Improving Analytical Methods by Uncertainty Evaluation with the Case of Determination of Aluminum in Starch Products by ICP-MS |
GAO Rui-feng1, 2, GAO Meng-chao3, LING Rui3, ZHANG Ning1* |
1. Department of Physics, Nanjing Normal University, Nanjing 210023, China2. Nanjing Institute of Product Quality Inspection, Nanjing 210019, China3. Nanjing Institute for Food and Drug Control, Nanjing 210038, China |
|
|
Abstract The measurement uncertainty is a non-negative parameter associated with the result of a measurement that characterizes the dispersion of the quantity values that could reasonably be attributed to the measurand. In the present study measurement uncertainty is estimated using the GUM (ISO/IEC Guide 98: 1993 Guide to the expression of uncertainty in measurement) bottom-up approach. The steps were followed: specifying the measurand; identifying all the associated sources of uncertainty; quantifying the uncertainty components; combining the uncertainty components; determining the extended combined standard uncertainty; reviewing the estimates and reporting the measurement uncertainty. In this process, the major uncertainty components with greater impact were identified; try to eliminate or to reduce the impact of these components can improve measurement methods. Examples were the determination of aluminum in starch and bread crumbs by inductively coupled plasma-mass spectrometry (ICP-MS). The uncertainties of aluminum contents were from measurement repeatability, variability of calibration curve, standard stock solution, dilution, solution volume and sample weighing. The data indicated that the major contributions to the uncertainty budget originating from urel(cAl)1 (the relative standard uncertainty of aluminum content derived from linear least squares calibration), urel(cAl)3 (the relative standard uncertainty of aluminum content derived from the dilution of the standard stock solutions) and urel(rep) (the relative standard uncertainty derived from the repeatability). Based on the analysis of the main individual contributions of each uncertainty source to the total uncertainty value, several modifications were proposed. Firstly helium collision mode was replaced by no gas mode to improve the sensitivity of mass spectrometry. Secondly the number of measurements was increased. Thirdly let the mean of data points in the calibration closer the measurand. Finally the relative error smaller gauges were used. After these modifications, urel(cAl)1, urel(cAl)3 and urel(rep) were from (0.035 8, 0.013 2, 0.008 5) down to (0.006 0, 0.010 5, 0.003 3), respectively; the combined relative standard uncertainty of aluminum was from 0.039 down to 0.013; the expanded uncertainty from 1.8 down to 0.4 mg·kg-1(coverage factor k=2). The improvement effect was significant.
|
Received: 2015-01-13
Accepted: 2015-04-19
|
|
Corresponding Authors:
ZHANG Ning
E-mail: zhangning@njnu.edu.cn
|
|
[1] Evaluation of Measurement Data-Guide to the Expression of Uncertainty in Measurement. JCGM 100, BIPM, IEC, IFCC, ILAC, ISO, IUPAC, IUPAP, OIML, 2008. [2] Ellison S L R, Williams A. Quantifying Uncertainty in Analytical Measurement. EURACHEM/CITAC Guide CG 4, Third Edition, 2012. [3] ISO/IEC 17025, General Requirements for the Competence of Testing and Calibration Laboratories. ISO, Geneva, Switzerland, 2005. [4] Maroto A, Boque R, Riu J, et al. Trends in Analytical Chemistry, 1999, 18: 577. [5] Moser J, Wegscheider W, Meisel T, et al. Analytical and Bioanalytical Chemistry, 2003, 377: 97. [6] Hund E, Massart D L, Smeyers-Verbeke J. Analytica Chimica Acta, 2003, 480: 39. [7] Luo Y, Gao S, Longerich H P, et al. Journal of Analytical Atomic Spectrometry, 2007, 22: 122. [8] XING Li, WANG Mei, ZHAO Feng-min, et al(兴 丽, 王 梅, 赵凤敏, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2014, 34(1): 226. [9] Ekanem E J, Lori J A, Okibe F G, et al. Journal of Food Technology, 2009, 7(2): 50. [10] Bratakos S M, Lazou A E, Bratakos M S, et al. Food Additives and Contaminants: Part B Surveillance, 2012, 5(1): 33. [11] Beauchemin D. Analytical Chemistry, 2010, 82(12): 4786. [12] Llorent-Martinez E J, Fernandez de Cordova M L, Ruiz-Medina A, et al. Microchemical Journal, 2012, 102: 23. [13] Zhu Y B,Chiba K. Talanta, 2012, 90: 57. [14] Chudzinska M, Debska A,Baralkiewicz D. Accreditation and Quality Assurance, 2012, 17(1): 65. [15] Yenisoy-Karakas. Food Chemistry, 2012, 132: 1555. [16] Chen H, Fan C L, Wang Z B, et al. Analytical Methods, 2013, 5: 3291. [17] Tanase I G, Popa D E, Udristioiu G E, et al. RSC Advances, 2015, 5: 11445. |
