Development and Performance Test of a Seya-Namioka Vacuum
Ultraviolet Spectroscopy System
SHEN Yong-cai1, NIU Yu-fan1, KONG De-feng2, YE Yang2*, ZHANG Shou-biao3, LI Da-chuang1, TAN Ming-sheng2, HUANG Yan-qing4, ZHAO Zhi-hao5, ZI Peng-fei3, ZHANG Xiao-hui5, WEN Fei3
1. School of Physics and Materials Engineering, Hefei Normal University, Hefei 230601, China
2. Institute of Energy, Hefei Comprehensive National Science Center, Hefei 230031, China
3. Institute of Plasma Physics, Chinese Academy of Sciences, Hefei 230031, China
4. School of Physics and Electronic Engineering, Hengyang Normal University, Hengyang 421008, China
5. School of Computer Science and Information Engineering, Hefei University of Technology, Hefei 230009, China
Abstract:Impurity behavior study is very important in magnetic confined fusion research as impurity may cause the dilution of fuel ions, affect the power balance, and degrade plasma performance. Spectroscopic diagnostic is important for impurity measurement and transport study in fusion devices. Spectroscopy in the vacuum ultraviolet (VUV) range offers a useful tool for investigating impurity radiation from low-temperatureareasof edge tokamak plasma. To meet the impurity measurement requirement in fusion research, a Seya-Namioka spectrometer was designed, and the main parts of the spectrometer are adjustable width incident slits, concave gratings, and a detector). The luminous flux of the spectrometer is adjusted through an adjustable width of the incident slit, which is of the linear guide type. The position of the incident slit baffle is adjusted through the linear guide to achieve slit width adjustment. The grating surface is coated with aluminum (Al) and magnesium fluoride (MgF2) to enhance the refractive efficiency. The usable wavelength range of the grating is 50~460 nm, and the optical path optimization design is carried out for the 50~50 nm wavelength range. The spectrometer's detector was chosen as a deeply cooled back-illuminated charge-coupled device (CCD). By turning the grating turntable to rotate the grating and change the diffraction angle, spectral observations in the 50~250 nm range can be achieved. Based on the parameters of the concave grating, the specific optical path was determined, and the relationship between wavelength and grating rotation angle, as well as the line dispersion rate at different wavelengths, was analyzed. According to the theory of concave grating imaging, the spectral resolution of the system was calculated and analyzed. The optimal exit arm was determined to be 205 mm by analyzing the spectral resolution under different exit arms. The effects of incident slit width, pixel size, aberration, and diffraction limit on spectral resolution were analyzed with an exit arm of 205 mm and an incident slit of 20 μm. The main contributions to the spectral resolution are the entrance slit width, pixel size, aberration, and diffraction limit. The diffraction limit has the smallest effect on spectral resolution, which can be ignored. Due to the size of the detector pixel, the width of the exit slit has a great impact on the spectral resolution, which remains at about 0.09~0.10 nm and increases slowly in the 50~250 nm wavelength range. The total spectral resolution is between 0.121 nm and 0.122 nm. The width of the incident slit will be adjustable from 10 to 1 000 μm, depending on the intensity of the incident light. The wider the width of the incident slit, the worse the overall spectral resolution will be. In actual measurement, the luminous flux and spectral resolution should be considered. Wavelength calibration and performance tests were performed by using a low-pressure mercury lamp and microwave plasma light source. The wavelength calibration of the spectrometer can be completed based on the zero-order spectrum and the characteristic wavelength of the mercury lamp (Hg Ⅰ 185 nm). The zero-order spectrum position is defined as the diffraction angle zero position. The angular position of Hg Ⅰ 185 nm in the spectrometer is determined according to the formula. The spectral resolution was 0.124 3 at 185 nm by Gauss fitting of the Hg Ⅰ (185 nm) spectral line, which is close to the calculated value. By comparing the spectral resolution of Hg Ⅰ 185 nm under different exit arms through experiments and theoretical calculations, it was further verified that the instrument achieved the best spectral resolution at the exit-arm length of 205 mm. Further performance tests of the spectrometer were carried out using a microwave plasma discharge light source device. It has been verified that the spectrometer has good detection ability in the 50~250 nm wavelength range, based on observing nitrogen, oxygen, and helium lines emitted by microwave plasma light sources.
