Influence of Photoacoustic Cell Geometrical Shape on the Performance of Photoacoustic Spectroscopy |
CHENG Gang1, CAO Ya-nan1, TIAN Xing1, CAO Yuan2, LIU Kun2* |
1. State Key Laboratory of Mining Response and Disaster Prevention and Control in Deep Coal Mines (Anhui University of Science and Technology), Huainan 232001, China
2. Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China |
|
|
Abstract With the rapid development of modern industry, the importance of trace gas detection technology is self-evident. At present, trace gas detection technology has been widely used in environmental protection, chemical industry, bio-ecological and medical detection and other fields. Photoacoustic spectroscopy (PAS) has become one of the most important analytical methods in spectroscopic detection due to its advantages of zero background detection, no wavelength limitation of detectors, simple optical elements, convenient system regulation and maintenance. In recent years, with the rapid development of weak signal detection and laser technology, photoacoustic spectroscopy has also attracted more attention of scholars. Relevant research results provide an important design reference for improving the detection performance of photoacoustic spectroscopy. However, the current literature report rarely involves the optimization of photoacoustic cells, such as the shape of photoacoustic cells. There is a lack of in-depth exploration on such problems as the shape structure. The photoacoustic cell is one of the most important core components in the photoacoustic spectroscopy detection system. It is the cavity for carrying the gas to be tested and the place where photoacoustic coupling occurs. Its shape greatly affects the relationship between light and sound. The coupling situation affects the signal-to-noise ratio and sensitivity of the whole system. Therefore, exploring the shape of the photoacoustic pool has important theoretical significance and engineering application value. Therefore, based on the design basis of the traditional cylindrical photoacoustic cell, this paper explores and studies the structural model of the photoacoustic cell with eight typical shapes such as circle, triangle, ellipse and so on, and simulates its sound field characteristics. 3D printing technology has produced all kinds of photoacoustic cells, and analyzed the performance indexes of 8 photoacoustic cells through experiments. The longitudinal length of the photoacoustic cell designed by the constraint is equal to the perimeter of the longitudinal section. The simulation results show that the longitudinal acoustic modes of the eight photoacoustic cells are the same. The experimental results show that the acoustic resonance frequencies of the eight photoacoustic cells are the same. The size is basically the same, and is affected by the coupling of the laser source and the acoustic mode in the cavity. Their quality factors are arranged from large to small: circular, short-axis ellipse, regular pentagon, square, large circular axis, and equilateral triangle. Small circle axis shape, long axis ellipse, their cell constants are arranged in order from large to small: circular, long axis ellipse, regular pentagon, square, large circular axis shape, small circular axis shape, regular triangle, short axis ellipse . The overall results show that for the design of the photoacoustic cell in photoacoustic spectroscopy, the shape of photoacoustic cell should be circular in priority without special requirements. The research process and results of this paper can provide a reference for the optimization design of the photoacoustic cell in photoacoustic spectroscopy.
|
Received: 2019-08-02
Accepted: 2019-12-15
|
|
Corresponding Authors:
LIU Kun
E-mail: liukun@aiofm.ac.cn
|
|
[1] Miklós A, Hess P, Bozóki Z. Review of Scientific Instruments, 2001, 72(4): 1937.
[2] Nodov E. Applied Optics, 1978, 17(7): 1110.
[3] Heritier J M, Fouquet J E, Siegman A E. Applied Optics, 1982, 21(1): 90.
[4] Kapitanov V A , Ponomarev Y N, Song K, et al. Applied Physics B (Lasers and Optics), 2001, 73(7): 745.
[5] Webber M E, Pushkarsky M B, Patel C K N. Proceedings of SPIE—The International Society for Optical Engineering, 2002, 4817: 1.
[6] Kosterev A A, Bakhirkin Y A, Curl R F, et al. Optics Letters, 2002, 27(21): 1902.
[7] Ma Yufei, He Ying, Zhang Ligong, et al. Applied Physics Letters, 2017, 110(3): 031107.
[8] Haouari R, Rochus V, Lagae L, et al. Rotterdam, Netherlands, 2017 COMSOL Conference.
[9] SHI Qiang, HU Shui-ming, CHEN Jun, et al(史 强,胡水明,陈 军,等). Chinese Journal of Chemical Physics(化学物理学报), 1998, 11(1): 20.
[10] CHENG Gang, CAO Yuan, LIU Kun, et al(程 刚,曹 渊,刘 锟,等). Acta Physica Sinica(物理学报), 2019, 68(7): 074202.
[11] ZHENG De-zhong, ZHAO Nan(郑德忠,赵 南). Chinese Journal of Lasers(中国激光), 2014, 41(4): 0415002.
[12] DONG Lei, WU Hong-peng, ZHENG Hua-dan, et al(董 磊,武红鹏,郑华丹,等). Chinese Journal of Lasers(中国激光), 2018, 45(9): 0911004.
