Temperature Measurement Method for Small Target Medium-Wave
Infrared Spectral Radiation Based on Distance Correction
LI Wen-kai1, 2, ZHOU Liang1*, LIU Zhao-hui1, CUI Kai1, LIU Kai1, LI Zhi-guo1,XIE Mei-lin1
1. Xi'an Institute of Optics and Precision Mechanics, Chinese Academy of Sciences, Xi'an 710119, China
2. University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:For long-distance space targets moving at high speeds, temperature is one of the important parameters to characterize their working state and performance. Accurately obtaining the temperature of the target has an important reference value for judging its motion state and predicting its situation development. At present, the commonly used processing method of surface target or point target is no longer applicable to the measurement of the radiation characteristics of small targets. At the same time, spectral detection increases the distinguishable information of the target in the wavelength dimension, which can accurately obtain the distribution of the target energy with wavelength, providing a possibility for the inversion of the target temperature, and has great application potential. The slitless spectrometer can reduce the requirements for tracking and stabilization accuracy of space targets, has the characteristics of simple structure, high frame rate and fast response speed, and has high application value in astronomical observation and spacecraft observation. In this paper, we analyzed the spectral calibration model of target infrared radiation characteristic measurement and determined the main parameters in the linear response model of infrared detector pixels. In order to reduce the influence of imaging distance on temperature measurement accuracy, we proposed a target temperature inversion model based on distance correction. The improved temperature measurement accuracy meets the accuracy requirements in practical engineering applications and greatly affects infrared radiation spectrum temperature measurement. Certain guiding significance.
Key words:Radiation characteristics measurement; Slitless spectrometer; Spectral calibration; Temperature inversion
李文凯,周 亮,刘朝晖,崔 凯,刘 凯,李治国,谢梅林. 基于距离修正的小目标中波红外光谱辐射测温方法[J]. 光谱学与光谱分析, 2024, 44(04): 1158-1164.
LI Wen-kai, ZHOU Liang, LIU Zhao-hui, CUI Kai, LIU Kai, LI Zhi-guo,XIE Mei-lin. Temperature Measurement Method for Small Target Medium-Wave
Infrared Spectral Radiation Based on Distance Correction. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(04): 1158-1164.
[1] de Jong A N. Proc SPIE, 1995, 2552: doi:10.1117/12.218217.
[2] CHANG Song-tao, ZHANG Yao-yu, SUN Zhi-yuan, et al(常松涛,张尧禹,孙志远,等). Acta Optica Sinica(光学学报), 2014, 34(5): 15.
[3] TIAN Qi-jie, LI Zhou, CHANG Song-tao, et al(田棋杰,李 周,常松涛,等). Acta Optica Sinica(光学学报), 2017, 37(10): 191.
[4] Pyatt E C. British Journal of Applied Physics, 1954, 5: 264.
[5] SUN Xiao-gang, HE Jin, DAI Jing-min, et al(孙晓刚,何 瑾,戴景民,等). Journal of Harbin Institute of Technology(哈尔滨工业大学学报), 1998, 30(6): 1.
[6] ZHANG Fu-cai, SUN Bo-jun, SUN Xiao-gang(张福才,孙博君,孙晓刚). Infrared and Laser Engineering(红外与激光工程), 2019, 48(2): 276.
[7] ZHANG Fu-cai, SUN Bo-jun, SUN Xiao-gang(张福才,孙博君,孙晓刚). Acta Optica Sinica(光学学报), 2019, 39(2): 208.
[8] SUN Bo-jun, SUN Xiao-gang, DAI Jing-min(孙博君,孙晓刚,戴景民). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2021, 41(5): 1336.
[9] Willott Chris J, Doyon René, Albert Loic, et al. Publications of the Astronomical Society of the Pacific, 2022, 134(1032): 025002.
[10] McHarg Matthew G, Stenbaek-Nielsen H C, Kammae T. Observations of the STARDUST Sample Return Capsule Entry Using a High Frame Rate Slit-Less Spectrograph, 46th Aiaa Aerospace Sciences Meeting and Exhibit, 2008.
[11] CAO Li-hua, WAN Chun-ming, ZHANG Yun-feng, et al(曹立华,万春明,张云峰,等). Journal of Infrared and Millimeter Waves(红外与毫米波学报), 2015, 34(4): 460.
[12] XING Qiang-lin, HUANG Hui-ming, XIONG Ren-sheng, et al(邢强林,黄惠明,熊仁生,等). Acta Photonica Sinica(光子学报), 2004, 33(7): 893.
[13] YANG Ci-yin, ZHANG Jian-ping, CAO Li-hua(杨词银,张建萍,曹立华). Chinese Journal of Scientific Instrument(仪器仪表学报), 2013, 34(2): 304.