Abstract In order to meet the needs of high-precision measurements of atmospheric trace constituents, accurate description the slitfunction of the detector for atmosperic element measurements is required. Considering the characteristic of large field, wide wavelength range, high spatial and spectral resolution, a slitfunction measurement stimulus has been developed. Firstly the theory of slitfunction measurement was supplied. The slitfunction measurement can output high-resolution and wide spectrum and the slit function of the detector for atmosperic element measurements was tested by building an experimental device, The mathematical expression was offered accurately, and the results were analyzed. The results show that: the spectral resolution is about 0.423~0.597 nm of the detector for hyperspectral atmospheric element measurements, and the slitfunction approximating Gaussian distribution. Since the large field of the detector for hyperspectral imaging instrument, a phenomenon of smiling was appealed. This result led to the spectral resolution in the edge of the field slightly below that in the center of the field. For accurate and complete measurement of the slitfunction an echelle monochromator has been developed, which offers a number of high-resolution spectral lines, that can be used to spectral calibration.
|
Received: 2015-12-02
Accepted: 2016-04-26
|
|
Corresponding Authors:
WANG Yong-mei
E-mail: wym@nssc.ac.cn
|
|
[1] WANG Yue-ming, LANG Jun-wei, WANG Jian-yu(王跃明, 郎均慰, 王建宇). Laser & Optoelectronics Progress(激光与光电子学进展), 2013, 50(1): 010008.
[2] ZHANG Da, ZHENG Yu-quan(张 达, 郑玉权). Optics & Optoelectronic Technology(光学与光电技术), 2013, 11(3): 67.
[3] XUE Qing-sheng(薛庆生). Acta Optica Sinica(光学学报), 2014, 8(34): 0822005.
[4] CHEN Wei, ZHENG Yu-quan, XUE Qing-sheng(陈 伟, 郑玉权, 薛庆生). Acta Optica Sinica(光学学报), 2014, 10(43): 1022001.
[5] CHEN Wei, ZHENG Yu-quan, XUE Qing-sheng(陈 伟, 郑玉权, 薛庆生). Acta Photonica Sinica(光子学报), 2014, 10(43):1022001(1).
[6] XUE Qing-sheng(薛庆生). Acta Optica Sinica(光学学报), 2013, 3(33): 0322001.
[7] ZHOU Bin, LIU Wen-qing, QI Feng, et al(周 斌, 刘文清, 齐 峰, 等). Acta Physica Sinica(物理学报), 2001, 9(50): 1818.
[8] LI Zhan-feng, WANG Shu-rong, HUANG Yu, et al(李占峰, 王淑荣, 黄 煜, 等). Acta Optica Sinica(光学学报), 2013, 2(33): 0228002.
[9] ZHENG Yu-quan(郑玉权). Optics and Precision Engineering(光学 精密工程), 2010, 18(11): 2347.
[10] Dirksen R, Dobber M, Voors, et al. Appl. Opt., 2006, 45(17): 3972.
[11] Dirksen R, Dobber M, Levelt P, et al. Proc. SPIE, 2004, 5234: 400.
[12] de Vries J, van den Oord G H J, Hilsenrath E, et al. Proc. SPIE, 2002, 4480: 315.
[13] Smorenburg K, Dobber M, Schenkeveld E, et al. Proc. SPIE, 2003, 4881: 511.
[14] Dobber M R, Dirksen R J, Levelt P F, et al. IEEE Transactions on Geoscience and Remote Sonsing, 2006, 44 (5): 1209. |
[1] |
CHENG Liang-xiao1, 2, TAO Jin-hua1*, ZHOU Hai-jin3, YU Chao1, FAN Meng1, WANG Ya-peng4, WANG Zhi-bao5, CHEN Liang-fu1. Evaluations of Environmental Trace Gases Monitoring Instrument (EMI) Level 1 Data[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3881-3886. |
[2] |
HUANG Shan1, 2, SI Fu-qi1*, ZHAO Min-jie1, ZHOU Hai-jin1, JIANG Yu1. Study on the Slit Function of Atmospheric Trace Gas Differential Optical Absorption Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2019, 39(07): 2008-2012. |
[3] |
LI Liu-cheng, WANG Zeng-qiang, LI Guo-fu, DUO Li-ping* . Study of Plasma Temperature Measurements for Oxygen Discharge [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(10): 2651-2654. |
|
|
|
|