Design of High-Throughput μ-EDXRF |
LIU Ming-bo1, 2, ZHAO Lei1, 2, HU Xue-qiang2, NI Zi-yue1, 2, YANG Li-xia1, 2,JIA Yun-hai1, 2, WANG Hai-zhou1, 2* |
1. Central Iron and Steel Research Institute, Beijing 100081, China
2. NCS Testing Technology Co., Ltd., Beijing 100081, China
|
|
|
Abstract A high-through put μ-EDXRF instrument named NX-mapping with polycapillary optics and 4~6 SDD detectors placed symmetrically around the emitted focal spot was developed. We designed the software to control the X-ray source, multiple detectors, and high-precision stage and carried out the work of data collection, signal synchronization and result in analysis simultaneously. Due to the use of multiple detectors, the intensity of this XRF instrument was improved several times, and the RSD was reduced to 40% hasn’t deteriorated due to the differences between several detectors and the moving stage.There is no significant difference between the RSD of scanning a uniform sample with 2 mm×2 mm square 400 pixels and the fixed-point test, indicating that the motion mechanism and control algorithm performed well. The fluorescent knife-edge test was carried out forthe effective focal size testing of NX-mapping. The minimum effective focal size of Fe, Ni and Mo elements are 52.4,49.3 and 39.03 μm. The law that each element’s effective focal spot size decreases with the increase of atomic number is summarized, which is consistent with the design principle of polycapillary optics.It is also found that the effective size of the focal spot of each element is sensitive to the change of height at the minimum value. Therefore, it is recommended to keep the surface of the sample flat in order to obtain uniform and clear images.Finally, the characteristic intensity of elements Ni, Ta, W and Re of a single crystal superalloy sample were scanned and analyzed by NX-mapping. The images of the dendritic structure can be distinguished, and the intensity of the Ni∶Ka characteristic line was up to 220 kcps, which was significantly higher than that of the ordinary XRF test. The NX-mapping μ-EDXRF instrument with multiple detectors, which provide high signal intensity, good test precision and shorter test time,can meet the requirements of high-through put testing.
|
Received: 2021-07-23
Accepted: 2021-10-26
|
|
Corresponding Authors:
WANG Hai-zhou
E-mail: wanghaizhou@ncschina.com
|
|
[1] YANG Ming-tai, TANG Hui(杨明太, 唐 慧). Nuclear Electronics & Detection Technology(核电子学与探测技术), 2011,(12): 1307.
[2] LUO Li-qiang, SHEN Ya-ting, MA Yan-hong, et al(罗立强, 沈亚婷, 马艳红, 等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2017,37(4): 1003.
[3] SUN Tian-xi, XU Guang-yu, LIU Zhi-guo, et al(孙天希, 徐光瑜, 刘志国, 等). Acta Optica Sinica(光学学报), 2008, 29(9): 1833.
[4] CHU Xue-lian, LIN Xiao-yan, CHENG Lin, et al(初学莲, 林晓燕, 程 琳, 等). Journal of Beijing Normal University·Natural Science(北京师范大学学报·自然科学版), 2007,(5): 530.
[5] ZHAO Ji-cheng(赵继成). Chinese Journal of Nature(自然杂志), 2014, 36(2): 89.
[6] Wang Haizhou, Zhao Lei, Jia Yunhai, et al. Engineering, 2020, 6(6): 621.
[7] SUN Tian-xi, DING Xun-liang, LIU Zhi-guo, et al(孙天希, 丁训良, 刘志国, 等). Chinese Physics C(中国物理C), 2005, 29(10): 1017.
[8] Vincze L,Wei F,Proost K,et al. J. Anal. At. Spectrom., 2002, 17(3): 177.
[9] SUN Tian-xi, DING Xun-liang, LIU Zhi-guo, et al(孙天希, 丁训良, 刘志国, 等). Acta Optica Sinica(光学学报), 2005, 25(10): 1436.
|
[1] |
ZHOU Qing-qing1, LI Dong-ling1, 2, JIANG Li-wu1, 3*, WAN Wei-hao1, ZENG Qiang4, XUE Xin4, WANG Hai-zhou1, 2*. Quantitative Statistical Study on Dendritic Component Distribution of Single Crystal Blade Based on Microbeam X-Ray Fluorescence Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(07): 2112-2118. |
[2] |
CAO Xiao-feng1, 2, 3, YU Ke-qiang1, 2, 3, ZHAO Yan-ru1, 2, 3*, ZHANG Hai-hui1, 2, 3*. Current Status of High-Throughput Plant Phenotyping for Abiotic Stress by Imaging Spectroscopy: A REVIEW[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(11): 3365-3372. |
[3] |
LÜ Jing-jing1,2, GONG Wei-jin1, DOU Yan-yan1, DUAN Xue-jun1, LIU Hai-fang1, ZHANG Lie-yu3, XI Bei-dou3, YU Shui-li2, HOU Li-an2,4. Spectroscopy and Microbiological Analysis of Soil Infiltration Clogging in Treating Aged Swine Wastewater[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(05): 1541-1546. |
|
|
|
|