Abstract:Fourier transform infrared spectroscopy(FTIR)can be used for the continuously on-line quantitative determination of the concentrations of several gas components in smoke gas. The process of acquiring accurate calibration of spectra is one of the most important steps in the quantitative analysis. In the present paper, a novel calibration technique was proposed and evaluated by the results of CO and HCl. Data obtained by this new technique were compared to those of conventional concentration-based calibration. The results showed that this calibration technique is feasible, and its reliability and validity were also confirmed. Further more, it has several other advantages, which will also be discussed here. Finally, the calibration problem that the substances are liquid at ambient temperature was solved by this technique.
连晨舟,吕子安,季春生,李定凯,徐旭常 . 一种标定新方法在FTIR定量分析中的应用[J]. 光谱学与光谱分析, 2004, 24(11): 1323-1326.
LIAN Chen-zhou, Lü Zi-an, JI Chun-sheng, LI Ding-kai, XU Xu-chang. A Novel Calibration Technique for Quantitative Analysis of FTIR Spectra . SPECTROSCOPY AND SPECTRAL ANALYSIS, 2004, 24(11): 1323-1326.
[1] Duquesne Sophie,Le Bras Michel. Journal of Fire Sciences,2000,18(6):456. [2] Marsanich Katia et al. Thermochimica Acta,2002, 390:153. [3] Maciejewski M et al. Thermochimica Acta,1997, 295:167.