Research on High Sensitive Detection of HCl Trace Gas Based on QEPAS Technology |
MA Yu-fei1, 2, HE Ying1, YU Xin1, YANG Chao-bo1, YANG Zhen1, BAI Xue-han1, SUN Rui2 |
|
|
Abstract Hydrogen chloride (HCl) is a toxic and harmful gas. It is of great significance to detecte it with high sensitivity. But up to now there has been a few research reports about the HCl detection using laser spectroscopy. Quartz-enhanced photoacoustic spectroscopy (QEPAS) technology was invented not long ago. This technique uses a commercially available mm sized piezoelectric quartz tuning fork (QTF) as an acoustic wave transducer. A high Q-factor and a ~32.7 kHz resonance frequency of the QTF improve QEPAS selectivity and immunity to environmental acoustic noise. QEPAS sensor has the advantages of high sensitivity, selectivity and compactness. In this paper, 5 000 ppm HCl∶N2 mixture was selected as the target analyte. A continuous-wave distributed feedback (DFB) single mode diode laser emitting at 1 742.38 μm was used as the exciting source and QEPAS technology was adopted as the detection method. In order to reduce the sensor background noise and simplify the data process, wavelength modulation spectroscopy and a 2nd harmonic detection technique were employed. In this research, acoustic micro-resonator (mR) in the acoustic detection module and the effect of acoustic wave improvement was discussed firstly. The “on beam” architecture of mR was selected and the length and inner diameter of the mR tubes were selected to be 4 and 0.5 mm, respectively. In the experiments, the influence of modulation depth of laser wavelength on QEPAS signal amplitude was investigated. When the acquisition time was 1 sec and modulation depth was 0.23 cm-1, a 1σ minimum detectable concentration limit of 815 ppb was obtained. The corresponding normalized noise equivalent absorption (NNEA) coefficient was 7.41×10-9 cm-1·W·Hz-1/2. In the future, the detection sensitivity can be further improved by adding water vapor into the target HCl gas to enhance the V-T (vibrational-translational) relaxation rate of HCl molecular.
|
Received: 2015-12-04
Accepted: 2016-04-12
|
|
|
[1] Kim S, Klimecky P, Jeffries J B, et al. Measurement Science and Technology, 2003, 14: 1662.
[2] Matsuguchi M, Kadowaki Y, Noda K, et al. Sensors and Actuators B, 2007, 120: 462.
[3] Nakagawa K, Sadaoka Y, Supriyatno H, et al. Sensors and Actuators B, 2001, 76: 42.
[4] Ortwein P, Woiwode W, Fleck S, et al. Experiments in Fluids, 2010, 49: 961.
[5] Kosterev A A, Bakhirkin Y A, Curl R F, et al. Optics Letters, 2002, 27: 1902.
[6] Ma Y F, Lewicki R, Razeghi M, et al. Optics Express, 2013, 21: 1008.
[7] Ma Y F, Yu X, Yu G, et al. Applied Physics Letters, 2015, 107: 021106.
[8] Dong L, Spagnolo V, Lewicki R, et al. Optics Express, 2012, 19: 24037.
[9] Dong L, Kosterev A A, Thomazy D, et al. Applied Physics B, 2010, 100: 627.
[10] Liu K, Guo X, Yi H, et al. Optics Letters, 2009, 34: 1594.
[11] Serebryakov D V, Morozov I V, Kosterev A A, et al. Quantum Electronics, 2010, 40: 167.
