光谱学与光谱分析 |
|
|
|
|
|
Study of Fundamental Parameters of Inductively Coupled Plasma for Slurry Nebulization——Electron Density Determination |
WANG Zheng1, 2, QIU De-ren2, TAO Guang-yi1, YANG Peng-yuan2* |
1. Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 200050, China 2. Department of Chemistry, Fudan University, Shanghai 200433, China |
|
|
Abstract The electron density of the plasma for slurry nebulization inductively coupled plasma emission spectrometry was determined and reported in the present paper. The Stark broadening method of Hβ line (486.1 nm) was chosen and used to determine the electron density of the plasma for nebulization into the inductively coupled plasma with both the aqueous solution and different concentration titanium slurry. There are approximately the same plasma electron density results of 1015 cm-3 for the two nebulization ways. The experiment verified that the plasma electron density only shows a litter decrease with 10% TiO2 suspension nebulization into the inductively coupled plasma. This means that the plasma electron density does not change remarkably with high content suspension nebulization into the inductively coupled plasma emission spectrometry instrument. It will help trace elements determination by using high concentration suspension nebulization into the inductively coupled plasma emission spectrometry instrument.
|
Received: 2008-01-28
Accepted: 2008-05-06
|
|
Corresponding Authors:
YANG Peng-yuan
E-mail: wangzheng@mail.sic.ac.cn
|
|
[1] Kornblum G R, Galan L D. Spectrochim. Acta, Part B, 1977, 32: 71. [2] Jarosz J, Mermet J M, Robin J P. Spectrochim. Acta, Part B, 1978, 33: 55. [3] Torres J, Sande M J, Mullen J J A M, et al. Spectrochim. Acta, Part B, 2006, 61: 58. [4] Yang P Y, Barnes R M. Spectrochim. Acta, Reviews, 1990, 13: 275. [5] Yang P Y, Horner J A, Sesi N N, et al. Spectrochim. Acta, Part B, 2000, 55: 1833. [6] LI Yi-jiu, ZENG Xin-ping, WANG Shi-long, et al(李义久,曾新平,汪世龙,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 1999,19(3):360. [7] WANG Zheng, CHEN Tian-yu, TAO Guang-yi, et al(汪 正, 陈天裕,陶光仪,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析) 2005,25(4):556. [8] WANG Zheng, QIU De-ren, TAO Guang-yi, et al(汪 正, 邱德仁,陶光仪,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析) 2006,26(3):542. [9] Sparkes S T, Ebdon L. J. Anal. At. Spectrom., 1988, 3: 563. [10] QIU De-ren(邱德仁). Analytical Atomic Spectrometry(原子光谱分析). Shanghai: Fudan University Press(上海: 复旦大学出版社), 2002. (Chapter 2). [11] Boumans P W J M. Inductively Coupled Plasma Emission Spectroscopy, Part New York: Wiley, 1987. [12] Montaser Akbar, Golightly D W. Inductively Coupled Plasmas in Analytical Atomic Spectrometry(感耦等离子体在原子光谱分析法中的应用). Translated by CHEN Long-mao, SHAO You-bin, et al(陈隆懋,邵友彬, 等). Beijing: People’s Medical Publishing House(北京:人民卫生出版社),1992. 207.
|
[1] |
WANG Zheng1, 2,QIU De-ren2,TAO Guang-yi1,YANG Peng-yuan2* . Study of Fundamental Parameters of Inductively Coupled Plasma with Slurry Nebulization ——I.Excitation Temperature Determination[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(03): 793-796. |
[2] |
WANG Yuan-zhong1,LI Tao2,LIU Hong-gao3*,LI Xing-kui3 . Direct Determination of Manganese in Edible Fungi Using Suspension Sample Introduction and Hypo-Sensitive Wavelength by Flame Atomic Absorption Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27(05): 1021-1023. |
|
|
|
|