[1] |
WANG Shuai1, XU Jun-ping1, WANG Nan1, LEI Wan-ying1, FAN Xi-yan1, DOU Sen2, 3*. Structural Characteristics of Mineral-Microbial Residues Formed by Microbial Utilization of Lignin Joined with Fe, Al, Mn-Oxides Based on FT-IR and SEM Techniques[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(07): 2086-2093. |
[2] |
CHANG Zhen1, 2, ZHAO Min-jie1*, WANG Yu1, SI Fu-qi1, ZHOU Hai-jin1, LIU Wen-qing1. Study on the Spectral Features of Space-Borne Aluminum Diffuser[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(04): 997-1000. |
[3] |
ZHANG Chun-yan1,2, LUO Jian-xin2*, OU Li-juan2, LIU Yong2, HU Bo-nian2, YU Gui-peng1*, PAN Chun-yue1*. Luminescence Properties of Carbazole-Substituted Alq3 Derivative: a Potential Multifunctional Fluorescent Sensing Material[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(04): 1290-1297. |
[4] |
HE Qiu-ju1,2, WANG Li-qin1*, ZHANG Ya-xu1. Study of Mechanism of Aluminum Sizing Precipitant on Xuan Paper Based on Spectral Analysis[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(02): 418-423. |
[5] |
ZHOU Dan-yi1, 2, LU Tai-jin2*, KE Jie2, CHEN Hua2, SHI Guang-hai1, LI Ke1. Study on the Spectral Characteristics and the Alexandrite Effect of Diaspore[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(11): 3504-3509. |
[6] |
TANG En-ling1, SONG Ji-qiu1, ZHANG Qing-ming2, HAN Ya-fei1, HE Li-ping1, WANG Meng1, XIANG Sheng-hai1, XIA Jin1, LIU Shu-hua1, GUO Kai1, ZHANG Shuang1, ZHANG Li-jiao1, YUAN Jian-fei1, WU Jin1. Characteristics of Spectral Radiation for Al+ During Hypervelocity Impact on 2A12 Aluminum Target[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(08): 2381-2388. |
[7] |
NONG Jin-peng1, 2, WEI Wei1, 2, 3*, ZHU Yong2*, TANG Lin-long3, 4, ZHANG Gui-wen2, JIANG Xiao2, CHEN Na2, WEI Da-peng3, 4. Resonance Spectra and Sensing Characteristics of Plasmons in Graphene-Covered Aluminum Nano-Grating[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(04): 997-1002. |
[8] |
MI Hai-peng1, ZHU Hong-ming1, LI Gen-rong1, YU Xiang1, MA Bing-bing2, ZHOU Xi-lin2, WANG Ya-sen2, SU Zhong-hua2, DENG Xiong2. Determination of Pb, Cr, Cd, and As in Aluminum-Plastic Packaging Materials via Inductively Coupled Plasma-Mass Spectrometry with Microwave Digestion[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(02): 646-650. |
[9] |
LI Shu1, DI Jing-ru1*, WANG Xiao-juan2, CAO Bai-hui1 . The Gemmological Characteristics of a New Kind of Imitation of Turquoise [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(09): 2978-2983. |
[10] |
WU Xiu-li1, ZHAO Zhi-zhong2, WU Dan2, CHEN Guang-ying3, LIU Hong1, HE Meng-xiong1, DING Guo-hua1, HE Wen-ying1*. Determination of Main Active Components and Metal Elements of Hainan Alpinia Katsumadai by Spectral Analysis [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(04): 1191-1196. |
[11] |
DUAN Liang-fei1,2, YANG Wen1,2, ZHANG Li-yuan1,2, LI Xue-ming1,2, CHEN Xiao-bo1,2, YANG Pei-zhi1,2* . Study on the Properties of the Pc-Si Films Prepared by Magnetron Co-Sputtering at Low Temperature[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(03): 635-639. |
[12] |
ZHANG Jun1, ZENG Jie1*, WANG Bo2, WANG Wen-juan3, LIANG Da-kai1, LIU Xiao-ying1 . The Research on Optic Fiber FBG Corrosion Sensor Based on the Analysis of the Spectral Characteristics [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(03): 853-856. |
[13] |
ZHAO Huan-bin, SUN Qin-jun*, ZHOU Miao, GAO Li-yan, HAO Yu-ying*, SHI Fang. Study on the Effects of Alq3∶CsF Composite Cathode Buffer Layer on the Performances of CuPc/C60 Solar Cells[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(02): 331-335. |
[14] |
WANG Xiao-wei1, LIU Jing-fu2, GUAN Hong3, WANG Xiao-yan1, SHAO Bing1*, ZHANG Jing1, LIU Li-ping1, ZHANG Ni-na1. Determination of Total Sulfur Dioxide in Chinese Herbal Medicines via Triple Quadrupole Inductively Coupled Plasma Mass Spectrometry [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(02): 527-531. |
[15] |
HE Xiao1,3, ZHANG Li-sheng2, ZU En-dong1*, YANG Xiao-yun3, DONG Kun3 . Research on the Relationship between Surface Structure and Fluorescence Intensity of Ca(1-x)Al2Si2O8∶Eux [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(01): 146-150. |
|
|
|
|