Abstract:The contents of platinum in Membrane Electrode Assembly(MEA) is very important to the cost of proton exchange membrane fuel cell. In this paper, the contents of platinum in MEA were determinated by graphite furnace atomic absorption spectrometry. The influences of the program for graphite furnace and the optimum conditum of experiment were systematically studied. The method is simple, rapid, accurate and practical. The result indicates that the relative standard deviation of Pt determination is less than 2.67%. The rate of recovery is 97.3%-101.2%.
[1] HUANG Zhuo et al(黄 倬等). Development and Application of the Proton Exchange Membrance Full Cell(质子交换膜燃料电池的研究开发与应用). Beijing: Metallurgical Industry Press(北京:冶金工业出版社),2000. [2] SHAO Zhi-gang et al(邵志刚等). Chinese Journal of Power Source(电源技术),2000, 24(1): 42. [3] FANG Hui-qun et al(方惠群等). Analysis of Instrument(仪器分析). Beijing: Science Press(北京:科学出版社),2002. 245. [4] Ro N K,Mistry M. Atomic Spectroscopy,2001, 22(4): 356.