光谱学与光谱分析 |
|
|
|
|
|
A Research on Filed-of-View (FOV) Widening and Thermal-Phase-Drift (TPD) Compensating Technology Applied in a Polarized Interference Imaging Spectrometer (PIIS) |
ZHAI Yang1, 2, XIAO Dong1, 2, LI Bo1, 2, ZHU Ri-hong3 |
1. National Astronomical Observatories/Nanjing Institute of Astronomical Optics & Technology, Chinese Academy of Sciences, Nanjing 210042, China 2. Key Laboratory of Astronomical Optics & Technology, Nanjing Institute of Astronomical Optics & Technology, Chinese Academy of Sciences, Nanjing 210042, China 3. School of Electronic Engineering and Photoelectric Technology, Nanjing University of Science & Technology, Nanjing 210094, China |
|
|
Abstract The Polarized Interference Imaging Spectrometer (PIIS), which is based on the theory of Fourier Transform Spectroscopy, is consisted of a series of birefringent crystals such as polarizers, a beam splitter as well as various lengths of birefringent crystals required to achieve large delays. The PIIS, compared with a traditional grating-based dispersion spectrometer, has various advantages of multiple-channel measurements, simultaneous information acquisition of both original images and fringes containing spectral details, large light flux, better light signal-to-noise ratio (SNR) as well as anti-vibration etc. Therefore, the PIIS has also been developed in a range of astronomy and astrophysics areas such as remote sensing, extrasolar planet radial velocity measurements, spacecraft design, lunar exploration etc. However, by reviewing of former works and references, two major drawbacks still remain in PIIS and need to be fixed. For one thing, the classic PIIS has a very limited field-of-view (FOV) around ±2°, which means the acquired fringes on the image plane will show quite strong non-linear distortion and hence degrade the accuracy of spectral reconstruction via Fourier transform. For another, the random thermal-phase-drift (TPD), given rise from both thermal expansion and birefringence variation caused by the environmental temperature fluctuation, has barely been studied before and will inevitably result in extra radial velocity error based on Doppler Spectroscopy. In this paper, a noble polarization interference imaging spectrometer with the emphasis on the FOV widening technology is introduced. This technology, using a compensated Savart plate containing a half-wave plate sandwiched between two orthogonally placed displacer plates as a compensated Savart plate, produces an angle-dependent phase shear to create parallel spatial interference fringes with a FOV around ±10°. This improvement not only enhances the accuracy of Fourier Transform algorithm but also increases input luminous flux and therefore even weak input spectrum detection and calibration results with high SNR can be fully accomplished. Also, a secondary set of birefringent plates (α-BBO and LiNbO3) with opposite thermal properties is proposed to passively diminish TPD caused by temperature fluctuation. The experiment shows that thermal-drift-phase error is perfectly restricted within 0.02 rad in the laboratory environment. As a consequence, this advanced PIIS is eligible to realize the fast and accurate measurement and calibration application in the field of large astronomical spectral instruments with ultra-high spectral resolution occasions such as Astronomical Frequency Comb.
