光谱学与光谱分析 |
|
|
|
|
|
Calibration Procedure of Laser Confocal Micro-Raman Spectrometer |
ZHAO Ying-chun1, 2, REN Ling-ling2*, WEI Wei-sheng1, YAO Ya-xuan2 |
1. School of Chemical Engineering, China University of Petroleum, Beijing 102249, China 2. Division of Nano-Merology and Materials Measurement, National Institute of Metrology, China, Beijing 100029, China |
|
|
Abstract As a common spectral characterization technique, Raman spectroscopy is widely used and has a specified calibration procedure. Based on laser confocal micro-Raman spectrometer, in this paper, we briefly introduced the principle, configuration and main components of Raman spectrometer. In addition, the calibration procedures were also presented, with an emphasis on the calibration of spectrometer (spectrograph) and that of excitation laser wavelength. On the basis of conventional calibration method, a novel and more accurate method was proposed to obtain the actual excitation wavelength, that is, calibration at the point of Raman shift Δν=0. Using this novel calibration method of excitation wavelength, Raman frequency shift values of sulfur were measured, and compared with the standard values from American Society Testing and Materials (ASTM). As a result, the measured values after calibration were consistent with those ASTM values, which indicated that the calibration method is accurate. Thus, a more reasonable calibration procedure of the laser confocal micro-Raman spectrometer was provided.
|
Received: 2014-05-12
Accepted: 2014-08-25
|
|
Corresponding Authors:
REN Ling-ling
E-mail: renll@nim.ac.cn
|
|
[1] Smith E, Dent G. Modern Raman Spectroscopy: A Practical Approach. John Wiley & Sons, 2005. 1. [2] Schmidt H, Bich Ha N, Pfannkuche J, et al. Marine Pollution Bulletin, 2004, 49(3): 229. [3] Abdallah W A, Yang Y. Energy & Fuels, 2012, 26(11): 6888. [4] Li M, Xu J, Romero-Gonzalez M, et al. Current Opinion in Biotechnology, 2012, 23(1): 56. [5] El-Abassy R M, Donfack P, Materny A. Journal of Raman Spectroscopy, 2009, 40(9): 1284. [6] Tedesco J M, Davis K L. Photonics East (ISAM, VVDC, IEMB). International Society for Optics and Photonics, 1999. 200. [7] ZHANG Shu-lin(张树霖). Raman Spectroscopy And Low-dimensional Semiconductors(拉曼光谱学与低维纳米半导体). Beijing:Science Press(北京:科学出版社),2008. 72. [8] Laane J, Kiefer W. The Journal of Chemical Physics, 2008, 72(10): 5305. [9] ASTM Standards. E1840—96. West Conshohocken, Pa.: ASTM USA,2007. |
[1] |
YUAN Liang-jing1, JIA Yun-hai1, 2, CHENG Da-wei1, 2*. Study on Methods of Detection Limit in XRF Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(02): 412-418. |
[2] |
LI Huan-tong1, 2, CAO Dai-yong3, ZOU Xiao-yan3, ZHU Zhi-rong1, ZHANG Wei-guo1, XIA Yan4. Raman Spectroscopic Characterization and Surface Graphitization Degree of Coal-Based Graphite With the Number of Aromatic Layers[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(08): 2616-2623. |
[3] |
LIU Zhen-wen1, XU Ling-jie2*, CHEN Xiao-jing3. Near Infrared Spectroscopy Transfer Based on Deep Autoencoder[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(07): 2313-2318. |
[4] |
LONG Yao-wei, SUN Hong, GAO De-hua, ZHENG Zhi-yong, LI Min-zan*, YANG Wei. Detection of Crop Chlorophyll Content Based on Spectrum Extraction from Coating Imaging Sensor[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(05): 1581-1587. |
[5] |
LIU Zi-long1,2*, SUN Li-qun1, GUO Yin1, YANG Huai-dong1*, DING Wu-wen1, ZHANG Qiao-xiang2. The Calibration Research of DOAS Based on Spectral Optical Density[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(04): 1302-1306. |
[6] |
DU Zhen-hui, ZHANG Zhe-yuan, LI Jin-yi, XIONG Bo, ZHEN Wei-meng . The Development of Ammonia Sensor Based on Tunable Diode Laser Absorption Spectroscopy with Hollow Waveguide [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(08): 2669-2673. |
[7] |
ZHANG Jian-feng1, PAN Sun-qiang1, LIN Xiao-lu1,2, HU Peng-bing1, CHEN Zhe-min1 . Research on On-Line Calibration Based Photoacoustic Spectrometry System for Monitoring the Concentration of CO2 in Atmosphere[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(01): 1-5. |
[8] |
GAO Peng-fei1, YANG Rui1, 2, JI Jiang1, GUO Han-ming1*, HU Qi1, ZHUANG Song-lin1 . Locally Dynamically Moving Average Algorithm for the Fully Automated Baseline Correction of Raman Spectrum [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(05): 1281-1285. |
[9] |
KOU Jie-ting1,2, Bayanheshig1*, TANG Yu-guo1, QI Xiang-dong1, YU Hong-zhu1 . Wavelength Calibration Research on the Automatic Grating Monochromator for Orthogonal Czerny-Turner Structure[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(04): 1127-1131. |
[10] |
WU Zheng-jie1, HUANG Yao-xiong1*, WANG Cheng2, LI Shao-fa2 . Raman Spectra Quantitative Analysis on Materials with Strong Fluorescence Background[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(07): 1798-1801. |
[11] |
WU Zheng-jie1, HUANG Yao-xiong1*, WANG Cheng2, LI Shao-fa2 . Normalization Methods for Ethanol Raman Spectra Quantitative Analysis [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(04): 971-974. |
[12] |
WU Xiao-li1, LI Yan-jun2, WU Tie-jun3* . Water Quality Analysis by Three-Dimensional Fluorescence Spectra Based on Selective Model Combination [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(04): 996-1001. |
[13] |
ZHANG Jun1,2, ZHU Yong1*, CHEN Jun-qing2, LIANG Bo1 . Study on Online Self-Calibration Technique for Trace Gas Analyzer Based on Tunable Diode Laser Absorption Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(04): 1030-1034. |
[14] |
MA Xiao-jun,GAO Dang-zhong*,FENG Jian-hong,LI Yu-hong,YE Cheng-gang,LIU Yuan-qiong . Quantitative Analysis of the Concentration of Br-Doping in Micro-Shell Coating with XRF[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(06): 1678-1681. |
[15] |
LING Xiao,BORJIHAN Ge-rel-tu*,ZHOU Hai-jun,NA Ri-su. Fluorescence Characteristics and Analytical Application of Mn(Ⅱ)-Piperine-Cetyltrimethylammonium Bromide Ternary System[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(01): 188-191. |
|
|
|
|