FI-KR-FAAS Twice Gas Isolation Elution Method for the Determination of Trace Lead in Marine Biological Samples |
WANG Zhong-yuan1, ZHANG Hong-kang2*, NI Zhi-xin1, LIU Jing-qin1, CHEN Xin1, LI Sheng-yong1, YANG Qi-jin2, ZHANG Can1 |
1. South China Sea Environment Monitoring Center, State Oceanic Administration, Guangzhou 510300, China
2. College of Light Industry and Food, Zhongkai University of Agriculture and Engineering, Guangzhou 510225, China |
|
|
Abstract In this paper, trace lead in marine biological samples was determined by using flow injection ( FI ) air mixture adsorption preconcentration and twice gas isolation elution method in knotted reactor (KR) coupled with flame atomic absorption spectrometry (FAAS) . Air, complexing reagent and lead solution were mixed on-line during the preconcentration step, and the introduction of air greatly improved the adsorption effects of the lead chelate on the inner walls of KR. By the introduction of air flow before elution and adding part of air flow during elution, through these two parts of the introduction of air flow bubble partition, the dispersion of the lead chelate in KR was greatly reduced, and the peak of the atomic absorption signal was increased, and the enhancement factor (Enhancement factor, EF) was increased. By the combination with the KR of the air mixing sorption preconcentration and twice gas isolation elution method, the KR adsorption effect and elution effect reached a higher level, and compared with high back pressure and short lifetime of the micro column, KR has the characteristics such as acid and alkali resistance, low back pressure and almost unlimited lifetime. What’s more, it has no critical requirement for the peristaltic pump and is not easy to cause leakage of the solution. In the elution process, 1 s air flow was used as a spacer at the fifth second, instead of entering the air flow after the entire elution process was completed. The advantage of this method is not only being eluted completely, but also greatly reducing the dispersion of the analyte during the elution process, and the value of the absorption signal is greatly increased. If the elution time is simply put down to 5 s, then the elution will be very incomplete, the tube being not clean, a large number of analytes still remaining in the tube, and the remnants of the analyte were brought into the analysis of the next sample. All the experimental data will be inaccurate, and the value of data utilization will be lost. So complete elution is necessary and must ensure a certain elution time. As for the elution process, the elution of the analyte decreased gradually, in the past elution method, until the entire process is performed after elution, the analyte in the elution dispersion has been quite serious, which leads to the great reduction of absorption signal value. Operational parameters such as types, concentration and pH of complexing agents, sample preconcentration time and flow rate, flow rate and time of air mixing sorption process, the first ventilation time, pre elution time, the second ventilation time were optimized for the purpose of determining trace lead. Using the optimized conditions, a linear relationship existed between the peak absorbance and the concentrations of lead in the range of 0.005~0.6 mg·L-1 respectively. The method was applied in the determination of trace lead in marine biological samples with a detection limit(3σ) of 2.2 μg·L-1. In this paper, the detection results of FAAS method, traditional FI-KR-FAAS air mixing sorption preconcentration method and normal elution method were compared with this method, and it showed that the new method obtained the highest enhancement factor value. This method was applied to detect the trace lead in biological samples such as eel, pawn, slipper lobster, sturgeon, tonguefish and mussel. The trace lead content in these biological samples were among 0.34~1.92 μg·g-1. When the 1.0 μg·g-1 standard Lead solution was added, the recovery was 93.5%~96.4%, and the relative standard deviation (RSD) was 0.52%~2.94%. The method of FI-KR-FAAS twice gas isolation elution for the determination of trace lead in marine biological samples obtains several merits. It has good preconcentration effect, high precision,high degree of accuracy and high enhancement factor. This method can satisfy the analysis requirement for the marine biological samples.
|
Received: 2017-11-24
Accepted: 2018-03-21
|
|
Corresponding Authors:
ZHANG Hong-kang
E-mail: zhkuzhk@163.com
|
|
[1] WANG Fei, HUANG Xiao-ping(汪 飞,黄小平). Chinese Journal of Ecology(生态学杂志), 2012, 31(1):207.
[2] Prachi S Shenai-Tirodkar, Mangesh U Gauns, Mohammad Wassim A Mujawar, et al. Ecotoxicology and Environmental Safety, 2017, 142: 87.
[3] Fernando P Carvalho. Journal of Environmental Radioactivity, 2011, 102: 462.
[4] Aanijela Joksimovic, Slavka Stankovic. J. Serb. Chem. Soc., 2012, 77(1): 105.
[5] Dung T T T, Cappuyns V, Swennen R, et al. Rev. Environ. Sci. Biotechnol., 2013, 12: 335.
[6] FENG Liu-xing, WANG Jun(冯流星,王 军). Chinese Journal of Analytical Chemistry(分析化学), 2014, 42(4):536.
[7] Luca Bamonti, Sarah Theiner, Nataliya Rohr-Udilova, et al. JAAS, 2016, 31: 2227.
[8] GB 17378.6—2007 The Specification for Marine Monitoring(海洋监测规范). Beijing:Standards Press of China(北京:中国标准出版社), 2007.
[9] CHEN Qing-hui, WAN Yao-yu, LI Qian, et al(陈清慧,万瑶宇,李 倩,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2016, 36(10):3351.
[10] XIE Fa-zhi, ZHANG Feng-jun, XUAN Han, et al(谢发之,张峰君,宣 寒,等). Chinese Journal of Analytical Chemistry(分析化学), 2012, 40(11):1720.