|
Received: 2015-06-11
Accepted: 2015-10-28
|
|
Corresponding Authors:
ZHAI Yang
E-mail: yzhai@niaot.ac.cn
|
|
[1] Casasent David, Psaltis Demetri. Applied Optics, 1980, 19(12): 2034. [2] Smith W L, Zhou D K, Harrison F W, et al. Proc. SPIE, 2001, 4151: 94. [3] MU Ting-kui, ZHANG Chun-min, LI Qi-wei,et al(穆廷魁, 张淳民, 李祺伟,等). Acta Phys. Sin.(物理学报), 2014, 63(11): 11704-1. [4] ZHANG Da-peng, LIANG Zhi-guo, YE Peng(张大鹏,梁志国,叶 蓬). Metrology & Measurement Technology(计测技术), 2014, 34(3): 43. [5] JIANG Ming-da, XIAO Dong, ZHU Yong-tian(姜明达,肖 东,朱永田). Progress in Astronomy(天文学进展), 2012, 30(2): 246. [6] ZHAO Fei, WANG Hui-juan, ZHAO Gang,et al(赵 斐,王汇娟,赵 刚,等). Astronomical Research & Technology(天文研究与技术), 2015, 12(1): 117. [7] WU Xue-jian, LI Yan, WEI Hao-yun,et al(吴学健,李 岩,尉昊赟,等). Laser & Optoelectronics Progress(激光与光电子进展), 2012, 49(3): 1. [8] Murphy M T, Udem T, Holzwarth R, et al. Monthly Notices of the Royal Astronomical Society,2007. 839. [9] John Howard, Jinil Chung. Review of Scientific Instruments, 2012, 83(10): 10. [10] John Howard. J. Phys. B: At. Mol. Opt. Phys., 2010, 43(14): 4075. [11] WU Lei, ZHANG Chun-min, YUAN Yan,et al(吴 磊, 张淳民, 袁 艳,等). Acta Optica Sinica(光学学报), 2005, 25(7): 885. [12] Zhou Xuetie, Li Yan, Wang Junde, et al. IEEE Trans. Plasma Science,2001,29(2): 360. [13] PENG Zhi-hong, ZHANG Chun-min, ZHAO Bao-chang(彭志红,张淳民,赵葆常). Acta Physica Sinica(物理学报), 2006, 55(12): 6374. |
[1] |
XIE Pei-yue1, 2, YANG Jian-feng1, XUE Bin1*, Lü Juan1, HE Ying-hong1, LI Ting1, MA Xiao-long1 . Research on an Equal Wavelength Spectrum Reconstruction Method of Interference Imaging Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2016, 36(03): 848-852. |
[2] |
WANG Shuang1, 2, XIANGLI Bin3, LI Li-bo1, 2, PI Hai-feng1, 2 . Research of Spectrum Signal-to-Noise Ratio of Large Aperture Static Imaging Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34(03): 851-856. |
[3] |
WANG Xiang1,2, CHEN Jian-ping1,2*, LI Jian-feng1,2, SHI Rui1,2, WU Zhao1,2 . Information Extraction of the Chang’E-1 Interference Imaging Spectrometer (IIM) 2C Data [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(04): 1145-1150. |
[4] |
FU Qiang1,2, XIANGLI Bin3, Lü Qun-bo3, JING Juan-juan1,2 . Design of a Dual-Channel Mach-Zehnder Lateral Shearing Interferometer for the Large Aperture Static Imaging Spectrometer [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(02): 553-557. |
[5] |
TANG Yi, TANG Qiu-yan, WANG Xiao-yan, CAO Wei-liang, WANG Jing . Simulation and Analysis of Interference Imaging Spectrometer Influenced by Satellite Vibration[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(11): 3149-3153. |
[6] |
CHEN Chao, QIN Qi-ming*, ZHANG Ning, WANG Jin-liang, LI Jun, JIANG Hong-bo . Research on Photometric Calibration and Reflectance Calculation of CE-1 IIS Data[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(07): 1985-1990. |
[7] |
ZHANG Xiu-bao1, YUAN Yan1, JING Juan-juan2,3, SUN Cheng-ming1, WANG Qian1 . Spectral Discrimination Method Information Divergence Combined with Gradient Angle [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(03): 853-857. |
[8] |
YAO Tao1,4, YIN Shi-min2, XIANGLI Bin3,Lü Qun-bo4 . Techniques for Pixel Response Nonuniformity Correction of CCD in Interferential Imaging Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(06): 1712-1716. |
[9] |
XIANGLI Bin1, Lü Qun-bo1, HUANG Min1, 3, YAO Tao2, 3 . Comparison of Two Types of Fourier Transform Imaging Spectrometry [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(05): 1422-1426. |
[10] |
HUANG Min1,2, XIANGLI Bin1,3, YUAN Yan4, SHEN Zhong5, Lü Qun-bo1,2,WANG Zhong-hou1, LIU Xue-bin1 . Spatially Modulated Fourier Transform Imaging Spectrometer Data Compression Research[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2010, 30(01): 278-282. |
|
|
|
|