光谱学与光谱分析 |
|
|
|
|
|
Moving Mirror Scanning System Based on the Flexible Hinge Support |
XIE Fei, FENG Fei, WANG Fu-bei, WU Qiong-shui*, ZENG Li-bo* |
School of Electronic Information, Wuhan University, Wuhan 430079,China |
|
|
Abstract In order to improve moving mirror drive of Fourier transform infrared spectrometer, we design a dynamic scanning system based on flexible hinge support. Using the flexible hinge support way and the voice coil motor drive mode. Specifically, Using right Angle with high accuracy high stability type flexible hinge support mechanism support moving mirror, dynamic mirror can be moved forward and backward driven by voice coil motor reciprocating motion, DSP control system to control the moving mirror at a constant speed. The experimental results show that the designed of moving mirror scanning system has advantages of stability direction, speed stability, superior seismic performance.
|
Received: 2014-09-02
Accepted: 2014-12-05
|
|
Corresponding Authors:
WU Qiong-shui, ZENG Li-bo
E-mail: qswu@whu.edu.cn; lbzeng@whu.edu.cn
|
|
[1] Tatsuyuki Tanahashi, Masaya Toda, Hidetoshi Miyashita, et al. IEEE Xplore,2013 W4D. 007. [2] Sun Xiaojie, Mao Jianhua, Dai Zuoxiao, et al. Optical System Technologies for Manufacturing and Testing,2012,doi: 10.1117/12.975181. [3] WENG Shi-fu(翁诗甫). Fourier Transform Infrared Spectroscopy Analysis(傅里叶变换红外光谱分析),2010. 62. [4] LIU Hua, LIU Wei-qi, FENG Rui, et al(柳 华,刘伟奇,冯 睿,等). Infrared and Laser Engineering(红外与激光工程), 2012, 41(1). [5] YANG Xue-feng, LI Wei,WANG Yu-qiao, et al(杨雪锋, 李 威, 王禹桥, 等). Nano Technology and Precision Engineering(纳米技术与精密工程), 2009, 7(4). [6] Milan Milosevic. Internal Reflection and Atr Spectroscopy: John Wiley&Sons,2012. [7] Vlikyl T, Kauppinen J. Applied Optics, 2011, 50(36): 6671. [8] LIU Ri-long, YIN De-kui(刘日龙,殷德奎). Infrared(红外),2009, 30(9). [9] Bernhardt B, Sorokin E, Jacquet P, et al. Applied Physics B, 2010, doi: 10.1007/s00340-010-4080-0. |
[1] |
LIANG Xiao-wen1, SHI Lei2*. Design of a Moving Mirror Scanning System for Portable Interferometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(10): 3255-3259. |
[2] |
DING Fan, ZENG Li-bo, WU Qiong-shui*. Automatic Regulating System Design for Fourier Infrared Spectrometer Beam Splitter[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2017, 37(09): 2937-2942. |
[3] |
XIE Fei, WU Qiong-shui*, ZENG Li-bo* . Biological Process Oriented Online Fourier Transform Infrared Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2015, 35(08): 2357-2361. |
[4] |
LIU En-chao, LI Xin, ZHANG Yan-na, LI Wen-wei, ZHENG Xiao-bing . Spectral Scanning Measurement of High-Precision Solar Irradiance-Meter [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34(03): 609-613. |
[5] |
REN Li-bing, YANG Hong-lei, WEI Hao-yun, LI Yan . The Realization of Moving Mirror Scanning in FTIR Spectrometer Using Completely Digital Control Method [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(08): 2263-2266. |
[6] |
SHI Lei, LI Kai, GAO Zhi-fan, ZENG Li-bo, WU Qiong-shui* . Design of A Compact Structure Interferometer [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2013, 33(08): 2294-2298. |
[7] |
LI Zhong-bing, XU Xian-ze*, LE Yi, XU Feng-qiu, LI Jun-wei. Motion Control of Moving Mirror Based on Fixed-Mirror Adjustment in FTIR Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(08): 2295-2298. |
[8] |
SHI Lei, LIU Jia, ZHENG Xin-wen, ZENG Li-bo, WU Qiong-shui*. An Interferometer Design for Fourier Transform Infrared Spectrometer[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(06): 1712-1716. |
[9] |
ZHANG Min-juan2, ZHANG Ji-long1, 2*, WANG Zhi-bin1, 2, JING Ning2, HAO Jian2 . Phase Correction Technology Research and Improvement Based on Single-Sided Interferograms in FTIR [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2012, 32(05): 1203-1208. |
[10] |
LI Sheng, GAO Min-guang, ZHANG Yu-jun, LIU Wen-qing, XU Liang, TONG Jing-jing, CHENG Si-yang, JIN Ling,WEI Xiu-li, WANG Ya-ping, CHEN Jun. Monitoring and Analysis of Urban Ozone Using Open Path Fourier Transform Infrared Spectrometry[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(12): 3256-3259. |
[11] |
FU Jian-guo1,2,LIANG Zhong-zhu1,LIANG Jing-qiu1* . Influence of Extended Light Source on Spectral Reconstruction in a Micro FTIR [J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31(06): 1723-1726. |
[12] |
LIU Shi-yuan1,ZHANG Chuan-wei2,SHEN Hong-wei2,GU Hua-yong2 . Model-Based FTIR Reflectometry Measurement System for Deep Trench Structures of DRAM[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2009, 29(04): 935-939. |
[13] |
LIU Zhi-ming,GAO Min-guang,LIU Wen-qing,LU Yi-huai,ZHANG Tian-shu,XU Liang,WEI Xiu-li. The Study of Instrument Response Function of FTIR Detectors[J]. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2008, 28(08): 1786-1789. |
|
|
|
|