光谱学与光谱分析 |
|
|
|
|
|
Trace Carbon Monoxide Detection with a Cavity Ring-Down Spectrometer |
CHEN Bing1, ZHOU Ze-yi2, KANG Peng1, LIU An-wen1, HU Shui-ming1* |
1. Hefei National Laboratory for Physical Sciences at Microscale, University of Science and Technology of China, Hefei 230026, China 2. National Institute of Metrology, Beijing 100013, China |
|
|
Abstract A cavity ring-down spectrometer (CRDS) was built based on telecom distributed feedback (DFB) diode lasers. The ring down cavity was formed by mirrors of 99.997% reflectivity with a separation of 130 cm, with an empty ring down time of about 150 μs. A minimum detectable absorption coefficient of 5×10-12 cm-1 was obtained by averaging about 1 000 recorded spectra. The ring down cavity is thermo-isolated, and used as an interferometer to calibrate the recorded spectrum. A feedback control scheme was applied to step scan the laser frequency, successively matching each of the longitudinal modes of the cavity. By measuring the CO contents in a standard gas sample, the quantitative capability of the CRDS instrument was demonstrated, and a CO detection limit of 4 ppb was achieved. The instrument was applied to monitor the CO concentration in ambient air.
|
Received: 2014-03-03
Accepted: 2014-06-14
|
|
Corresponding Authors:
HU Shui-ming
E-mail: smhu@ustc.edu.cn
|
|
[1] Logan J A, Prather M J, Wofsy S C, et al. Journal of Geophysical Research, 1981, 86(C8): 7210. [2] Tracey H, Hiram L, Prasad K. Journal of Geophysical Research,2000, 105:12123. [3] Sawa Y, Tanimoto H, Yonemura S, et al. Journal of Geophysical Research: Atmospheres, 2007, 112, D22S26. [4] Ouyang Changfeng, Lin Yuchi, Lin Nenghuei, et al. Atmospheric Environment, 2009, 43, 6449. [5] Sachse G W, Hill G F, Wade L O, et al. Journal of Geophysical Research: Atmospheres, 1987, 92:2071. [6] Fried A, Henry B, Parrish D, et al. Atmospheric Environment, Part A. General Topics, 1991, 25: 2277. [7] Chen H, Karion A, Rella C W, et al. Atmospheric Measurement Techniques, 2013, 6: 1031. [8] Thiebaud J, Fttschen C. Applied Physics B: Lasers and Optics, 2006, 85:383. |
[1] |
BAI Bing1, 2, 3, CHEN Guo-zhu2, 3, YANG Wen-bin2, 3, CHE Qing-feng2, 3, WANG Lin-sen2, 3, SUN Wei-min1*, CHEN Shuang1, 2, 3*. The Study on Precise and Quantitative Measurement of Flame OHConcentration by CRDS-CARS-PLIF Techniques[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(12): 3955-3962. |
[2] |
ZHANG Le-wen1, 2, WANG Qian-jin1, 3, SUN Peng-shuai1, PANG Tao1, WU Bian1, XIA Hua1, ZHANG Zhi-rong1, 3, 4, 5*. Analysis of Interference Factors and Study of Temperature Correction Method in Gas Detection by Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 767-773. |
[3] |
LOU Deng-cheng, RAO Wei*, SONG Jun-ling, WANG Kai, JIANG Ya-jing, GUO Jian-yu. Research of Carbon Monoxide Concentration Measurement in Combustion Field by Off-Axis Integrated Cavity Output Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(12): 3678-3684. |
[4] |
LI Qing-yuan, LI Jing, WEI Xin, SUN Mei-xiu*. Performance Evaluation of a Portable Breath Isoprene Analyzer Based on Cavity Ringdown Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(08): 2415-2419. |
[5] |
XU Xing-wei1, 2, WANG Wei1*, LIU Cheng3, SHAN Chang-gong4, SUN You-wen1, HU Qi-hou1, TIAN Yuan1, HAN Xue-bing1, YANG Wei1. Observations of Total Columns of CO Based on Solar Absorption Spectra[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(05): 1329-1334. |
[6] |
ZHOU Sheng1, 2, HAN Yan-ling1, LI Bin-cheng1, 3*. Calibration Method of Pressure Gauges Based on Cavity Ring-Down Spectroscopy Technique[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(04): 1031-1035. |
[7] |
HU Chun-dong1,2, YAN Jing-yang1,2, WANG Yan1,2, LIANG Li-zhen1*. Study on Cavity Ring-Down Spectroscopy in High Power Ion Source[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2018, 38(02): 346-351. |
[8] |
ZOU Jie-shu, WANG fei*, YAN Jian-hua, CEN Ke-fa . Simulation of Concentration Measurement of SO2, NO2 and Particles Simultaneously by Differential Optical Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(12): 3830-3835. |
[9] |
ZHOU Sheng1, 2, HAN Yan-ling1, LI Bin-cheng1, 3* . Trace Moisture Measurement with 5.2 μm Quantum Cascade Laser Based Continuous-Wave Cavity Ring-Down Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(12): 3848-3852. |
[10] |
LI Chun-guang1, DANG Jing-min1, CHEN Chen2*, WANG Yi-ding1*. Multi-Pass Absorption Spectroscopy for CO Detection Using a Quantum Cascaded Laser[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(05): 1308-1312. |
[11] |
CHEN Chen1, WANG Biao4, LI Chun-guang2, LI Jian2, 3*, WANG Yi-ding2* . A Trace Gas Sensor Using Mid-Infrared Quantum Cascaded Laser at 4.8 μm to Detect Carbon Monoxide[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34(03): 838-842. |
[12] |
WU Bin1, 2, SHI Xue-shun1, 2, SUN Qing3, DENG Yu-qiang3, WANG Heng-fei2, CHEN Kun-feng2 . Frequency Correction in Terahertz Absorption Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(09): 2343-2347. |
[13] |
CHEN Xiao, SUI Qing-mei, MIAO Fei, JIA Lei, WANG Jing, JIANG Ming-shun . Research on Carbon Monoxide Multi-Parameter Detection Based on Ultra-Narrow-Linewidth Laser[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(10): 2829-2833. |
[14] |
Qiherima1, 3,YUAN Hui2,ZHANG Yun-hong1,LI Hui-feng2,XU Guang-tong2* . In Situ FTIR and XPS Study on Selective Hydrodesulfurization Catalyst of FCC Gasoline [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(07): 1752-1757. |
[15] |
CHEN Jin1, 2, DUAN Fa-jie1, TONG Ying2, GAO Qiang3 . Genetic Programming Used for the Measurement of CO Concentration Based on Nondispersive Infrared Absorption Spectroscopy [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(07): 1758-1761. |
|
|
|
|