Achievement of the Noninvasive Measurement for Human Blood Glucose with NIR Diffusion Reflectance Spectrum Method
ZHANG Hong-yan1, DING Dong2, SONG Li-qiang3, GU Lin-na1, YANG Peng1, TANG Yu-guo1
1. Changchun Institute of Optics and Fine Mechanic and Physics, Chinese Academy of Sciences, Changchun 130021, China 2. Jilin University,Changchun 130012, China 3. Yunnan Normal University, Kunming 650092, China
Abstract:The noninvasive measurement of human blood glucose was achieved with NIR diffusion reflectance spectrum method. The thumb fingertip NIR diffusion reflectance spectra of six different age healthy volunteers were collected using Nexus-870 and its NIR fiber port smart accessory. The test was implemented with changing the blood glucose concentration for the limosis and satiation of every volunteer. The calibration model was set up using PLS method with the smoothing, baseline correction and first derivatives pretreatment spectrum in the 7 500-8 500 cm-1 region for single volunteer, the same age combination and that of different age. When the spectrum was obtained, the actual blood glucose value of every spectrum sample was demarcated using ultraviolet spectrophotometer. The correlation between the calibration value and true value for single volunteer is better than that for the combination of volunteers, the correlative coefficients are all over 0.904 71, RMSECs are all less than 0.171.
[1] SHEN Yao-chun, WANG Chun,Mmac Kenzie H A,et al(沈耀春,王 春,Mmac Kenzie H A等). Journal of Nanjing University(Natural Science)(南京大学学报·自然科学), 2000,36(5):632. [2] Michael J Mcshane, Gerard L, Clifford Spiegelman. SPIE,1997,2982:189. [3] Vidi A Saptari,Kamal Youcef-toumi. SPIE,2000,4163:45. [4] LU Wan-zhen,YUAN Hong-fu,XU Guang-tong,et al(陆婉珍,袁洪福,徐广通, 等). Modern Analysis Technique of Near-infrared Spectrum(现代近红外光谱分析技术). Beijing: China Petrochemical Press(北京: 中国石化出版社),2000. [5] ZHU Er-yi, YANG Peng-yuan(朱尔一 ,杨芃原). Chemistry Metrology Technique and Application(化学计量学技术及应用). Beijing:Sciences Press(北京: 科学出版社),2001. [6] CUI Ji-cheng,DING Dong,ZHANG Hong-yan,et al(崔继承,丁 东,张洪艳,等). Chinese Journal of Spectroscopy Laboratory(光谱实验室),2003,20(2):257. [7] Stephen F Malin,Timothy L Ruchti, Thomas B Blank,et al. Clinical Chemistry, 1999, 45(9): 1651. [8] HUANG Lan, DING Hai-shu, WANG Guang-zhi(黄 岚, 丁海曙, 王广志). Spectroscopy and Spectral Analysis(光谱学与光谱分析), 2002, 22(3): 387.