1. Key Laboratory of Environmental Optics and Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China 2. University of Science and Technology of China, Hefei 230026, China
Abstract:Fourier transform infrared spectrometer can be realized in high temperature flue gas multicomponent measurement at the same time, has wide application prospects in the field. And one of the important factors to determine the success of application, lies in the measuring system of infrared interference figure sampling phase error control. This paper discusses the main reasons of the appearance of phase error in the system, through the analysis of Helium-neon laser interference signal zero uniformity, illustrates the produce phase error is the main reason of the laser signal and reference signal phase difference. At the meantime, the quantitative analysis of the phase error influence on instrument signal to noise ratio (SNR), also the Mertz phase correction method for the instrument improves the thousands of times of the original signal to noise ratio. And the related experiment, the experimental results show that the system based on the interference figure sampling method satisfy the needs of high temperature flue gas measurements.
Key words:Fourier transform spectroscopy;Helium-neon laser sampling;High temperature flue gas;Phase correction;Signal-to-noise ratio
[1] Qian Weikang,Mei Junhui,Qian Jianqiu. Applied Mechanics and Materials, 2013,421: 468. [2] XIE Pin-hua, LIU Wen-qing, ZHENG Zhao-hui, et al(谢品华,刘文清,郑朝辉,等). Acta Photonica Sinica(光子学报),2000,29(3): 271. [3] CHEN Dai-sheng, ZHANG Li-jun, BAO Jian-guo(陈代胜,张莉君,鲍建国). Envitonmental Science and Technology(环境科学与技术),2007, 30(9): 50. [4] LIU Zhi-ming, GAO Min-guang, ZHANG Tian-shu, et al(刘志明,高闽光,张天舒,等). Infrared Technology(红外技术),2007, 29(4): 243. [5] Schutze C, Lau S, Reiche N, et al. Energy Procedia,2013, 37: 4276. [6] ZHANG Min-juan, ZHANG Ji-long, WANG Zhi-bin, et al(张敏娟,张记龙,王志斌,等). Spectroscopy and Spectral Analysis(光谱学与光谱分析),2012, 32(5): 1203. [7] Peter R Griffiths, James A de Haseth. Fourier Transform Infrared Spectrometry, Second Edition. John Wiley & Sons, 2007. 56. [8] Lawrence Mertz. Correction of Phase Errors in Interferograms Applied Optics, 1963, 12(2): 1332. [9] XING Ting, WANG Mo-chang, GONG Hui-xing(刑 廷,王模昌,龚惠兴). Acta Optica Sinica(光学学报),1999,19(3): 355. [10] Robert Furstenberg, Jeffrey O. Vibrational Spectroscopy, 2006, 42: 226. [11] Michael L Forman. Spectroscopy Optical Sciety of America,1966, 56(1): 56. [12] Avishai Ben-David,Agustin Ifarraguerri. Optical Society of America,2002, 41(6): 1181. [13] LI Sheng, GAO Min-guang, ZHANG Yu-jun, et al(李 胜,高闽光,张玉钧,等). Journal of Atmospheric and Environmental Optica(大气与环境光学学报), 2011, 6(3): 224. [14] Griffiths P R. Science, 1983, 222(4621): 297.