materials (CRMs) were digested and the elemental concentrations were determined by ICP-MS instrument, and the results include not only the trace key elements such as As, Pb, Hg and Cd but also the major components such as P, Ca, Fe, K, Na etc. The digestion method, the recoveries, and the detection limits are discussed to prove the reliability of ICP-MS and EPA200.8 method for the determination of the elements in food and plant samples. The ICP-MS instrument with ORS (octupole reaction system) was applied to the determination of As, Se etc, and the results are compared with those from the normal ICP-MS without ORS. The removal of ArCl and ArAr interference is discussed, and the advantage of ORS is proved by measuring the As and Se in some CRMs.
施燕支1,王英锋1,贺闰娟1,陈登云2 . 应用普通ICP-MS和八极杆碰撞/反应池(ORS)技术测定动植物样品中的As和Se的比较[J]. 光谱学与光谱分析, 2005, 25(06): 955-959.