Abstract:To improve the quality of laser-induced breakdown spectroscopy, flat-mirror device was proposed. The effects of flat-mirror device on the radiation characteristics of laser-induced plasma were studied. The experimental results showed that when the device consisted of three flat-mirrors placed around the plasma, the spectral line intensity of Mg, Fe, Ba, Ti and Al increases by about 116.2%, 96.43%, 90.93%, 102.1% and 98.57% than that without flat-mirror device, and the signal-to-noise raises by around 39.17%, 32.48%, 38.07%, 39.95% and 21.30%,respectively. By measuring the plasma parameters,the mechanism of the radiation enhancement obtained with the device consisting of three flat-mirrors was explained. This method was an effective way to improve the detection capacity of LIBS.
Key words:Laser-induced plasma;Flat-mirror device;Spectral line intensity;SNR;Electron temperature;Electron density
[1] Galmed A H, Kassem A K, Von Bergmann H, et al. Appl. Phys. B, 2011, 102: 197. [2] Gupta G P, Suri B M, Verma A, et al. Alloys Compd., 2011, 509(9): 3740. [3] Dyar M D, Tucker J M, Humphries S, et al. Spectrochim. Acta, Part B, 2011, 66(1): 39. [4] Hamzaoui S, Khleifia R, Jaidane N, et al. Lasers Med. Sci., 2011, 26: 79. [5] Pandhija S, Rai N K, Rai A K, et al. Appl. Phys. B, 2010, 98: 231. [6] Popov A M, Colao F, Fantoni R. J. Anal. At. Spectrom., 2010, 25: 837. [7] Ermalitskaia K F, Voropay Y S, Zajogin A P. Journal of Applied Spectroscopy, 2010, 77(2): 153. [8] Guo L B, Hu W, Zhang B Y, et al. Optics Express, 2011, 19(15): 14067. [9] CHEN Jin-zhong, ZHANG Lin-jing, SUN Jiang,et al(陈金忠,张琳晶,孙 江,等). Chinese Science Bulletin(科学通报),2011, 56(4-5): 299. [10] Shen X K, Sun J, Ling H, et al. Appl. Phys. Lett., 2007, 91: 081501; http://dx.doi.org/10.1063/1.2770772. [11] Guo L B, Zhang Y B, He X N, et al. Optics Express, 2012, 20(2): 1436. [12] Pandhija S, Rai N K, Rai A K, et al. Appl. Phys. B, 2010, 98(1): 231. [13] XIN Ren-xuan(辛仁轩). Plasma Emission Spectral Analysis(等离子体发射光谱分析). Beijing: Chemical Industry Press(北京: 化学工业出版社), 2010. [14] NIST Atomic Spectra Database, http://physics.nist.gov. Kurucz output Atomic Spectral Line database from R.L.Kurucz’s CD-ROM 23.