A New NIR Calibration Transfer Method Based on Parameter Correction
HU Yun1, LI Bo-yan2*, ZHANG Jin2, PENG Qian-rong1
1. Technology Center, China Tobacco Guizhou Industrial Co., Ltd., Guiyang 550009, China
2. College of Food Science, Guizhou Medical University, Guiyang 550025, China
Abstract:Model transfer plays an important role in solving the problem of the difference between near infrared (NIR) spectroscopic instruments and the prediction difficulty of models. The spectral differences between the same samples taken on different NIR instruments were identified using the principal component-Mahalanobis distance method. Based on the constrain conditions of Tikhonov regularization (TR) and model parameter correction, a new algorithm (called new Tikhonov regularization-based calibration transfer, NTRCT) was proposed for calibration transfer between NIR instruments, so as to facilitate the share and use of the calibration models. The spectra of a set of standard samples were first utilized to establish a specific function that could minimize the prediction errors obtained from the master and slave instrumental models. By constraining the difference of the model parameters, the parameters of the slave instrument model were then determined, to achieve the purpose of model transfer from the master instrument to the slave one. This method was applied to analyze the content of the active pharmaceutical ingredient (API) of tablets and quantify the contents of total alkaloids and total sugars in tobacco leaves respectively, by means of their NIR spectra acquired on different instruments. The results showed that the root means square error of prediction (RMSEP) of samples taken on the salve instrument was reduced from 8.3 mg, 0.49% and 1.91% to 3.9 mg, 0.09% and 0.83% respectively; when 15 standard samples were employed for modelling. As the calibration transferred all the resulting RPD values were larger than 3.0, and the sample predictions from the salve instrumental spectra were thus significantly improved. The method was explicit and intuitive in theory, and had good accuracy in sample prediction in practical applications. It provided a new idea for calibration transfer method with standard samples.
胡 芸,李博岩,张 进,彭黔荣. 基于参数校正的近红外光谱模型转移新方法[J]. 光谱学与光谱分析, 2020, 40(06): 1804-1808.
HU Yun, LI Bo-yan, ZHANG Jin, PENG Qian-rong. A New NIR Calibration Transfer Method Based on Parameter Correction. SPECTROSCOPY AND SPECTRAL ANALYSIS, 2020, 40(06): 1804-1808.
[1] Feudale R N, Woody N A, Tan H, et al. Chemometrics and Intelligent Laboratory Systems, 2002, 64(2): 181.
[2] Workman J J. Applied Spectroscopy, 2018, 72(3): 340.
[3] SHI Yun-ying, LI Jing-yan, CHU Xiao-li(史云颖, 李敬岩, 褚小立). Chinese Journal of Analytical Chemistry(分析化学), 2019, 47(4): 479.
[4] ZHANG Jin, CAI Wen-sheng, SHAO Xue-guang(张 进, 蔡文生, 邵学广). Progress in Chemistry(化学进展), 2017, 29(8): 902.
[5] Wang Y, Veltkamp D J, Kowalski B R. Analytical Chemistry, 1991, 63(23): 2750.
[6] Du W, Chen Z, Zhong L, Wang S, et al. Analytica Chimica Acta, 2011, 690(1): 64.
[7] Zhang J, Guo C, Cui X, et al. Analytica Chimica Acta, 2019, 1050: 25.
[8] Chen W R, Bin J, Lu H M, et al. Analyst, 2016, 141(6): 1973.
[9] Forina M, Drava G, Armanino C, et al. Chemometrics and Intelligent Laboratory Systems, 1995, 27(2): 189.
[10] Kalivas J H. Journal of Chemometrics, 2012, 26(6): 218.
[11] Liu Y, Cai W, Shao X. Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy, 2016, 169: 197.
[12] Williams P C, Sobering D C. Journal of Near Infrared Spectroscopy, 1993, 1(1): 25.