光谱学与光谱分析 |
|
|
|
|
|
A New Method for Eliminating Background Signal Drift to Improve the Detection Precision in Continuous Harmonic Detection |
ZHAO Ying1, ZHAO Xue-hong1,2*, WANG Zhe1, ZHANG Rui1,3, WANG Yan1 |
1. Collegy of Precision Instrumonts and Opto-Electronics Eningeering, Tianjin University, Tianjin 300072, China 2. Tianjin Key Laboratory of Biomedical Detecting Techniques & Instruments, Tianjin 300072,China 3. College of Electronic Information and Automation, Tianjin University of Science and Technology, Tianjin 300222, China |
|
|
Abstract To overcome the second harmonic background signal drift in gas continuous detection based on tunable diode laser absorption spectroscopy(TDLAS), a new method of background elimination was proposed by changing center current of the laser. This method can eliminated the effects of background signal drifting on concentration inversion. Based on the wavelength modulation theory, the theoretical expression of the second harmonic of the background signal is derived. In addition,the second harmonic background signal components and the factors that affect it are described. In different working temperatures, the relationship curve between thelaser current and output light intensity. In the process of continuous detection, we analyzes the feasibility ofchangingthe laser center current extraction of background signal. Combined with the principle of the background signal searching, the LabView flow chart was designed to search background center current. Ammonia(NH3) was detected by the TDLAS experimental system, which demonstrated the feasibility of the new method. The range of experiment parameters and the range of background center current searches were determined, in the situation that there was only one absorption line in the full laser workspace. The experimental results showed that this method realizes background signal extraction, reduces the error of concentration inversion and the effects of concentration inversion bybackground signal. Thus it improved the detection accuracy of the concentration. In the continuous detection experiment, the standarddeviation of inversion concentration reduced from 2.688 3 to 1.856 1, which demonstrated that the degree of dispersion of detected concentration is reduced, even eliminate the effects of background signal drifting on concentration inversion and the accuracy of detection is improved. This method provides an effective background drifting elimination approach for improving the detection accuracy of the concentration.
|
Received: 2014-08-19
Accepted: 2014-11-28
|
|
Corresponding Authors:
ZHAO Xue-hong
E-mail: zhaoxh@tju.ecu.cn
|
|
[1] TAO Bo, YE Jing-feng, ZHAO Xin-yan, et al(陶 波, 叶景峰, 赵新艳, 等). Chinese J. Laser(中国激光), 2011, 38(12): 1215002. [2] GAO Nan, DU Zhen-hui, QI Ru-bin, et al(高 楠,杜振辉,齐汝宾,等). Acta Optica Sinica(光学学报), 2012, 32(4): 0430004. [3] HE Ying, ZHANG Yu-jun, WANG Li-ming, et al(何 莹,张玉钧,王立明,等). Optical Technique(光学技术), 2012, 38(4): 421. [4] Gustafsson J, Chekalin N, Axner Q. Spectrochimica Acta Part B, 2003, 58(1): 123. [5] Ruxton K, Chakraborty A L, Johnstone W,et al. Sensors and Actuators B Chemical,2010, 150(1): 367. [6] Reid J, El-Sherbiny M, Garside B K. Appl. Optica Acta,1980, 19(19): 3349. [7] Wu S Q, Kimishima T, Hiroaki K, et al. Japanese Journal of Applied Physics, 2000, 39(7A): 4034. [8] Werle P W, Mazzinghi P, D’Amato F, et al. Spectrochimica Acta Part A,2004, 60(8-9): 1685. [9] GUO Yuan, ZHAO Xue-hong, ZHANG Rui, et al(郭 媛, 赵学虹,张 锐,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2013, 33(8):2172. [10] HU Ya-jun, ZHAO Xue-hong, ZHANG Rui, et al(胡雅君, 赵学虹,张 锐,等). Acta Optica Sinica(光学学报), 2013, 33(11):288. [11] Schilt S, Thevenaz L,Robert P. Appl. Opt., 2003, 42(33): 6728. [12] Ghafouri-Shiraz H. Distributed Feedback Laser Diodes and Optical Tunable Filters. Jone Wiley & Sons Ltd, 2003. [13] Kevin Duffin, Andrew James McGettrick, WalterJohnstone, et al. Journal of Lightwave Technology,2007, 25(10): 3114. |
