Research on Gas Sensing System Using Photoacoustic Spectroscopy Based on a Hemispherical Acoustic Resonant Cavity
WANG Gao-xuan1, 2, DING Zhi-hua3*, GAO Xiao-ming4*
1. Ningbo Innovation Center, College of Optical Science and Engineering, Zhejiang University, Ningbo 315100,China
2. School of Information Science and Engineering, Ningbo Tech University, Ningbo 315100,China
3. Ningbo Nuclear Core Optoelectronic Technology Co., Ltd., Ningbo 315100,China
4. Anhui Institute of Optics and Fine Mechanics, Hefei Institutes of Physical Science, Chinese Academy of Sciences, Hefei 230026,China
Abstract:Acoustic resonant cavity is an important element in photoacoustic absorption spectroscopy. The spherical acoustic resonant cavity is widely reported to have the advantages of small volume, high-quality factor, and easier improvement of gas absorption path compared with a traditional cylindrical acoustic resonator. To further reduce the cavity volume and realize the miniaturization of the cavity, this paper first reports a gas sensing system using photoacoustic spectroscopy based on a hemispherical acoustic resonant cavity with a radius of 15 mm and a volume of 7.07 mL. The frequency response of the cavity was characterized to optimize the operating frequency and microphone position through theoretical simulation and experimental verification. To further improve the performance of the photoacoustic spectroscopy gas sensing system, the gas absorption path is further enhanced by combining with a multi-pass cell by increasing the light inlet dimension of the acoustic resonant cavity. The multi-pass cell comprises two mirrors with a diameter of 25.4 mm, a radius of 100 mm, and a distance of 20 cm, which results in a light spot of single line mode on the mirror surface. The photoacoustic signal amplitude of methane and formaldehyde were enhanced by 6 and 4 times, respectively. Using a multi-microphone method, the photoacoustic signal was increased by four times. The developed photoacoustic spectrometer was coupled to two tunable lasers emitting at 1 653 and 3 640 nm, which were irradiated into the cavity in two opposite directions to realize dual-gas detection (methane and formaldehyde) with measurement sensitivity of 2.11×10-6(methane, 20 s) and 0.71×10-6(formaldehyde, 20 s), respectively. The photoacoustic spectroscopy system was calibrated by measuring different methane concentrations, which showed a good linear relationship between photoacoustic signal and methane concentration. Allan's standard deviation method was used to evaluate the instrument stability of the photoacoustic spectroscopy system under long-term operation. The methane detection sensitivity was enhanced to be 0.4×10-6 with an optimal integration time of 660 s. The hemispherical acoustic resonant cavity, due to its good light source applicability and small volume et al., can be employed with good application prospects in the field of gas sensing using photoacoustic spectroscopy.
王高旋,丁志华,高晓明. 基于半球形声学共振腔的光声光谱气体传感系统研究[J]. 光谱学与光谱分析, 2025, 45(03): 608-615.
WANG Gao-xuan, DING Zhi-hua, GAO Xiao-ming. Research on Gas Sensing System Using Photoacoustic Spectroscopy Based on a Hemispherical Acoustic Resonant Cavity. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2025, 45(03): 608-615.
[1] GONG Zheng-feng, WU Guo-jie, XING Jia-wei, et al(宫振峰,吴国杰,幸佳伟,等). Acta Optica Sinica(光学学报),2023, 43(18): 1899911.
[2] Patimisco P, Sampaolo A, Dong L, et al. Applied Physics Reviews, 2018, 5(1): 011106.
[3] DONG Lei, WU Hong-peng, ZHENG Hua-dan, et al(董 磊,武红鹏,郑华丹,等). Chinese Journal of Lasers(中国激光), 2018, 45(9): 0911004.
[4] Wu H, Dong L, Zheng H, et al. Nature Communications, 2017, 8: 15331.
[5] Ma Y. Frontiers in Physics, 2020, 8: 268.
[6] Wang J, Wu H, Sampaolo A, et al. Light: Science & Applications, 2024, 13: 77.
[7] Wei T, Zifarelli A, Dello Russo S, et al. Appl. Phys. Rev., 2021, 8(4): 041409.
[8] Lang Z, Qiao S, Liang T, et al. Optics Express, 2024, 32: 379.
[9] WANG Qi, WANG Shi-chao, LIU Tai-yu, et al(王 琦,王世超,刘泰余,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2022, 42(1): 1.
[10] Fang C, Liang T, Qiao S, et al. Optics Letters, 2024, 49: 770.
[11] Zhang C, Qiao S, He Y, et al. Optics Express, 2024, 32: 848.
[12] Li J S, Chen W D, Yu B L. Applied Spectroscopy Reviews, 2011, 46: 440.
[13] Wang G X, Zhang T, Jiang Y M, et al. Optics Express, 2022, 30(5): 7053.
[14] Qiao S, He Y, Sun H, et al. Light: Science & Applications, 2024, 13: 100.
[15] Rouxel J, Coutard J G, Gidona S, et al. Sensors and Actuators B, 2016, 236: 1104.
[16] Li Z G, Si G S, Ning Z Q, et al. Sensors, 2022, 22: 281.
[17] Huang Q, Zhou S, He T, et al. Microwave and Optical Technology Letters, 2021, 63(10): 2528.
[18] Wei Y, Huang Q, Li J. Sensors & Actuators A: Physical, 2023, 360: 114542.
[19] Yang K, Shen F Y, Chen K, et al. Applied Sciences, 2021, 11: 4997.
[20] Wittstock V, Scholza L, Bierer B, et al. Sensors and Actuators B, 2017, 247: 930.
[21] Miklós A, Hess P. Review of Science Instruments, 2001, 72: 1937.
[22] Yuan W, Yao J, Gong P, et al. Atmosphere, 2023, 14: 704.
[23] Kost B, Baumann B, Germer M, et al. Applied Physics B: Lasers and Optics, 2011, 102: 87.
[24] Gordon I E, Rothman L S, Hargreaves R J, et al. Journal of Quantitative Spectroscopy & Radiative Transfer, 2022, 277: 107949.
[25] Bohren A, von Lerber T, Sigrist M W. Advanced Solid State Lasers, 2000, 34: 229.
[26] Wang G, Meng L, Gou Q, et al. Analytical Chemistry, 2023, 95(6): 3460.