光谱学与光谱分析 |
|
|
|
|
|
Study of Fundamental Parameters of Inductively Coupled Plasma with Slurry Nebulization ——I.Excitation Temperature Determination |
WANG Zheng1, 2,QIU De-ren2,TAO Guang-yi1,YANG Peng-yuan2* |
1. Shanghai Institute of Ceramics, Chinese Academy of Sciences, Shanghai 200050, China 2. Department of Chemistry, Fudan University, Shanghai 200433, China |
|
|
Abstract The excitation temperature of plasma with slurry nebulization into inductively coupled plasma emission spectrometry was determined and reported in the present paper. Ti-lines and multi spectrum lines method were chosen and used to determine the excitation temperature with nebulization into the inductively coupled plasma with both aqueous solution and 0.05% titanium slurry. There were approximately the same plasma excitation temperatures of 5 000-6 000 K for the two nebulization ways. The experiment verified that the excitation temperature only showed a little increase with RF power increasing. This means that the atomization efficiency did not change remarkably with slurry nebulization into the inductively coupled plasma if only the RF power was increased. Therefore, the analytical results were not much improved in the inductively coupled plasma emission spectrometry with slurry nebulization when only changing RF power.
|
Received: 2007-11-16
Accepted: 2008-02-18
|
|
Corresponding Authors:
YANG Peng-yuan
E-mail: pyyang@fudan.edu.cn
|
|
[1] Yang P Y, Barnes R M. Spectrochimica Acta Reviews, 1990, 13: 275. [2] Yang P Y, Horner J A, Sesi N N, et al. Spectrochim. Acta, Part B, 2000, 55: 1833. [3] Benson C M, Gimelshein S F, Levin D A, et al. Spectrochim. Acta, Part B, 2001, 56: 1097. [4] Horner J A, Lehn S A, Hieftje G M. Spectrochim. Acta, Part B,2002, 57: 1025. [5] Shan Y G, Mostaghimi J. Spectrochim. Acta, Part B,2003, 58: 1959. [6] WANG Zheng, CHEN Tian-yu, TAO Guang-yi, et al(汪 正, 陈天裕,陶光仪,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2005,25(4):556. [7] WANG Zheng, QIU De-ren, TAO Guang-yi, et al(汪 正, 邱德仁,陶光仪,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2006,26(3):542. [8] Sparkes S T, Ebdon L. J. Anal. Atomic Spectrom., 1988, 3: 563. [9] Ebdon L, Foulkes M E. Chimia Analityczna, 1990, 35: 109. [10] QIU De-ren(邱德仁). Analytical Atomic Spectrometry(原子光谱分析). Shanghai: Fudan University Press(上海: 复旦大学出版社), 2002. [11] Reader J, Corliss C H, Weise W L, et al. Wavelengths and Transition Probabilities for Atoms and Atomic Ions. Washington: DC, 1980. [12] Boumans P W J M. Inductively Coupled Plasma Emission Spectroscopy, Part Ⅱ, New York: Wiley., 1987. ch. 10. |
[1] |
WANG Zheng1, 2, QIU De-ren2, TAO Guang-yi1, YANG Peng-yuan2* . Study of Fundamental Parameters of Inductively Coupled Plasma for Slurry Nebulization——Electron Density Determination[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(05): 1402-1404. |
[2] |
WANG Yuan-zhong1,LI Tao2,LIU Hong-gao3*,LI Xing-kui3 . Direct Determination of Manganese in Edible Fungi Using Suspension Sample Introduction and Hypo-Sensitive Wavelength by Flame Atomic Absorption Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2007, 27(05): 1021-1023. |
|
|
|
|