[12] Kosterev A A, Tittel F K, Serebryakov D V, et al. Review of Scientific Instruments,2005, 76: 043105. |
[1] |
WANG Chun-hui1, 2, YANG Na-na2, 3, FANG Bo2, WEI Na-na2, ZHAO Wei-xiong2*, ZHANG Wei-jun1, 2. Frequency Locking Technology of Mid-Infrared Quantum Cascade Laser Based on Molecule Absorption[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(08): 2363-2368. |
[2] |
TIAN Si-di1, WANG Zhen1, DU Yan-jun2, DING Yan-jun1, PENG Zhi-min1*. High Precision Measurement of Spectroscopic Parameters of CO at 2.3 μm Based on Wavelength Modulation-Direct Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(07): 2246-2251. |
[3] |
WANG Yi-hong, ZHOU Bin*, ZHAO Rong, WANG Bu-bin. Calibration-Free Wavelength Modulation Spectroscopy for Gas Properties Measuring Basedon 2nd and 4th Harmonics[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(02): 368-373. |
[4] |
JIANG Ya-jing, SONG Jun-ling*, RAO Wei, WANG Kai, LOU Deng-cheng, GUO Jian-yu. Rapid Measurement of Integrated Absorbance of Flow Field Using Extreme Learning Machine[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1346-1352. |
[5] |
HUANG An1, 2, XU Zhen-yu1, XIA Hui-hui1, YAO Lu1, RUAN Jun1, HU Jia-yi1, ZANG Yi-peng1, 2, KAN Rui-feng1*. Measurement Method of Two-Dimensional Distribution of Temperature and Components in Gas Turbine Combustor Based on Wavelength Modulated Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(04): 1144-1150. |
[6] |
WANG Jing-jing1, 2, DONG Yang2, TIAN Xing2, CHEN Jia-jin2, TAN Tu2, ZHU Gong-dong2, MEI Jiao-xu2, GAO Xiao-ming1, 2*. Application of White Noise Perturbation in Wavelength Modulated Off-Axis Integrated Cavity Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(09): 2657-2663. |
[7] |
LI Jin-yi1, FAN Hong-qing1, TIAN Xin-li1, LI Hong-lian2, WU Zhi-chao1, SONG Li-mei1. Pressure Correction for Calibration-Free Measurement of Wavelength Modulation Spectroscopy in Atmospheric Environment[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(05): 1407-1412. |
[8] |
ZHANG Bu-qiang1, 2, XU Zhen-yu1, LIU Jian-guo1, XIA Hui-hui1, FAN Xue-li1, NIE Wei1, 2, YUAN Feng1, 2, KAN Rui-feng1. Modulation Characteristics of Laser Based on Wavelength Modulation Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(09): 2702-2707. |
[9] |
LI Chun-guang1, 2, 3, 4, DONG Lei2*, ZHENG Chuan-tao3, WANG Yi-ding3, LIN Jun1*. Mid-Infrared Trace Ethane Sensor Design and Stability Analysis[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(03): 959-963. |
[10] |
QIU Xuan-bing1, SUN Dong-yuan1, LI Chuan-liang1*, WU Ying-fa1, ZHANG En-hua1, WEI Ji-lin1, WANG Gao2*, YAN Yu3*. Wavelet Denoising Research for the Tunable Laser Diode Absorption Spectroscopy of the CO at 1.578 μm[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(02): 628-633. |
[11] |
YANG Yan-fang1, 2, PEI Kai-long1, 2, YIN Xu-kun1, 2, WU Hong-peng1, 2, LI Shang-zhi1, 2, CUI Ru-yue1, 2, MA Wei-guang1, 2, ZHANG Lei1, 2, YIN Wang-bao1, 2, DONG Lei1, 2*, JIA Suo-tang1, 2. Photoacoustic Spectroscopy Based Methane Sensor Using a Double-Pass Photoacoustic Cell[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(02): 616-620. |
[12] |
ZHU Gao-feng1,2, YANG Chun-hua1, ZHU Hong-qiu1*, GUI Wei-hua1. Oxygen Concentration Detection and Calibration Method Improvement in Pharmaceutical Vial Based on Wavelength Modulation Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(10): 3133-3137. |
[13] |
MA Yu-fei1,2, TONG Yao1, ZHANG Li-gong1, HE Ying1, ZHANG Jing-bo1, WANG Long1, LONG Jin-hu1, YU Xin1, SUN Rui2. Study on High Sensitive Detection of Acetylene Trace Gas Based on QEPAS[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(09): 2869-2872. |
[14] |
QU Dong-sheng, HONG Yan-ji*, WANG Guang-yu, PAN Hu, WANG Ming-dong. Wavelength-Modulation Spectroscopy for Measurements of Gas Pressure, Temperature and H2O Concentration in High-Temperature Environment[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(05): 1339-1344. |
[15] |
QU Shi-min, WANG Ming, LI Nan . Mid-Infrared Trace CH4 Detector Based on TDLAS-WMS [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(10): 3174-3178. |
|
|
|
|