沈永才,钮雨凡,孔德峰,叶 扬,张寿彪,李大创,谭名昇,黄艳清,赵志豪,訾鹏飞,张小辉,文 斐. Seya-Namioka类型真空紫外光谱系统的研制和性能测试[J]. 光谱学与光谱分析, 2024, 44(08): 2158-2165.
SHEN Yong-cai, NIU Yu-fan, KONG De-feng, YE Yang, ZHANG Shou-biao, LI Da-chuang, TAN Ming-sheng, HUANG Yan-qing, ZHAO Zhi-hao, ZI Peng-fei, ZHANG Xiao-hui, WEN Fei. Development and Performance Test of a Seya-Namioka Vacuum
Ultraviolet Spectroscopy System. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(08): 2158-2165.
[1] Isler R C. Nuclear Fusion, 1984, 24(12): 1599.
[2] Shen Yongcai, Lyu Bo, Du Xuewei, et al. Fusion Engineering and Design, 2013, 88(11): 3072.
[3] Shen Yongcai, Du Xuewei, Zhang Wei, et al. Nuclear Instruments & Methods in Physics Research, 2013, 700(1): 86.
[4] Field A, Fink R J, Dux R, et al. Review of Scientific Instruments, 1995, 66(12): 5433.
[5] Maggi C F, Elder J D, Fundamenski W, et al. Journal of Nuclear Materials, 1997, 241-243(1): 414.
[6] Morita S, Goto M. Reviews of Scientific Instrument, 2003, 74(3): 2036.
[7] Cui Z, Morita S, Fu B, et al. Review of Scientific Instruments, 2010, 81(4): 043503.
[8] HE Liang, ZHANG Hong-ming, CAO Jin-jia, et al(何 梁, 张洪明, 曹锦佳, 等). Nuclear Fusion and Plasma Physics(核聚变与等离子体物理), 2021, 41(2): 110.
[9] Li Y, Cheng Z F, Zhang X L, et al. Journal of Instrumentation, 2017, 12(6): C06043.
[10] ZHANG Kai, ZHENG Dian-lin, DONG Chun-feng, et al(张 凯,郑典麟,董春凤,等). Nuclear Fusion and Plasma Physics(核聚变与等离子体物理), 2021,41(2): 527.
[11] Kong Defeng, Zhuang Ge, Lan Tao, et al. Plasma Science and Technology. 2023, 25: 065601.
[12] MENG Zhu, TAN Xin, Bayanheshig, et al(孟 竹,谭 鑫,巴音贺希格,等). Chinese Optics(应用光学), 2018, 11: 623.
[13] WANG Qiu-ping, ZHANG Yun-wu, LIU Shao-yi, et al(王秋平,张允武,刘绍义,等). Vacuum(真空), 1994,4:40.
[14] XU Xiang-dong, LIU Zheng-kun, QIU Ke-qiang, et al(徐向东,刘正坤,邱克强,等). Optics and Precision Engineering(光学精密工程), 2012, 12(1): 1.
[15] Zhang Hongming, Lyu Bo, He Liang, et al. Plasma Science and Technology. 2020, 22: 084001.
[16] XU Chao-yin(徐朝银). Optics and Engineering of Synchrotron Radiation(同步辐射光学与工程). Hefei: University of Science and Technology of China Press(合肥:中国科技大学出版社), 2013: 85.
[17] He Liang, Zhang Hongming, Lyu Bo, et al. Plasma Science and Technology, 2022, 24: 064003.
[18] West J B, Padmore H A. Handbook on Synchrotron Radiation, Chapter 2-Optical Engineering, 1987: 21.
[19] LI Xin, ZHANG Guo-wei, XUN Li-na, et al(李 新,张国伟,寻丽娜,等). Acta Optica Sinica(光学学报), 2008, 28(5): 902.
[20] HUANG Tao, LU Hui, WANG Mao-bi,et al(黄 桃,卢 辉,王茂碧,等). Vacuum Electronics(真空电子技术), 2013, 307(6): 73.
[21] Xin Renxuan. Chinese Journal of Inorganic Analytical Chemistry, 2012, 2(4): 1.
[22] Kramida A. Atomic Spectra Database From National Institute of Standards and Technology[EB/OL]. 2022. http://physics.nist.gov/PhysRefData/ASD/index.html