Determination of Various Elements in Ceramic Materials by Wavelength Dispersive X-Ray Fluorescence Spectrometry With Fusion Sample
Preparation |
NI Zi-yue, LIU Ming-bo*, ZHENG Qi, HU Xue-qiang, YUE Yuan-bo, YANG Bo-zan, FAN Zhen, LI Cheng |
NCS Testing Technology Co., Ltd., Beijing 100094, China
|
|
|
Abstract The chemical composition of ceramic materials directly affects their properties, and due to their diverse composition, when measured by X-ray fluorescence spectroscopy after being pressed directly, the matrix greatly influences the determination of samples.In this paper, the fused pelletswerepreparedfor various types of ceramic materials, the accurate correction of the concentration was discussed by the application of loss on ignition(LOI)for the elements to be measured in the sample, the relative error of concentration was analyzed for different dilution ratios and different sampling approaches, and the test of multiple components was studied in various types of matrices by wavelength dispersive X-ray fluorescence spectrometry. Then, the calibration curve is verified by the validation group. When the actual samplesareprepared, the accurate value of the ignition loss can be used to correct the concentration of the element to be measured in the sample. Different correction formulas are required to correct the content when the sample is weighed before or after the ignition. When the loss of ignition is not corrected, the concentration error is significantly lower when weighted before ignition than when weighted after ignition. When the sample dilution ratio is 1∶10 and 1∶20, the relative error caused by 50% burning loss is only 0.43% and 0.12% after using a dilution ratio 1∶20 for sample preparation and applying the estimated standard error to evaluate the improvement of the working curve. After correction, the linear determination coefficient of each component reached more than 0.99. At the same time, the verification group was used to test the correctness of the method.
|
Received: 2024-02-29
Accepted: 2024-08-21
|
|
Corresponding Authors:
LIU Ming-bo
E-mail: Liumingbo@ncschina.com
|
|
[1] ZHANG Hong-quan, XU Shao-hua, ZHENG Qiu-rong, et al(张宏泉, 徐绍华, 郑秋荣, 等). Journal of Wuhan University of Technolog(武汉理工大学学报), 2019, 42(11): 27.
[2] LÜ Jun-dong(吕俊栋). Ceramics(陶瓷), 2022, (4): 29.
[3] WANG Chuang, ZHANG Kuai, YAN Hong-yan, et al(王 闯, 张 快, 严红艳, 等). Multipurpose Utilization of Mineral Resources(矿产综合利用), 2020, (4): 8.
[4] JIANG Xiao-liang, WANG Yang, HU Da-wei, et al(蒋小良, 王 旸, 胡大伟, 等). Foshan Ceramics(佛山陶瓷), 2019, 29(4): 29.
[5] HE Peng, CHEN Lian, LÜ Xiao-yuan, et al(贺 鹏, 陈 练, 吕小园, 等). Journal of Food Safety and Quality(食品安全质量检测学报), 2016, (1): 177.
[6] LIN Shan(林 珊). Foshan Ceramics(佛山陶瓷), 2012, (6): 28.
[7] Nicola Forster, Peter Grave, Nancy Vickery, et al. X-Ray Spectrometry, 2011, 40(5): 389.
[8] SHENG Cheng, ZHUO Shang-jun, JI Ang, et al(盛 成, 卓尚君, 吉 昂, 等). Physical Testing and Chemical Analysis-Part B: Chemical Analysis(理化检测-化学分册), 2012, 48(6): 629.
[9] LIU Ming-bo, WANG Ying-qian, NI Zi-yue, et al(刘明博, 王英倩, 倪子月, 等). Metallurgical Analysis(冶金分析), 2023, 43(1): 1. |
[1] |
WANG Guang-xi, LI Dan*, GE Liang-quan, CHEN Cheng, LAI Wan-chang, ZHAI Juan, LAN Yong, HU Jie, LONG Fei . Rapid Determination of Cu and Zn in PM2.5 with Wavelength Dispersive X-Ray Fluorescence Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(04): 1240-1244. |
[2] |
WANG Zheng1, WANG Shi-wei1, QIU De-ren2, YANG Peng-yuan2* . The Application of Inductively Coupled Plasma Atomic Emission Spectrometry/Mass Spectrometry to the Analysis of Advanced Ceramic Materials[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(10): 2838-2842. |
[3] |
SONG Yi, GUO Fen, GU Song-hai*. Method of Fused Sample Preparation after Nitrify—Determination of Primary and Minor Elements in Manganese Ore by X-Ray Fluorescence Spectrometry [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27(02): 404-407. |
|
|
|
|