[1] |
ZENG Si-xian1, REN Xin1, HE Hao-xuan1, NIE Wei1, 2*. Influence Analysis of Spectral Line-Shape Models on Spectral Diagnoses Under High-Temperature Conditions[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(09): 2715-2721. |
[2] |
LIANG Wen-ke, WEI Guang-fen, WANG Ming-hao. Research on Methane Detection Error Caused by Lorentzian Profile Approximation[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(06): 1683-1689. |
[3] |
PENG Wei, YANG Sheng-wei, HE Tian-bo, YU Ben-li, LI Jin-song, CHENG Zhen-biao, ZHOU Sheng*, JIANG Tong-tong*. Detection of Water Vapor Concentration in Sealed Medicine Bottles Based on Digital Quadrature Phase-Locked Demodulation Algorithm and TDLAS
Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 698-704. |
[4] |
ZHANG Le-wen1, 2, WANG Qian-jin1, 3, SUN Peng-shuai1, PANG Tao1, WU Bian1, XIA Hua1, ZHANG Zhi-rong1, 3, 4, 5*. Analysis of Interference Factors and Study of Temperature Correction Method in Gas Detection by Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(03): 767-773. |
[5] |
WANG Yi-hong, ZHOU Bin*, ZHAO Rong, WANG Bu-bin. Calibration-Free Wavelength Modulation Spectroscopy for Gas Properties Measuring Basedon 2nd and 4th Harmonics[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2023, 43(02): 368-373. |
[6] |
ZHANG Bo-han, YANG Jun, HUANG Qian-kun, XIE Xing-juan. Research on Gas Pressure Measurement Method Based on Absorption Spectroscopy and Laser Interference Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(12): 3692-3696. |
[7] |
LONG Jiang-xiong1, 2, ZHANG Yu-jun1*, SHAO Li1*, YE Qing1, 2, HE Ying3, YOU Kun3, SUN Xiao-quan1, 2. Traceable Measurement of Optical Path Length of Gas Cell Based on Tunable Diode Laser Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(11): 3461-3466. |
[8] |
LI Cong-cong1, LUO Qi-wu2, ZHANG Ying-ying1, 3*. Determination of Net Photosynthetic Rate of Plants Based on
Environmental Compensation Model[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1561-1566. |
[9] |
LIU Jiang-qing1, YU Chang-hui2, 3, GUO Yuan2, 3, LEI Sheng-bin1*, ZHANG Zhen2, 3*. Interaction Between Dipalmityl Phosphatidylcholine and Vitamin B2
Studied by Second Harmonic Spectroscopy and Brewster Angle
Microscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42(05): 1484-1489. |
[10] |
CHEN Hao1, 2, JU Yu3,HAN Li1. Research on the Relationship Between Modulation Depth and Center of High Order Harmonic in TDLAS Wavelength Modulation Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3676-3681. |
[11] |
CHEN Yang, DAI Jing-min*, WANG Zhen-tao, YANG Zong-ju. A Near-Infrared TDLAS Online Detection Device for Dissolved Gas in Transformer Oil[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(12): 3712-3716. |
[12] |
WANG Guo-shui1, GUO Ao2, LIU Xiao-nan1, FENG Lei1, CHANG Peng-hao1, ZHANG Li-ming1, LIU Long1, YANG Xiao-tao1*. Simulation and Influencing Factors Analysis of Gas Detection System Based on TDLAS Technology[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(10): 3262-3268. |
[13] |
CHEN Hao1, 2, JU Yu3, HAN Li1, CHANG Yang3. Curve Fitting of TDLAS Gas Concentration Calibration Based on Relative Error Least Square Method[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(05): 1580-1585. |
[14] |
HUANG An1, 2, XU Zhen-yu1, XIA Hui-hui1, YAO Lu1, RUAN Jun1, HU Jia-yi1, ZANG Yi-peng1, 2, KAN Rui-feng1*. Measurement Method of Two-Dimensional Distribution of Temperature and Components in Gas Turbine Combustor Based on Wavelength Modulated Absorption Spectroscopy[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(04): 1144-1150. |
[15] |
ZOU Li-chang1, HUANG Jun4, LI Zheng-hui1, DENG Yao1, SHAO Guo-dong1, RUAN Zhen4, LU Zhi-min1, 2, 3, YAO Shun-chun1, 2, 3*. Research on Correction Method of Background Signal Drift in Mid-Infrared Harmonic Detection[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2021, 41(02): 408-413. |
|
|
|
|