Influence of Collimation System on Static Fourier Transform Spectrometer
JIANG Cheng-zhi2, 3, LIANG Jing-qiu1*, LIANG Zhong-zhu1, SUN Qiang2,WANG Wei-biao1
1. State Key Laboratory of Applied Optics, Changchun Institute of Optics Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China 2. Opto-Electronics Technology Center, Changchun Institute of Optics, Fine Mechanics and Physics, Chinese Academy of Sciences, Changchun 130033, China 3. University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:Collimation system provides collimated light for the static Fourier-transform spectroscopy (SFTS). Its quality is crucial to the signal to noise ratio (SNR) of SFTS. In the present paper, the physical model of SFTS was established based on the Fresnel diffraction theory by means of numerical software. The influence of collimation system on the SFTS was discussed in detail focusing on the aberrations of collimation lens and the quality of extended source. The results of simulation show that the influences of different kinds of aberrations on SNR take on obvious regularity, and in particular, the influences of off-axis aberrations on SNR are closely related to the location of off-axis point source. Finally the extended sources maximum radius allowed was obtained by simulation, which equals to 0.65 mm. The discussion results will be used for the design of collimation system.
Key words:Spectroscopy;Design of collimation system;Fresnel diffraction;Aberration;Extended source