[11] WANG Zhong-yuan, ZHANG Hong-kang, YU Han-sheng, et al(王中瑗,张宏康,余汉生,等). Marine Environmental Science(海洋环境科学), 2016, 35(4):618.
[12] SUN Ling-ling, SONG Jin-ming, LIN Qiang, et al(孙玲玲,宋金明,林 强,等). Marine Sciences(海洋科学), 2014, 38(8):26.
[13] Yong Liang, Xiao Wenyan, Zhao Xinli, et al. Anal. Chem.,2014, 86: 3688.
[14] Sukanta Maity, Sahu S K, Pandit G G. Soil and Sediment Contamination, 2016, 25(3): 332.
[15] Beltrán B G, Leal L O, Ferrer L, et al. Intern. J. Anal. Chem., 2016: 3095120. |
[1] |
YI Min-na1, 2, 3, CAO Hui-min1, 2, 3*, LI Shuang-na-si1, 2, 3, ZHANG Zhu-shan-ying1, 2, 3, ZHU Chun-nan1, 2, 3. A Novel Dual Emission Carbon Point Ratio Fluorescent Probe for Rapid Detection of Lead Ions[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(12): 3788-3793. |
[2] |
AN Bai-song1, 2, WANG Xue-mei1, 2*, HUANG Xiao-yu1, 2, KAWUQIATI Bai-shan1, 2. Hyperspectral Estimation of Soil Lead Content Based on Random Frog Band Selection Algorithm[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(10): 3302-3309. |
[3] |
CHAI Lin-lin, Areyi Mulati, Shawket Abliz*. Analysis the Adsorption Behaviors of Acetic Acid Modified Sand Grains for Lead Ions by Atomic Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(09): 2775-2778. |
[4] |
CHENG Fang-beibei1, 2, GAN Ting-ting1, 3*, ZHAO Nan-jing1, 4*, YIN Gao-fang1, WANG Ying1, 3, FAN Meng-xi4. Rapid Detection of Heavy Metal Lead in Water Based on Enrichment by Chlorella Pyrenoidosa Combined With X-Ray Fluorescence Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(08): 2500-2506. |
[5] |
ZHANG Chao1*, SU Xiao-yu1, XIA Tian2, YANG Ke-ming3, FENG Fei-sheng4. Monitoring the Degree of Pollution in Different Varieties of Maize Under Copper and Lead Stress[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(04): 1268-1274. |
[6] |
LAI Si-han1, LIU Yan-song1, 2, 3*, LI Cheng-lin1, WANG Di1, HE Xing-hui1, LIU Qi1, SHEN Qian4. Study on Hyperspectral Inversion of Rare-Dispersed Element Cadmium Content in Lead-Zinc Ores[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(04): 1275-1281. |
[7] |
ZHU Gui-jun, WANG Gan-zhen, PENG Jun*, TIAN Zong-ping, HOU Zhi-hua. The Mineralogical and Spectroscopic Characteristics of Phosphohedyphane From Chenzhou of Hunan Province[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(10): 3017-3023. |
[8] |
HUANG Xiao-juan1, 2, YAN Jing1, ZHANG Yang-li-zheng1, WANG Li-qin2*, XU Wei-hong1. Technological Analysis of the Chess Pieces of Liubo Uncovered From the Warring States Tomb in Shaanxi Province[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(10): 3187-3192. |
[9] |
WANG Ruo-su1, SUN Feng1, 2*, XIAN Yi-heng1. Study on the Composition and Weathered Layer of a Silicate Bead
Excavated From Ma-Jia-Yuan Cemetery M21[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(10): 3193-3197. |
[10] |
HOU Ya-ru, LU Ji-long*, FAN Yu-chao, Abudusalamu·KADIER, TANG Xiao-dan, WEI Qiao-qiao, GUO Jin-ke, ZHAO Wei. Uncertainty Evaluation and Method Improvement of Determination of Copper, Lead, and Zinc in Rocks by Atomic Absorption Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(07): 2101-2106. |
[11] |
FAN Chun-hui1, 2, ZHENG Jin-huan3, LIU Hong-xin1. FTIR, 2D-IR and XPS Analyses on the Mechanism of Protoplast Derived From Calendula Officinalis in Response to Lead and Cadmium in Soil[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1420-1425. |
[12] |
ZHENG Pei-chao, ZHONG Chao, WANG Jin-mei*, LUO Yuan-jiang, LAI Chun-hong, WANG Xiao-fa, MAO Xue-feng. Evaluation of Flow Injection-Solution Cathode Glow Discharge With an Interference Filter Wheel for Spectral Discrimination[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(03): 842-847. |
[13] |
LUO Wei, TIAN Peng, DONG Wen-tao, HUANG Yi-feng, LIU Xue-mei, ZHAN Bai-shao, ZHANG Hai-liang*. Detection of Pb Contents in Soil Using LIBS Coupled With Univariate Calibration Curve Methods[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(03): 886-891. |
[14] |
LIU Qun-wei, SHAO Yong-ni, CHEN Xiao-wan, JIANG Lin-hua*. Study on the Changes of Material in Scenedesmus Obliquus Based on Fourier Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(12): 3890-3894. |
[15] |
LI Chao1, 2, LI Bin2, 3*, ZHANG Li-qiong2, YE Da-peng1*, ZHENG Shu-he1. Terahertz Spectrum Inversion Modeling of Lead Content in Different pH Soils[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(08): 2397-2402. |
|
|
|
|