Abstract:The Se in the blood and tissue of animal samples was determined by ICP-MS instrument. The method was not only proved simple and rapid, but also showed satisfying precision and low detection limit. The optimum conditions of ICP-MS for the method were investigated. Results by the digestion method, the isotopes selection, and the difference between the normal ICP-MS without ORS (Octopole Reaction System) and the ICP-MS with ORS were compared, which afford reference to the determination of the same kinds of sample. The standard materials were determined to proved the reliability of the normal ICP-MS and the ORS-ICP-MS methods. The ORS-ICP-MS can effectively remove the interference based on polyatomic ions, so the the detection limits were lowed, and the accuracy was improved. The detection limits of the method for normal ICP-MS is 0.024 ng·g-1 and that for ORS-ICP-MS is 0.004 6 ng·g-1, the RSDs are between 1.8% and 5.5%, and the recoveries of the method are in the range of 90.8%-107.2%.
王英锋1,刘翠梅1,刘少轻1,施燕支1*,陈玉红2. 普通/ORS模式-ICP-MS法测定动物血液和组织中硒的比较[J]. 光谱学与光谱分析, 2008, 28(09): 2173-2176.
WANG Ying-feng1,LIU Cui-mei1,LIU Shao-qing1,SHI Yan-zhi1*,CHEN Yu-hong2. Determination of Se in Blood and Tissue of Animal Samples by Normal ICP-MS and ICP-MS with Octupole Reaction System. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2008, 28(09): 2173-2176.
[1] ZHAO Yu, LIU Yu-wen, KANG Jing(赵 宇,刘玉文,康 敬). Chinese Primary Health Care(中国初级卫生保健), 2005, 19(6): 57. [2] WANG Xiu-hong(王秀红). Studies of Trace Elements and Health(微量元素与健康研究), 2006, 23(1): 66. [3] QIU Jian-hua, WANG Qiu-quan, HUANG Ben-li(邱建华,王秋泉,黄本立). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2006, 26(9): 1692. [4] YU Hong-wei, CHEN Chun-ying, GAO Yu-xi, et al(喻宏伟,陈春英,高愈希,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2006, 26(12): 2355. [5] XU Fang, QIU De-ren, YANG Peng-yuan, et al(徐 芳,邱德仁,杨原, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2002, 22(2): 331. [6] CHEN Hao, LIANG Pei, HU Bin, et al(陈 浩,梁 沛,胡 斌,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2002, 22(6): 1019. [7] LI Fang-shi, Walter Gossler, Kurt J Irgolic(李方实, Walter Gossler, Kurt J Irgolic). Chinese Journal of Chromatography(色谱), 1999, 17(3): 240. [8] YU Hong-wei, CHEN Chun-ying, GAO Yu-xi, et al(喻宏伟,陈春英,高愈希,等). Chinese Journal of Analytical Chemistry(分析化学), 2006, 34(6): 749. [9] DONG Yin-gen, SHEN Hui-jun(董银根,沈惠君). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2002, 22(4): 691. [10] CHEN Bai-ling, YAN Ke-mei, LIAO Quan-bin, et al(陈百玲,颜克美,廖全斌,等). Science of Guangdong Trace Elements(广东微量元素科学),2006, 13(11): 48. [11] XU Hui, HE Ping(许 卉,贺 萍). Chinese Journal of Analytical Chemistry(分析化学), 2003, 31(10): 1244. [12] HE Xiao-qing, LI Gong-ke, LIAO Zhu-hua(何小青,李攻科,廖祝华). Journal of Shenzhen University(Science and Engineering)(深圳大学学报·理工版), 1999, 16(2): 89. [13] CHEN Shu-yu, ZHANG Zhi-feng, YU Hua-ming(陈树榆,张志锋,余华明). Chinese Journal of Analysis Laboratory(分析试验室), 2004, 23(5): 9. [14] WEN Xin-yu, SONG Jian-wu, TAN Ya-ling, et al(文新宇,宋建武,谭亚翎, 等). Environmental Monitoring in China(中国环境监测), 2005, 21(3): 22. [15] PENG Kun, WU Heng(彭 琨, 吴 珩). Physical Testing and Chemical Analysis. Part B: Chemical Analysis(理化检验-化学分册), 2000, 36(12): 557. [16] ZHANG Li-min(张丽敏). Chin. J. Ind. Hyg. Occup. Dis.(中华劳动卫生职业病杂志), 2006, 24(1): 50. [17] SHI Yan-zhi, WANG Ying-feng, HE Rui-juan, et al(施燕支,王英锋,贺闰娟,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2005, 25(6): 955. [18] LI Zheng-jun, HUANG Jin-feng, LIU Jian-bin, et al(李政军,黄金凤,刘建斌, 等). Chinese Journal of Analysis Laboratory(分析试验室), 2007, 26(1): 76.