[13] ZHA Shen-long, LIU Kun, ZHU Gong-dong, et al(查申龙,刘 锟,朱公栋,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2017, 37(9): 2673.
[14] Bauer R, Stewart G, Johnstone W, et al. Optics Letters, 2014, 39(16): 4796.
[15] He Y, Ma Y F, Tong Y, et al. Optics Express, 2018, 26(8): 9666. |
[1] |
ZHENG Hong-quan, DAI Jing-min*. Research Development of the Application of Photoacoustic Spectroscopy in Measurement of Trace Gas Concentration[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(01): 1-14. |
[2] |
XU Qiu-yi1, 3, 4, ZHU Wen-yue3, 4, CHEN Jie2, 3, 4, LIU Qiang3, 4 *, ZHENG Jian-jie3, 4, YANG Tao2, 3, 4, YANG Teng-fei2, 3, 4. Calibration Method of Aerosol Absorption Coefficient Based on
Photoacoustic Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2024, 44(01): 88-94. |
[3] |
FU Wen-xiang, DONG Li-qiang, YANG Liu*. Research Progress on Detection of Chemical Warfare Agent Simulants and Toxic Gases by Photoacoustic Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(12): 3653-3658. |
[4] |
CHENG Gang1, CAO Ya-nan1, TIAN Xing1, CAO Yuan2, LIU Kun2. Simulation of Airflow Performance and Parameter Optimization of
Photoacoustic Cell Based on Orthogonal Test[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(12): 3899-3905. |
[5] |
CHEN Tu-nan1, 2, LI Kang1, QIU Zong-jia1, HAN Dong1, 2, ZHANG Guo-qiang1, 2*. Simulation Analysis and Experiment Verification of Insulating Material-Based Photoacoustic Cell[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(09): 2922-2927. |
[6] |
JIN Hua-wei1, 2, 3, WANG Hao-wei1, 2, LUO Ping1, 2, FANG Lei1, 2. Simulation Design and Performance Analysis of Two-Stage Buffer
Photoacoustic Cell[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(08): 2375-2380. |
[7] |
ZHANG Yu-feng1, WU Yu-ling2, WU Yuan-qing1*, JIA Hui2, LIU Wen-hao2, DAI Jing-min3. Effect of Surface Structure on Emissivity of Area Blackbody[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(02): 389-393. |
[8] |
LI Zhen-gang1, 2, SI Gan-shang1, 2, NING Zhi-qiang1, 2, LIU Jia-xiang1, FANG Yong-hua1, 2*, CHENG Zhen1, 2, SI Bei-bei1, 2, YANG Chang-ping1, 2. Research on Long Optical Path and Resonant Carbon Dioxide Gas
Photoacoustic Sensor[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(01): 43-49. |
[9] |
WANG Qi, WANG Shi-chao, LIU Tai-yu, CHEN Zi-qiang. Research Progress of Multi-Component Gas Detection by Photoacoustic Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(01): 1-8. |
[10] |
REN Zhong1, 2*, LIU Tao1, LIU Guo-dong1, 2. Classification and Identification of Real or Fake Blood Based on OPO Pulsed Laser Induced Photoacoustic Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(09): 2734-2741. |
[11] |
WAN Liu-jie1, 2, ZHEN Chao3, QIU Zong-jia1, LI Kang1, MA Feng-xiang3, HAN Dong1, 2, ZHANG Guo-qiang1, 2*. Research of High Precision Photoacoustic Second Harmonic Detection Technology Based on FFT Filter[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(10): 2996-3001. |
[12] |
CHEN Jie1, 2, 3, QIAN Xian-mei1, 3, LIU Qiang1, 3*, ZHENG Jian-jie1, 2, 3, ZHU Wen-yue1, 3, LI Xue-bin1, 3. Research on Optical Absorption Characteristics of Atmospheric Aerosols at 1 064 nm Wavelength[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(10): 2989-2995. |
[13] |
XIE Ying-chao1,2, WANG Rui-feng1,2, CAO Yuan1,2, LIU Kun1*, GAO Xiao-ming1,2. Research on Detecting CO2 With Off-Beam Quartz-Enhanced Photoacoustic Spectroscopy at 2.004 μm[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(09): 2664-2669. |
[14] |
WANG Qiao-yun, YIN Xiang-yu, YANG Lei, XING Ling-yu. Geometrical Optimization of Resonant Ellipsoidal Photoacoustic Cell in Photoacoustic Spectroscopy System[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(05): 1351-1355. |
[15] |
LIU Li-xian1, 2, 3, HUAN Hui-ting1, 2, Mandelis Andreas2, SHAO Xiao-peng1*. Multiple Dissolved Gas Analysis in Transformer Oil Based on Fourier Transform Infrared Photoacoustic Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(03): 684-687. |
|
|
|
|