Abstract:Spectral calibration of space-born imaging spectrometers based on spectrum-matching technique is presented, which adopts atmospheric absorption lines as the matching lines, and chooses correlation coefficient method as the criteria. In order to simulation the onboard spectral calibration, the spectrum-matching technique is applied on the imaging spectrometers that after the vibration test. The vibration test is able to simulation the launching. The spectral resolution, center wavelength of two-dimensional pixel is determined during onboard spectral calibration. As the calibration results show, the spectral resolution of imaging spectrometers after the vibration test is 0.40 nm, it is the same comparing to the value before the vibration, the wavelength shifts 0.08 nm towards the long wave for the spectral pixels, and the spectral smile is determined for all spatial elements, which shifts towards the short wave direction, with the max smile value is 0.96 nm, the result is similar to that before the vibration. As a result, the spectrum-matching technique is tested and verified.
[1] Robert A Neville, Sun Lixin, Karl Staenz. Proc. of SPIE, 2003, 5093: 144. [2] Gao Bo-cai, Marcos J Montes, Curtiss O Davis. Remote Sensing of Environment, 2004: 424. [3] Marcel R Dobber, Ruud J Dirksen, Pieternel F Levelt, et al. Transactions on Geoscience and Remote Sensing, 2006, 44(5): 1209. [4] Haring R E, Williams F L, Hartmann U G, et al. Proc. of SPIE, 2004, 4135: 421. [5] Robert O Green. Applied Optics, 1998, 37(4): 683. [6] Barry P S, Shepanski J, Segal C. Proc. of SPIE, 2002, 4480: 231. [7] Luis Guanter, Rudolf, Jose Moreno. Applied Optics, 2006, 45(10): 2360. [8] ZHAO Min-jie, SI Fu-qi, JIANG Yu, et al(赵敏杰,司福祺,江 宇,等). Optics and Precision Engineering(光学 精密工程),2013, 21(3): 567. [9] ZHOU Hai-jin, LIU Wen-qing, SI Fu-qi, et al(周海金,刘文清,司福祺,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2012, 32(11): 2881. [10] Zhou Jiankang, Liu Xiaoli, Ji Yiqun, et al. Proc. of SPIE, 2012, 8557: 85571T-1.