Study on the Influence of Excitation Voltage and Anode on the Contribution of Scattering Effects to the Intensity of X-Ray Fluorescence
HAN Xiao-yuan1, 2,ZHUO Shang-jun1*,SHEN Ru-xiang1,WANG Pei-ling1
1. Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 200050, China 2. Unit 63653 of the Chinese People’s Liberation Army,Xinjiang 841700, China
Abstract:The influence of X-ray tube spectral distribution related to the X-ray tube voltage and target anode on the contribution of scattering effects to the intensity of fluorescence was studied by using some fused disk samples. Three scattering effects were considered. They are coherent scattering effect, incoherent scattering effect and primary fluorescence scattered into the direction of detector respectively. For the fused beads under investigation, the results show that the contributions of coherent scattering effect and the primary fluorescence scattered into the direction of detector decreased with increasing the tube voltage, and the contribution was larger when excited by the X-ray from Cr target than that from Rh target. On the contrary, the contribution of incoherent scattering effect to the intensity of fluorescence increased with increasing the voltage, and was larger when excited by the X-ray from Rh target than that from Cr target. The sum of the contribution of scattering effects to the intensity of fluorescence increased with increasing the voltage, and was larger when excited by the X-ray from Rh target than that from Cr target.
韩小元1, 2,卓尚军1*,申如香1,王佩玲1 . XRF中激发电位和靶材对散射效应增强荧光强度的影响研究[J]. 光谱学与光谱分析, 2007, 27(01): 194-197.
HAN Xiao-yuan1, 2,ZHUO Shang-jun1*,SHEN Ru-xiang1,WANG Pei-ling1 . Study on the Influence of Excitation Voltage and Anode on the Contribution of Scattering Effects to the Intensity of X-Ray Fluorescence. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27(01): 194-197.
[1] Toshio Shiraiwa, Nobukatsu Fujino, Adv. In X-ray Anal., 1969, 12: 446. [2] Mantler M. Anal. Chim. Acta, 1986, 188: 25. [3] Huang T C. X-Ray Spectrometry, 1991, 20: 29. [4] de Boer D K G. X-Ray Spectrometry, 1990, 19: 145. [5] Martin E. Nordberg Ⅲ. X-Ray Spectrometry, 1991, 20: 245. [6] Mantler M. Adv. In X-ray Anal., 1986, 29: 97. [7] Shigeyuki Mori, Mantler M. Adv. In X-ray Anal., 1993, 36: 47. [8] Pella P A, Feng Liangyuan, Small J A. X-Ray Spectrometry, 1985, 14: 125. [9] McMaster W H, Kerr N Del Grande, Mallett J H. Compilation of X-ray Cross Sections, UCRL-50174 Sec.Ⅱ (1969), Lawrence Radiation Laboratory, University of California, Livermore. [10] Han X Y, Zhuo S J, Wang P L, et al. Anal. Chim. Acta,2005, 538: 297.